Technical Note: Commissioning of an ultrasound-compatible surrogate vaginal cylinder for transvaginal ultrasound-based gynecologic high-dose-rate brachytherapy

被引:4
|
作者
Van Elburg, Devin [1 ,2 ]
Roumeliotis, Michael [1 ,2 ,3 ]
Fenster, Aaron [4 ,5 ]
Phan, Tien [3 ]
Meyer, Tyler [1 ,2 ,3 ]
机构
[1] Univ Calgary, Dept Phys & Astron, Calgary, AB T2N 1N4, Canada
[2] Tom Baker Canc Clin, Med Phys Dept, Calgary, AB, Canada
[3] Univ Calgary, Dept Oncol, Calgary, AB, Canada
[4] Univ Western Ontario, Sch Biomed Engn, London, ON, Canada
[5] Univ Western Ontario, Robarts Res Inst, London, ON, Canada
关键词
brachytherapy; commissioning; gynecologic; ultrasound; GUIDED ADAPTIVE BRACHYTHERAPY; TRANSRECTAL ULTRASOUND; IMAGE REGISTRATION; CERVIX CANCER; QUALITY;
D O I
10.1002/mp.15559
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To provide a comprehensive set of commissioning tests for clinical implementation of three-dimensional transvaginal ultrasound (3D TVUS) as a replacement of computed tomography (CT) for applicator reconstruction in gynecologic (GYN) intracavitary high-dose-rate brachytherapy (HDRBT) with a multi-channel vaginal cylinder (MCVC). Methods We introduce an ultrasound-compatible "surrogate" vaginal cylinder (SVC) for reconstruction of Elekta's CT-MR Multi Channel Applicator (MCVC) in 3D TVUS. The MCVC is digitized over the SVC in 3DUS using digital library model overlay. Consulting guidelines from various sources (CPQR, GEC-ESTRO, AAPM), we identify and describe three tests specific to commissioning the SVC: (1) verification of SVC outer dimensions, (2) source position accuracy of MCVC digitization over the SVC in 3D TVUS, and (3) MRI/US registration error. Results The SVC outer dimensions (diameter and A-D marker locations) were well matched to the MCVC, however a 0.6 mm discrepancy in length between cylinder tips was observed. Source position accuracy was within 1 mm (tolerance recommended by CPQR) when reconstructing the MCVC in 3D TVUS. Dice similarity coefficients and target registration error for MRI/3D TVUS registration was similar to MRI/CT registration, which is the clinical standard. Conclusions These commissioning tests are performed using institutional equipment but provide the framework for any practitioner to repeat in their own setup, to demonstrate safe adoption of the 3D TVUS system for patient treatments. We demonstrate that MRI/US-based workflow achieves similar source position accuracy and image registration error as standard MRI/CT, which is consistent with standard tolerances. This is a critical step toward replacement of CT with US in GYN HDRBT treatments with the MCVC.
引用
收藏
页码:2203 / 2211
页数:9
相关论文
共 50 条
  • [41] Automated Needle Digitization in Ultrasound-based Prostate High Dose-rate Brachytherapy Using a Deep Learning Algorithm
    Wang, F.
    Xing, L.
    Bagshaw, H. P.
    Buyyounouski, M. K.
    Han, B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2020, 108 (03): : S129 - S130
  • [42] Enabling intraoperative implant assessment during high-dose-rate interstitial gynecologic brachytherapy using a 3D ultrasound guidance system
    Rodgers, Jessica
    Bax, Jeffrey
    Surry, Kathleen
    Hrinivich, William Thomas
    Rascevska, Elina
    Leung, Erica
    Velker, Vikram
    D'Souza, David
    Fenster, Aaron
    MEDICAL PHYSICS, 2019, 46 (11) : 5399 - 5399
  • [43] Dose Uncertainty Analysis for Needle-Tip Identification Error in Transrectal Ultrasound-Based High Dose-Rate Prostate Brachytherapy
    Fan, J.
    Veltchev, I.
    Lin, T.
    Horwitz, E. M.
    Hallman, M. A.
    Price, R. A., Jr.
    Ma, C. M. C.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E657 - E658
  • [44] The challenges facing deep learning-based catheter localization for ultrasound guided high-dose-rate prostate brachytherapy
    Liu, Derek
    Tupor, Shayantonee
    Singh, Jaskaran
    Chernoff, Trey
    Leong, Nelson
    Sadikov, Evgeny
    Amjad, Asim
    Zilles, Sandra
    MEDICAL PHYSICS, 2022, 49 (04) : 2442 - 2451
  • [45] Field distortion compensation for electromagnetic tracking of ultrasound probes with application in high-dose-rate prostate brachytherapy
    Lugez, Elodie
    Sadjadi, Hossein
    Joshi, Chandra P.
    Hashtrudi-Zaad, Keyvan
    Akl, Selim G.
    Fichtinger, Gabor
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (03):
  • [46] Ultrasound Sonographic Guided High-Dose-Rate Intracavitary Brachytherapy for Medically Inoperable Uterine Endometrial Cancer
    Kaneyasu, Y. O.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S419 - S419
  • [47] Spatial, Dosimetric, and Radiobiological Robustness of Template-Based Interstitial High-Dose-Rate Gynecologic Brachytherapy
    Ostyn, M.
    Burke, A.
    Fields, E.
    Todor, D.
    MEDICAL PHYSICS, 2019, 46 (06) : E232 - E232
  • [48] Template-based high-dose-rate interstitial brachytherapy in gynecologic cancers: A single institutional experience
    Mahantshetty, Umesh
    Shrivastava, Shyamkishore
    Kalyani, Nikhil
    Banerjee, Susovan
    Engineer, Reena
    Chopra, Supriya
    BRACHYTHERAPY, 2014, 13 (04) : 337 - 342
  • [49] Dose Optimization for Single-Channel Vaginal Cylinder High-Dose-Rate Brachytherapy: A Double Prescription Method for Patients With Endometrial Adenocarcinoma
    Wu, Xingen
    McDonald, Andrew
    Shen, Sui
    Stanley, Dennis
    Popple, Richard
    Marcrom, Samuel
    Kim, Robert
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (06)
  • [50] Observer preference of artificial intelligence-generated versus clinical prostate contours for ultrasound-based high dose rate brachytherapy
    King, Martin T.
    Kehayias, Christopher E.
    Chaunzwa, Tafadzwa
    Rosen, Daniel B.
    Mahal, Amandeep R.
    Wallburn, Tyler D.
    Milligan, Michael G.
    Dyer, M. Aiven
    Nguyen, Paul L.
    Orio, Peter F.
    Harris, Thomas C.
    Buzurovic, Ivan
    Guthier, Christian V.
    MEDICAL PHYSICS, 2023, 50 (10) : 5935 - 5943