Short communication: Kilovoltage measurement with rhodium target and filters on mammography X-ray machines

被引:3
|
作者
Underwood, AC
Law, J
Goodman, DA
Robinson, A
Rust, A
机构
[1] UNIV EDINBURGH,WESTERN GEN HOSP,DEPT MED PHYS & MED ENGN,EDINBURGH EH4 2XU,MIDLOTHIAN,SCOTLAND
[2] ADDENBROOKES HOSP,REG RADIAT PROTECT SERV,CAMBRIDGE CB3 7QJ,ENGLAND
[3] ST LUKES HOSP,REG RADIAT PROTECT SERV,GUILDFORD GU1 3NT,SURREY,ENGLAND
来源
BRITISH JOURNAL OF RADIOLOGY | 1996年 / 69卷 / 824期
关键词
D O I
10.1259/0007-1285-69-824-769
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
X-ray machines used for mammography have, until recently, almost exclusively employed molybdenum (Mo) as anode material and filtration. In the UK, the RMI model 232 kVp meter is used extensively for the measurement of kilovoltage on such equipment. This unit is provided with switchable internal calibration only for molybdenum and molybdenum, or tungsten and aluminium, as anode and filtration, respectively. However, rhodium (Rh) has recently been introduced for filtration with either a molybdenum or rhodium anode in mammography equipment but, as yet, calibration facilities are not available for rhodium spectra. In this work, appropriate corrections for readings obtained with the RMI 232 kVp meter are derived for use with rhodium as filtration material with either molybdenum or rhodium anode material. An intercomparison between measurements made with four RMI model 232 kVp meters and nine IGE DMR X-ray sets was undertaken. The reproducibility of the instruments was confirmed and measurements of tube potential made on each of the X-ray sets pooled. Measurements were made from a nominal 25-35 kVp using Mo/Mo, Mo/Rh and Rh/Rh target/filter combinations. Corrections for readings obtained with Mo/Rh and Rh/Rh were produced by comparison with readings obtained with Mo/Mo, assuming stability of tube potential between anodes. The results are compared with data recently produced by the manufacturer of the meter.
引用
收藏
页码:769 / 773
页数:5
相关论文
共 50 条
  • [21] Study of dosimetry consistency for kilovoltage x-ray beams
    Ma, CM
    Li, XA
    Seuntjens, JP
    MEDICAL PHYSICS, 1998, 25 (12) : 2376 - 2384
  • [22] X-RAY MAMMOGRAPHY AND BREAST COMPRESSION
    COUPAR, J
    LANCET, 1992, 340 (8815): : 366 - 366
  • [23] X-RAY MAMMOGRAPHY AND BREAST COMPRESSION
    WATMOUGH, DJ
    QUAN, KM
    LANCET, 1992, 340 (8811): : 122 - 122
  • [24] HISTORY AND STATUS OF X-RAY MAMMOGRAPHY
    HENDEE, WR
    HEALTH PHYSICS, 1995, 69 (05): : 636 - 648
  • [25] X-RAY BEAM FILTRATION FOR MAMMOGRAPHY
    SIEDBAND, MP
    JENNINGS, RJ
    EASTGATE, RJ
    ERGUN, DL
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1978, 130 (02) : 384 - 384
  • [26] OPTIMIZATION OF X-RAY MAMMOGRAPHY CONDITIONS
    BLINOV, NN
    KOZLOVSKY, EB
    LEONOV, BI
    MEDICAL PROGRESS THROUGH TECHNOLOGY, 1992, 18 (1-2) : 69 - 72
  • [27] X-RAY MAMMOGRAPHY AND BREAST COMPRESSION
    FAULKNER, K
    ROBSON, KJ
    KOTRE, CJ
    LANCET, 1992, 340 (8822): : 797 - 798
  • [28] A cermet X-ray tube for mammography
    Valuev N.N.
    Kulikov N.A.
    Kochetkov M.D.
    Taubin M.L.
    Biomedical Engineering, 2005, 39 (5) : 251 - 254
  • [29] X-RAY MAMMOGRAPHY WITH SYNCHROTRON RADIATION
    BURATTINI, E
    GAMBACCINI, M
    MARZIANI, M
    RIMONDI, O
    INDOVINA, PL
    POCEK, M
    SIMONETTI, G
    BENASSI, M
    TIRELLI, C
    PASSARIELLO, R
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (01): : 638 - 640
  • [30] USE OF X-RAY SCREENS IN MAMMOGRAPHY
    STEVELS, ALN
    PINGAULT, F
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1977, 124 (09) : 1400 - 1404