3D catheter reconstruction in HDR prostate brachytherapy for pre-treatment verification using a flat panel detector

被引:9
|
作者
Smith, Ryan L. [1 ,2 ]
Haworth, Annette [2 ,3 ]
Panettieri, Vanessa [1 ]
Millar, Jeremy L. [1 ,2 ,4 ]
Franich, Rick D. [2 ]
机构
[1] Alfred Hosp, Alfred Hlth Radiat Oncol, Melbourne, Vic 3004, Australia
[2] RMIT Univ, Sch Sci, Melbourne, Vic 3000, Australia
[3] Univ Sydney, Sch Phys, Camperdown, NSW 2006, Australia
[4] Monash Univ, Cent Clin Sch, Melbourne, Vic 3004, Australia
关键词
HDR brachytherapy; Pre-treatment imaging; Pre-treatment verification; Prostate; Flat panel detector; RATE AFTERLOADING BRACHYTHERAPY; DOSE-RATE BRACHYTHERAPY; ISOCENTRIC C-ARM; SEED RECONSTRUCTION; CANCER; MOVEMENT; DISPLACEMENT; GUIDELINES; FRACTIONS; ALGORITHM;
D O I
10.1016/j.ejmp.2017.06.008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: High dose rate prostate brachytherapy is a widely-practiced treatment, delivering large conformal doses in relatively few treatment fractions. Inter-and intra-fraction catheter displacements have been reported. Unrecognized displacement can have a significant impact on dosimetry. Knowledge of the implant geometry at the time of treatment is important for ensuring safe and effective treatment. In this work we demonstrate a method to reconstruct the catheter positions pre-treatment, using a 'shift' imaging technique, and perform registration with the treatment plan for verification relative to the prostate. Methods: Two oblique 'shift' images were acquired of a phantom containing brachytherapy catheters, representing the patient immediately pre-treatment. Using a back projection approach, the catheter paths were reconstructed in 3D and registered with the planned catheter paths. The robustness of the reconstruction and registration process was investigated as a function of phantom rotation. Catheter displacement detection was performed and compared to known applied displacements. Results: Reconstruction of the implant geometry in 3D immediately prior to treatment was achieved. A mean reconstruction uncertainty of 0.8 mm was determined for all catheters with a mean registration uncertainty of 0.5 mm. A catheter displacement detection threshold of 2.2 mm was demonstrated. Catheter displacements were all detected to within 0.5 mm of the applied displacements. Conclusion: This technique is robust and sensitive to assess catheter displacements throughout the implant volume. This approach provides a method to detect, in 3D, changes in catheter positions relative to the prostate. The method has sufficient sensitivity to enable clinically significant decisions immediately prior to treatment delivery. (C) 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
下载
收藏
页码:121 / 131
页数:11
相关论文
共 50 条
  • [31] 3D Reconstruction using HDR-based SLAM
    Yeh, Chia-Hung
    Lin, Min-Hui
    Lu, Wei-Chieh
    2019 ASIA-PACIFIC SIGNAL AND INFORMATION PROCESSING ASSOCIATION ANNUAL SUMMIT AND CONFERENCE (APSIPA ASC), 2019, : 1976 - 1980
  • [32] Early toxicity outcomes: A single 15Gy fraction HDR brachytherapy as pre-treatment EBRT boost in prostate cancer
    Chicas Sett, R.
    Soler, A.
    Fernandez, J.
    Burgos, J.
    Pons, O.
    Roldan, S.
    Celada, F.
    Gimeno, J.
    Tormo, A.
    Perez-Calatayud, J.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : V - V
  • [33] Early toxicity outcomes: A single 15Gy fraction HDR brachytherapy as pre-treatment EBRT boost in prostate cancer
    Chicas Sett, R.
    Soler, A.
    Fernandez, J.
    Burgos, J.
    Pons, O.
    Roldan, S.
    Celada, F.
    Gimeno, J.
    Tormo, A.
    Perez-Calatayud, J.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S357 - S357
  • [34] Real-Time Catheter Tracking for HDR Prostate Implant Using an Electromagnetic 3D Guidance Device
    Zhou, J.
    Sebastian, E.
    Ghilezan, M.
    Martinez, A.
    Yan, D.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [35] Clinical Outcomes of 3D CRT with HDR brachytherapy in patients with localized Prostate cancer.
    Rafi, M.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S848 - S849
  • [36] Dosimetric evaluation of a 3D printed phantom for patient-specific pre-treatment plan verification
    Makris, D. N.
    Zoros, E.
    Boursianis, T.
    Pappas, E.
    Maris, T. G.
    Efstathopoulos, E. P.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S39 - S40
  • [37] Evaluation of 3D pre-treatment verification for volumetric modulated arc therapy plan in head region
    Ruangchan, S.
    Oonsiri, S.
    Suriyapee, S.
    13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [38] 3D dose verification in 192Ir HDR prostate monotherapy using polymer gels and MRI
    Kipouros, P
    Papagiannis, P
    Sakelliou, L
    Karaiskos, P
    Sandilos, P
    Baras, P
    Seimenis, I
    Kozicki, M
    Anagnostopoulos, G
    Baltas, D
    MEDICAL PHYSICS, 2003, 30 (08) : 2031 - 2039
  • [39] Toward 3D dose verification of an electronic brachytherapy source with a plastic scintillation detector
    Georgi, Peter
    Kertzscher, Gustavo
    Nyvang, Lars
    Solc, Jaroslav
    Schneider, Thorsten
    Tanderup, Kari
    Johansen, Jacob Graversen
    MEDICAL PHYSICS, 2022, 49 (05) : 3432 - 3443
  • [40] Technical feasibility of CT assisted 3D HDR brachytherapy in the treatment of skin cancers
    Lakosi, F.
    Kovacs, A.
    Cselik, Z.
    Antal, G.
    Kotek, G.
    Battyani, Z.
    Hadjiev, J.
    Bogner, P.
    Repa, I.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S482 - S482