Ring Artifact Correction for Phase-Insensitive Ultrasound Computed Tomography

被引:4
|
作者
Baker, Christian [1 ,2 ]
Sarno, Daniel [2 ]
Eckersley, Robert J. [1 ]
Zeqiri, Bajram [2 ]
机构
[1] Kings Coll London, Biomed Engn Dept, Div Imaging Sci, London SE1 7EH, England
[2] Natl Phys Lab, Ultrasound & Underwater Acoust Grp, Teddington TW11 0LW, Middx, England
关键词
Transducers; Sensitivity; Ultrasonic imaging; Acoustics; Breast; Attenuation; Ultrasonic variables measurement; Artifact correction; computed tomography (CT); medical diagnostic imaging; ultrasonic imaging; SOUND SPEED; FLAT-FIELDS; RECONSTRUCTION; NORMALIZATION; ATTENUATION; REMOVAL;
D O I
10.1109/TUFFC.2019.2948429
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
An algorithm was developed for the correction of ring artifacts in phase-insensitive ultrasound computed tomography attenuation images. Differences in the measurement sensitivity between the ultrasound transducer array elements cause discontinuities in the sinogram which manifest as rings and arcs in the reconstructed image. The magnitudes of the discontinuities are potentially time-varying and dependent on the attenuation being measured. The algorithm dynamically determines the measurement sensitivity of each transducer in the array during the scan by comparison with both the elements to its left and the elements to its right. Elements at either end of the array are corrected, assuming a zero-attenuation path. The two estimates of sensitivity are combined using a weighted mean similar to a Kalman filter. The algorithm was tested on simulated and experimentally acquired data. It was demonstrated to reduce the root-mean-square error (RMSE) of simulated images against ground-truth images by up to a factor of 50 compared with uncorrected images and to visibly reduce artifacts on images reconstructed from the experimentally acquired data.
引用
收藏
页码:513 / 525
页数:13
相关论文
共 50 条
  • [1] Pulse Pileup Correction of Signals From a Pyroelectric Sensor for Phase-Insensitive Ultrasound Computed Tomography
    Baker, Christian
    Sarno, Daniel
    Eckersley, Robert J.
    Zeqiri, Bajram
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2019, 68 (10) : 3920 - 3931
  • [2] Phase-Insensitive Ultrasound Computed Tomography for Acoustic Attenuation Imaging of Breast Phantoms
    Sarno, D.
    Baker, C.
    Hodnett, M.
    Zeqiri, B.
    MEDICAL IMAGING 2021: ULTRASONIC IMAGING AND TOMOGRAPHY, 2021, 11602
  • [3] Phase-Insensitive Ultrasound Tomography of the Attenuation of Breast Phantoms
    Baker, Christian
    Sarno, Daniel
    Eckersley, Robert
    Hodnett, Mark
    Zeqiri, Bajram
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1219 - 1222
  • [4] Quantitative ultrasonic computed tomography using phase-insensitive pyroelectric detectors
    Zeqiri, Bajram
    Baker, Christian
    Alosa, Giuseppe
    Wells, Peter N. T.
    Liang, Hai-Dong
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (15): : 5237 - 5268
  • [5] Phase-insensitive versus phase-sensitive ultrasound absorption tomography in the frequency domain
    Kaupinmaki, Santeri
    Cox, Ben
    Arridge, Simon
    APPLIED MATHEMATICS IN SCIENCE AND ENGINEERING, 2023, 31 (01):
  • [6] Robustness of homodyne tomography to phase-insensitive noise
    DAriano, GM
    Sterpi, N
    JOURNAL OF MODERN OPTICS, 1997, 44 (11-12) : 2227 - 2232
  • [7] Ring artifact correction for X-ray computed tomography
    Ma, Jiming
    Song, Yan
    Wang, Qunshu
    Song, Guzhou
    Zhang, Jianqi
    Xia, Jingtao
    Han, Changcai
    Duan, Baojun
    Du, Jiye
    Zhou, Ming
    Qiangjiguang Yu Lizishu/High Power Laser and Particle Beams, 2014, 26 (12):
  • [8] Feature-Sensitive Ring Artifact Correction for Computed Tomography
    Markus Wedekind
    Susana Castillo
    Marcus Magnor
    Journal of Nondestructive Evaluation, 2023, 42
  • [9] Feature-Sensitive Ring Artifact Correction for Computed Tomography
    Wedekind, Markus
    Castillo, Susana
    Magnor, Marcus
    JOURNAL OF NONDESTRUCTIVE EVALUATION, 2023, 42 (01)
  • [10] PHASE-INSENSITIVE DETECTION FOR MEASUREMENT OF BACKSCATTERED ULTRASOUND
    JOHNSTON, PH
    MILLER, JG
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1986, 33 (06) : 713 - 721