Evaluation of volumetric blood collection devices for the measurement of phenylalanine and tyrosine to monitor patients with phenylketonuria

被引:16
|
作者
Carling, Rachel S. [1 ,2 ,5 ]
Emmett, Erin C. [2 ]
Moat, Stuart J. [3 ,4 ]
机构
[1] Kings Coll London, GKT Sch Med Educ, London, England
[2] Guys & St Thomas NHSFT, Biochem Sci, London, England
[3] Univ Hosp Wales, Dept Med Biochem Immunol & Toxicol, Cardiff, Wales
[4] Cardiff Univ, Univ Hosp Wales, Sch Med, Cardiff, Wales
[5] St Thomas Hosp, Biochem Sci, 4th Floor,North Wing,Westminster Bridge Rd, London SE1, England
关键词
Blood collection devices; Dried blood spots; DBS; Phenylalanine; Phenylketonuria; PKU; Microsampling; Tyrosine; Standardization; AMINO-ACIDS; HEMATOCRIT; PUNCH;
D O I
10.1016/j.cca.2022.08.005
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Measurement of dried blood spot (DBS) phenylalanine (Phe) is central to the monitoring of patients with phenylketonuria. However, the volume and hematocrit (Hct) of the blood applied to conventional DBS cards significantly affects analytical results. Volumetric blood collection devices are reported to be more accurate, precise and less prone to Hct effects.Methods: Accuracy, imprecision, effect of blood volume and Hct were evaluated for measurement of Phe and tyrosine using three volumetric devices and compared with the conventional PerkinElmer-226 filter-paper collection devices. i.e. conventional DBS cards. Applicability for use in a clinical laboratory was assessed qualitatively.Results: Blood volume did not impact on the performance of the volumetric devices; however, significant biases were observed with the conventional DBS card. A higher Hct introduced unacceptable bias for Neoteryx-Mitra and conventional DBS card. All devices had a mean relative standard deviation (RSD) <= 4.1 %, except for the Neoteryx-Mitra (<= 6.2 %). Relative to liquid blood, the mean biases of Phe for the various devices were-5.1 (HemaXis-DB10),-7.8 (Capitainer-qDBS),-12.0 (Neoteryx-Mitra) and-32.6 % (conventional DBS card).Conclusions: Introducing volumetric collection devices will overcome the significant pre-analytical issues asso-ciated with conventional DBS collection and improve the biochemical monitoring of patients with PKU.
引用
下载
收藏
页码:157 / 166
页数:10
相关论文
共 50 条
  • [31] PLASMA PHENYLALANINE AND TYROSINE RESPONSES TO DIFFERENT NUTRITIONAL CONDITIONS (FASTING/POSTPRANDIAL) IN PATIENTS WITH PHENYLKETONURIA - EFFECT OF SAMPLE TIMING
    VANSPRONSEN, FJ
    VANRIJN, M
    VANDIJK, T
    SMIT, GPA
    REIJNGOUD, DJ
    BERGER, R
    HEYMANS, HSA
    PEDIATRICS, 1993, 92 (04) : 570 - 573
  • [32] The Effect of Various Doses of Phenylalanine Supplementation on Blood Phenylalanine and Tyrosine Concentrations in Tyrosinemia Type 1 Patients
    van Ginkel, Willem G.
    van Reemst, Hannah E.
    Kienstra, Nienke S.
    Daly, Anne
    Rodenburg, Iris L.
    MacDonald, Anita
    Burgerhof, Johannes G. M.
    de Blaauw, Pim
    van de Krogt, Jennifer
    Santra, Saikat
    Heiner-Fokkema, M. Rebecca
    van Spronsen, Francjan J.
    NUTRIENTS, 2019, 11 (11)
  • [33] Blood Phenylalanine Levels in Patients with Phenylketonuria from Europe between 2012 and 2018: Is It a Changing Landscape?
    Pinto, Alex
    Ahring, Kirsten
    Almeida, Manuela Ferreira
    Ashmore, Catherine
    Belanger-Quintana, Amaya
    Burlina, Alberto
    Coskun, Turgay
    Daly, Anne
    van Dam, Esther
    Dursun, Ali
    Evans, Sharon
    Feillet, Francois
    Gizewska, Maria
    Goekmen-oezel, Hulya
    Hickson, Mary
    Hoekstra, Yteke
    Ilgaz, Fatma
    Jackson, Richard
    Lesniak, Alicja
    Loro, Christian
    Malicka, Katarzyna
    Patalan, Michal
    Rocha, Julio Cesar
    Sivri, Serap
    Rodenburg, Iris
    van Spronsen, Francjan
    Straczek, Kamilla
    Tokatli, Ayseguel
    Macdonald, Anita
    NUTRIENTS, 2024, 16 (13)
  • [34] Evaluation of blood collection tubes specific for homocysteine measurement
    de Graaf, Aniek O.
    de Jonge, Robert
    Velmans, Mathieu H.
    Swaanenburg, Joost C. J. M.
    Janssen, Marcel J. W.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2008, 46 (11) : 1652 - 1654
  • [35] Optimization and validation of a bioanalytical HPLC-UV technique for simultaneous determination of underivatized phenylalanine and tyrosine in the blood for phenylketonuria diagnosis and monitoring
    Rabti, Hadjira
    Amrane, Mounira
    Lalaouna, Abdeldjalil
    Flilissa, Abdenacer
    Benguerba, Yacine
    BIOMEDICAL CHROMATOGRAPHY, 2024, 38 (01)
  • [36] Use of sodium measurement as a surrogate volume measurement in unmetered-blood collection devices
    Maguire, GA
    Hales, CN
    CLINICAL CHEMISTRY, 2002, 48 (03) : 592 - 593
  • [37] The evaluation of phenylalanine levels in Estonian phenylketonuria patients during eight years by electronic laboratory records
    Lillevali, Hardo
    Reinson, Karit
    Muru, Kai
    Saarsalu, Siret
    Kunnapas, Kadi
    Kahre, Tiina
    Murumets, Ulle
    Ounap, Katrin
    MOLECULAR GENETICS AND METABOLISM REPORTS, 2019, 19
  • [38] Predicting the phenylalanine blood concentration from urine analyses. An approach to noninvasive monitoring of patients with phenylketonuria
    Langenbeck, U
    Baum, F
    Mench-Hoinowski, A
    Luthe, H
    Behbehani, A
    JOURNAL OF INHERITED METABOLIC DISEASE, 2005, 28 (06) : 855 - 861
  • [39] Estimating the probability of IQ impairment from blood phenylalanine for phenylketonuria patients: a hierarchical meta-analysis
    Fonnesbeck, Christopher J.
    McPheeters, Melissa L.
    Krishnaswami, Shanthi
    Lindegren, Mary Louise
    Reimschisel, Tyler
    JOURNAL OF INHERITED METABOLIC DISEASE, 2013, 36 (05) : 757 - 766
  • [40] Evaluation of the interchangeability of blood collection tubes for the measurement of trace elements
    Delgado Rodriguez, J. A.
    Forteza Ferrer, F.
    Gutierrez Sanchez, S.
    Robles Bauza, J.
    Bauca, J. M.
    CLINICA CHIMICA ACTA, 2022, 530 : S367 - S367