Evaluation of volumetric blood collection devices for the measurement of phenylalanine and tyrosine to monitor patients with phenylketonuria

被引:16
|
作者
Carling, Rachel S. [1 ,2 ,5 ]
Emmett, Erin C. [2 ]
Moat, Stuart J. [3 ,4 ]
机构
[1] Kings Coll London, GKT Sch Med Educ, London, England
[2] Guys & St Thomas NHSFT, Biochem Sci, London, England
[3] Univ Hosp Wales, Dept Med Biochem Immunol & Toxicol, Cardiff, Wales
[4] Cardiff Univ, Univ Hosp Wales, Sch Med, Cardiff, Wales
[5] St Thomas Hosp, Biochem Sci, 4th Floor,North Wing,Westminster Bridge Rd, London SE1, England
关键词
Blood collection devices; Dried blood spots; DBS; Phenylalanine; Phenylketonuria; PKU; Microsampling; Tyrosine; Standardization; AMINO-ACIDS; HEMATOCRIT; PUNCH;
D O I
10.1016/j.cca.2022.08.005
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Measurement of dried blood spot (DBS) phenylalanine (Phe) is central to the monitoring of patients with phenylketonuria. However, the volume and hematocrit (Hct) of the blood applied to conventional DBS cards significantly affects analytical results. Volumetric blood collection devices are reported to be more accurate, precise and less prone to Hct effects.Methods: Accuracy, imprecision, effect of blood volume and Hct were evaluated for measurement of Phe and tyrosine using three volumetric devices and compared with the conventional PerkinElmer-226 filter-paper collection devices. i.e. conventional DBS cards. Applicability for use in a clinical laboratory was assessed qualitatively.Results: Blood volume did not impact on the performance of the volumetric devices; however, significant biases were observed with the conventional DBS card. A higher Hct introduced unacceptable bias for Neoteryx-Mitra and conventional DBS card. All devices had a mean relative standard deviation (RSD) <= 4.1 %, except for the Neoteryx-Mitra (<= 6.2 %). Relative to liquid blood, the mean biases of Phe for the various devices were-5.1 (HemaXis-DB10),-7.8 (Capitainer-qDBS),-12.0 (Neoteryx-Mitra) and-32.6 % (conventional DBS card).Conclusions: Introducing volumetric collection devices will overcome the significant pre-analytical issues asso-ciated with conventional DBS collection and improve the biochemical monitoring of patients with PKU.
引用
下载
收藏
页码:157 / 166
页数:10
相关论文
共 50 条
  • [41] Diurnal variation of phenylalanine and tyrosine concentrations in adult patients with phenylketonuria: subcutaneous microdialysis is no adequate tool for the determination of amino acid concentrations
    Gruenert, Sarah C.
    Brichta, Corinna M.
    Krebs, Andreas
    Clement, Hans-Willi
    Rauh, Reinhold
    Fleischhaker, Christian
    Hennighausen, Klaus
    Sass, Joern Oliver
    Schwab, K. Otfried
    NUTRITION JOURNAL, 2013, 12
  • [42] Evaluation of the Performance and Hematocrit Independence of the HemaPEN as a Volumetric Dried Blood Spot Collection Device
    Deprez, Sigrid
    Paniagua-Gonzalez, Lucia
    Velghe, Sofie
    Stove, Christophe P.
    ANALYTICAL CHEMISTRY, 2019, 91 (22) : 14467 - 14475
  • [43] Diurnal variation of phenylalanine and tyrosine concentrations in adult patients with phenylketonuria: subcutaneous microdialysis is no adequate tool for the determination of amino acid concentrations
    Sarah C Grünert
    Corinna M Brichta
    Andreas Krebs
    Hans-Willi Clement
    Reinhold Rauh
    Christian Fleischhaker
    Klaus Hennighausen
    Jörn Oliver Sass
    K Otfried Schwab
    Nutrition Journal, 12
  • [44] A low abundance of genus Bacteroides in gut microbiota is negatively correlated with blood phenylalanine levels in Uygur patients with phenylketonuria
    Su, Yajie
    Shadike, Qiaolibang
    Wang, Mingbang
    Jiang, Haili
    Liu, Wanying
    Liu, Jingfang
    Tuerdi, Rena
    Zhou, Wenhao
    Li, Long
    TRANSLATIONAL PEDIATRICS, 2021, 10 (10) : 2521 - +
  • [45] No evidence for individual blood-brain barrier phenylalanine transport to influence clinical outcome in typical phenylketonuria patients
    Pietz, J
    Rupp, A
    Burgard, P
    Boesch, C
    Kreis, R
    ANNALS OF NEUROLOGY, 2002, 52 (03) : 382 - 383
  • [46] Simultaneous measurement of phenylalanine and tyrosine in phenylketonuric plasma and dried blood by high-performance liquid chromatography
    Dale, Y
    Mackey, V
    Mushi, R
    Nyanda, A
    Maleque, M
    Ike, J
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 788 (01): : 1 - 8
  • [47] Sapropterin therapy increases stability of blood phenylalanine levels in patients with BH4-responsive phenylketonuria (PKU)
    Burton, Barbara K.
    Bausell, Heather
    Katz, Rachel
    LaDuca, Holly
    Sullivan, Christine
    MOLECULAR GENETICS AND METABOLISM, 2010, 101 (2-3) : 110 - 114
  • [48] The 48-hour tetrahydrobiopterin loading test in patients with phenylketonuria: Evaluation of protocol and influence of baseline phenylalanine concentration
    Anjema, K.
    Venema, G.
    Hofstede, F. C.
    Weber, E. C. Carbasius
    Bosch, A. M.
    Ter Horst, N. M.
    Hollak, C. E. M.
    Jonkers, C. F.
    Rubio-Gozalbo, M. E.
    van der Ploeg, E. M. C.
    de Vries, M. C.
    Janssen-Regelink, R. G.
    Janssen, M. C. H.
    Essen, H. Zweers-van
    Boelen, C. C. A.
    van der Herberg-van de Wetering, N. A. P.
    Heiner-Fokkema, M. R.
    van Rijn, M.
    van Spronsen, F. J.
    MOLECULAR GENETICS AND METABOLISM, 2011, 104 : S60 - S63
  • [49] AN AUTOMATED ENZYMATIC METHOD ON THE ROCHE COBAS MIRA(TM)-S FOR MONITORING PHENYLALANINE IN DRIED BLOOD SPOTS OF PATIENTS WITH PHENYLKETONURIA
    RANDELL, EW
    LEHOTAY, DC
    CLINICAL CHEMISTRY, 1995, 41 (06) : S96 - S97
  • [50] An automated enzymatic method on the Roche COBAS MIRA(TM) S for monitoring phenylalanine in dried blood spots of patients with phenylketonuria
    Randell, EW
    Lehotay, DC
    CLINICAL BIOCHEMISTRY, 1996, 29 (02) : 133 - 138