Ring Artifact Correction for Phase-Insensitive Ultrasound Computed Tomography

被引:4
|
作者
Baker, Christian [1 ,2 ]
Sarno, Daniel [2 ]
Eckersley, Robert J. [1 ]
Zeqiri, Bajram [2 ]
机构
[1] Kings Coll London, Biomed Engn Dept, Div Imaging Sci, London SE1 7EH, England
[2] Natl Phys Lab, Ultrasound & Underwater Acoust Grp, Teddington TW11 0LW, Middx, England
关键词
Transducers; Sensitivity; Ultrasonic imaging; Acoustics; Breast; Attenuation; Ultrasonic variables measurement; Artifact correction; computed tomography (CT); medical diagnostic imaging; ultrasonic imaging; SOUND SPEED; FLAT-FIELDS; RECONSTRUCTION; NORMALIZATION; ATTENUATION; REMOVAL;
D O I
10.1109/TUFFC.2019.2948429
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
An algorithm was developed for the correction of ring artifacts in phase-insensitive ultrasound computed tomography attenuation images. Differences in the measurement sensitivity between the ultrasound transducer array elements cause discontinuities in the sinogram which manifest as rings and arcs in the reconstructed image. The magnitudes of the discontinuities are potentially time-varying and dependent on the attenuation being measured. The algorithm dynamically determines the measurement sensitivity of each transducer in the array during the scan by comparison with both the elements to its left and the elements to its right. Elements at either end of the array are corrected, assuming a zero-attenuation path. The two estimates of sensitivity are combined using a weighted mean similar to a Kalman filter. The algorithm was tested on simulated and experimentally acquired data. It was demonstrated to reduce the root-mean-square error (RMSE) of simulated images against ground-truth images by up to a factor of 50 compared with uncorrected images and to visibly reduce artifacts on images reconstructed from the experimentally acquired data.
引用
收藏
页码:513 / 525
页数:13
相关论文
共 50 条
  • [21] PHASE-INSENSITIVE PULSE-ECHO IMAGING
    ODONNELL, M
    ULTRASONIC IMAGING, 1982, 4 (04) : 321 - 335
  • [22] Prior-based artifact correction (PBAC) in computed tomography
    Heusser, Thorsten
    Brehm, Marcus
    Ritschl, Ludwig
    Sawall, Stefan
    Kachelriess, Marc
    MEDICAL PHYSICS, 2014, 41 (02)
  • [23] Evaluation of the dose distribution of tomotherapy using polymer gel dosimeters and optical computed tomography with ring artifact correction
    Chou, Ying-Hsiang
    Lu, Yu-Chan
    Peng, Shin-Lei
    Lee, Shan-Chih
    Hsieh, Ling-Ling
    Shih, Cheng-Ting
    RADIATION PHYSICS AND CHEMISTRY, 2020, 168
  • [24] Spatial pattern summation is phase-insensitive in the fovea but not in the periphery
    Chen, CC
    Tyler, CW
    SPATIAL VISION, 1999, 12 (03): : 267 - 285
  • [25] Sparsity-based method for ring artifact elimination in computed tomography
    Selim, Mona
    Rashed, Essam A.
    Atiea, Mohammed A.
    Kudo, Hiroyuki
    PLOS ONE, 2022, 17 (06):
  • [26] Iterative Ring Artifact Removal Method for Helical Computed Tomography Scans
    Balogh, Zsolt Adam
    Huszar, Tamas
    Kis, Benedek Janos
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2020, 44 (05) : 796 - 805
  • [27] Probabilistic phase-insensitive optical squeezer in compliance with causality
    Gagatsos, C. N.
    Karpov, E.
    Cerf, N. J.
    PHYSICAL REVIEW A, 2012, 86 (01):
  • [28] Correction for the 'Black Hole' Artifact in Cone-Beam Computed Tomography
    Sun, M.
    Nagy, T.
    Oelhafen, M.
    Star-Lack, J.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [29] Simulation-based artifact correction (SBAC) for metrological computed tomography
    Maier, Joscha
    Leinweber, Carsten
    Sawall, Stefan
    Stoschus, Henning
    Ballach, Frederic
    Mueller, Tobias
    Hammer, Michael
    Christoph, Ralf
    Kachelriess, Marc
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2017, 28 (06)
  • [30] MAGNETIC FIELD INDEPENDENT PHASE-INSENSITIVE DETECTOR OF HYPERSONICS
    WIGMORE, JK
    VONGUTFE.RJ
    IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1969, SU16 (01): : 26 - &