Technical Note: Commissioning of an ultrasound-compatible surrogate vaginal cylinder for transvaginal ultrasound-based gynecologic high-dose-rate brachytherapy

被引:4
|
作者
Van Elburg, Devin [1 ,2 ]
Roumeliotis, Michael [1 ,2 ,3 ]
Fenster, Aaron [4 ,5 ]
Phan, Tien [3 ]
Meyer, Tyler [1 ,2 ,3 ]
机构
[1] Univ Calgary, Dept Phys & Astron, Calgary, AB T2N 1N4, Canada
[2] Tom Baker Canc Clin, Med Phys Dept, Calgary, AB, Canada
[3] Univ Calgary, Dept Oncol, Calgary, AB, Canada
[4] Univ Western Ontario, Sch Biomed Engn, London, ON, Canada
[5] Univ Western Ontario, Robarts Res Inst, London, ON, Canada
关键词
brachytherapy; commissioning; gynecologic; ultrasound; GUIDED ADAPTIVE BRACHYTHERAPY; TRANSRECTAL ULTRASOUND; IMAGE REGISTRATION; CERVIX CANCER; QUALITY;
D O I
10.1002/mp.15559
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To provide a comprehensive set of commissioning tests for clinical implementation of three-dimensional transvaginal ultrasound (3D TVUS) as a replacement of computed tomography (CT) for applicator reconstruction in gynecologic (GYN) intracavitary high-dose-rate brachytherapy (HDRBT) with a multi-channel vaginal cylinder (MCVC). Methods We introduce an ultrasound-compatible "surrogate" vaginal cylinder (SVC) for reconstruction of Elekta's CT-MR Multi Channel Applicator (MCVC) in 3D TVUS. The MCVC is digitized over the SVC in 3DUS using digital library model overlay. Consulting guidelines from various sources (CPQR, GEC-ESTRO, AAPM), we identify and describe three tests specific to commissioning the SVC: (1) verification of SVC outer dimensions, (2) source position accuracy of MCVC digitization over the SVC in 3D TVUS, and (3) MRI/US registration error. Results The SVC outer dimensions (diameter and A-D marker locations) were well matched to the MCVC, however a 0.6 mm discrepancy in length between cylinder tips was observed. Source position accuracy was within 1 mm (tolerance recommended by CPQR) when reconstructing the MCVC in 3D TVUS. Dice similarity coefficients and target registration error for MRI/3D TVUS registration was similar to MRI/CT registration, which is the clinical standard. Conclusions These commissioning tests are performed using institutional equipment but provide the framework for any practitioner to repeat in their own setup, to demonstrate safe adoption of the 3D TVUS system for patient treatments. We demonstrate that MRI/US-based workflow achieves similar source position accuracy and image registration error as standard MRI/CT, which is consistent with standard tolerances. This is a critical step toward replacement of CT with US in GYN HDRBT treatments with the MCVC.
引用
收藏
页码:2203 / 2211
页数:9
相关论文
共 50 条
  • [21] A Dosimetric Study by Using a Multichannel Cylinder in High-Dose-Rate Brachytherapy Treatment of the Vaginal Cuff
    Syh, J. H.
    Zhang, M.
    White, M.
    Patel, B.
    Syh, J.
    Song, X.
    Wu, H.
    Rosen, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S501 - S501
  • [22] High-Dose-Rate Interstitial Brachytherapy for Deeply Situated Gynecologic Tumors Guided by Combination of Transrectal and Transabdominal Ultrasonography: A Technical Note
    Shimizu, Yuri
    Murakami, Naoya
    Chiba, Takahito
    Kaneda, Tomoya
    Okamoto, Hiroyuki
    Nakamura, Satoshi
    Takahashi, Ayaka
    Kashihara, Tairo
    Takahashi, Kana
    Inaba, Koji
    Okuma, Kae
    Nakayama, Yuko
    Itami, Jun
    Igaki, Hiroshi
    FRONTIERS IN ONCOLOGY, 2022, 11
  • [23] Evaluation of Dose Uncertainty for in Trans-Rectal Ultrasound (TRUS) - Based High-Dose-Rate Prostate Brachytherapy
    Fan, J.
    Veltchev, I.
    Lin, T.
    Chibani, O.
    Price, R.
    Ma, C.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [24] A consensus-based, process commissioning template for high-dose-rate gynecologic treatments
    Browns, Derek W.
    Damato, Antonio L.
    Sutlief, Steven
    Morcovescu, Serban
    Park, Sang-June
    Reiff, Jay
    Shih, Allen
    Scanderbeg, Daniel J.
    BRACHYTHERAPY, 2016, 15 (05) : 570 - 577
  • [25] Building a High-Dose-Rate Prostate Brachytherapy Program With Real-Time Ultrasound-Based Planning: Initial Safety, Quality, and Outcome Results
    Zhang, Hong
    Kang, Sohyun
    Ali, Naba
    Baran, Andrea
    Bylund, Kevin
    Gentile, David
    Previte, Gregory
    Matloubieh, Ahmad
    Gray, Alex
    Marou, Nancy
    ADVANCES IN RADIATION ONCOLOGY, 2020, 5 (03) : 388 - 395
  • [26] Use of ultrasound in image-guided high-dose-rate brachytherapy: enumerations and arguments
    Banerjee, Susovan
    Kataria, Tejinder
    Gupta, Deepak
    Goyal, Shikha
    Bisht, Shyam Singh
    Basu, Trinanjan
    Abhishek, Ashu
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2017, 9 (02) : 146 - 150
  • [27] Needle Identification in High-Dose-Rate Prostate Brachytherapy using Ultrasound Imaging Modality
    Buzurovic, Ivan
    Misic, Vladimir
    Yu, Yan
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 476 - 479
  • [28] An Optimization Study and Comparison of a Multichannel Vaginal Cylinder Versus Single Channel Cylinder Applicator in High-dose-rate Brachytherapy
    Syh, J. H.
    Wu, T.
    Syh, J.
    Patel, B.
    Rosen, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S456 - S457
  • [29] 360-degree 3D transvaginal ultrasound system for high-dose-rate interstitial gynaecological brachytherapy needle guidance
    Rodgers, Jessica R.
    Surry, Kathleen
    Souza, David D'
    Leung, Eric
    Fenster, Aaron
    MEDICAL IMAGING 2017: ULTRASONIC IMAGING AND TOMOGRAPHY, 2017, 10139
  • [30] Ultrasound-planned high-dose-rate prostate brachytherapy: Dose painting to the dominant intraprostatic lesion
    Crook, Juanita
    Ots, Ana
    Gaztanaga, Miren
    Schmid, Matt
    Araujo, Cynthia
    Hilts, Michelle
    Batchelar, Deidre
    Parker, Brent
    Bachand, Francois
    Milette, Marie-Pierre
    BRACHYTHERAPY, 2014, 13 (05) : 433 - 441