Collimator; image quality; image sampling; instrumentation; molecular imaging; multiplexing; overlap; pinhole; single photon emission computed tomography (SPECT); system design; SLIT-SLAT COLLIMATOR; SAMPLING COMPLETENESS; RECONSTRUCTION; PROJECTIONS; BIODISTRIBUTION; TOMOGRAPHY; SIMULATION; ARTIFACTS; ALGORITHM; LOFTHOLE;
D O I
10.1109/TMI.2014.2361051
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Multi-pinhole collimators are often used in pre-clinical SPECT systems because they have a better resolution-sensitivity tradeoff than parallel hole collimators when imaging small objects. Most multi-pinhole collimators are designed to allow no or only a limited amount of overlap between the different pinhole projections because the ambiguity introduced by multiplexing pinholes can result in artifacts. The origin of these artifacts is still not fully understood, but previous research has already shown that data incompleteness could be part of the explanation. Therefore, we developed a method to investigate data completeness in multiplexing multi-pinhole systems and showed that a certain activity distribution can be successfully reconstructed when the nonmultiplexed data is complete or when the overlap can be sufficiently de-multiplexed. We validated this method using computer simulated phantom data of different multiplexing systems. We also studied contrast-to-noise and nonprewhitening matched filter signal-to-noise ratio (NPW-SNR) to compare the image quality in a single pinhole system with multiplexing systems. We found that our method can indeed be used to evaluate data completeness in multiplexing systems and found no artifacts in the systems that had complete data. Sensitivity increased significantly with multiplexing but we found only small, nonsignificant differences in contrast-to-noise ratio. However, the NPW-SNR did slightly improve in the multiplexing setups. We conclude that more multiplexing does not necessarily result in more artifacts and that even a high amount of multiplexing can still result in artifact-free images if the nonmultiplexed data is complete or when the overlap can be sufficiently de-multiplexed.
引用
收藏
页码:474 / 486
页数:13
相关论文
共 50 条
  • [1] Data completeness in multiplexing multi-pinhole SPECT
    Van Audenhaege, Karen
    Vanhove, Christian
    Vandenberghe, Stefaan
    Van Holen, Roel
    [J]. 2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [2] Evaluation of multiplexing in multi-pinhole SPECT
    Jorgensen, Aaron K.
    Zeng, Gengsheng L.
    [J]. 2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2213 - 2217
  • [3] The effects of object activity distribution on multiplexing multi-pinhole SPECT
    Mok, Greta S. P.
    Tsui, Benjamin M. W.
    Beekman, Freek J.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (08): : 2635 - 2650
  • [4] Effects of object activity distribution on multiplexing multi-pinhole SPECT
    Mok, Greta
    Tsui, Benjamin
    Beekman, Freek
    Wang, Yuchuan
    Ahuja, Anil
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [5] Development and evaluation of quantitative multi-pinhole SPECT image reconstruction methods
    Yu, Jianhua
    Xu, Jingyan
    Wang, Yuchuan
    Lian, Bing
    Hugg, James
    Tsui, Benjamin
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [6] Improved Quantitative DAT Imaging with Multiplexing in a Multi-pinhole SPECT System
    Konik, Arda
    Zeraatkar, Navid
    Kalluri, Kesava
    Auer, Benjamin
    Furenlid, Lars R.
    King, Michael A.
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [7] Targeted multi-pinhole SPECT
    Branderhorst, Woutjan
    Vastenhouw, Brendan
    van der Have, Frans
    Blezer, Erwin L. A.
    Bleeker, Wim K.
    Beekman, Freek J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 (03) : 552 - 561
  • [8] Design and Sampling Completeness Evaluation of Scanning Orbits in Multi-pinhole Small Animal SPECT Imaging
    Dai, Tiantian
    Ma, Tianyu
    Wei, Qingyang
    Wang, Shi
    Liu, Hui
    Cui, Junjian
    Huang, Qijie
    Liu, Yaqiang
    [J]. 2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2836 - 2839
  • [9] Targeted multi-pinhole SPECT
    Woutjan Branderhorst
    Brendan Vastenhouw
    Frans van der Have
    Erwin L. A. Blezer
    Wim K. Bleeker
    Freek J. Beekman
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2011, 38 : 552 - 561
  • [10] Performance evaluation of a novel multi-pinhole SPECT system
    Jiang, Nianming
    Liu, Hui
    Xue, Mengle
    Li, Cheng
    Gao, Lilei
    Liu, Fan
    Wu, Jing
    Liu, Yaqiang
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2024, 1059