Neutral carrier based needle-type sensor for the determination of calcium ion

被引:1
|
作者
Li, XL [1 ]
Hong, DQ [1 ]
Li, YY [1 ]
Lin, ZH [1 ]
Ren, S [1 ]
机构
[1] Huazhong Univ Sci & Technol, Tongji Med Coll, Dept Chem, Wuhan 340030, Peoples R China
关键词
calcium ion; needle-type; neutral carrier; in vivo monitoring;
D O I
10.1117/12.440166
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Calcium ion is a very important messenger not only between cell and cell, but also in meridian main and collateral channels. A new neutral carrier PVC membrane based needle-type potential sensor was prepared, which shows a good selectivity for Ca2+. Responses characteristics, pH, response time of sensor, other metal ions such as Mg2+, Na+, and K+ in measuring calcium ion were investigated. The resulting sensor has a linear range from 1.0 x 10(-6) to 9.1 x 10(-3) mol/L. The sensor will be used in the detection of Ca2+ in vivo in muscular tissues of the mice.
引用
收藏
页码:115 / 117
页数:3
相关论文
共 50 条
  • [41] A needle-type optical enzyme sensor system for determining glucose levels in fish blood
    Endo, Hideaki
    Yonemori, Yuki
    Musiya, Kazuya
    Maita, Masashi
    Shibuya, Toru
    Ren, Huifeng
    Hayashi, Tetsuhito
    Mitsubayashi, Kohji
    ANALYTICA CHIMICA ACTA, 2006, 573 : 117 - 124
  • [42] Two Decades' advancements and Research trends in needle-type Sensor technology: A scientometric analysis
    Yang, Guangyi
    Xu, Wenjing
    Xu, Botan
    Yang, Yi
    Li, Pengwei
    Yu, Aotian
    Ning, Simin
    Fu, Qixuan
    Zhang, Rong
    Liu, Xiaohan
    HELIYON, 2024, 10 (06)
  • [43] Detection of plant water content with needle-type in-situ water content sensor
    Katayanagi H.
    Miki N.
    IEEJ Transactions on Sensors and Micromachines, 2010, 130 (09) : 452 - 453
  • [44] A cobalt-coated needle-type microelectrode array sensor for in situ monitoring of phosphate
    Lee, Jin-Hwan
    Lee, Woo Hyoung
    Bishop, Paul L.
    Papautsky, Ian
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2009, 19 (02)
  • [45] Needle-type sensor array for simultaneous measurement of dissolved oxygen and redox potential in situ
    Papautsky, Ian
    Lee, Jin-Hwan
    Seo, Youngwoo
    Lim, Tae-Sun
    Bishop, Paul L.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2007, 234
  • [46] PRODUCTION OF NEEDLE-TYPE LIQUID-METAL ION SOURCES AND THEIR APPLICATION IN A SCANNING ION-MICROSCOPE
    KNAPP, H
    RUBESAME, D
    NIEDRIG, H
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1991, 59 : 131 - 134
  • [47] THE EMISSION CHARACTERISTICS OF AN INDIUM NEEDLE-TYPE LIQUID-METAL ION-SOURCE
    HORNSEY, RI
    APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1989, 49 (03): : 293 - 297
  • [48] DEVELOPMENT OF A NEEDLE-TYPE GLUCOSE SENSOR-BASED ON A TITANIUM-DIOXIDE OXYGEN-ELECTRODE FOR THE ARTIFICIAL PANCREAS
    IKEDA, S
    HARA, M
    HISHIDA, Y
    KIMURA, M
    ITO, K
    SENSORS AND ACTUATORS B-CHEMICAL, 1993, 13 (1-3) : 315 - 318
  • [49] FEASIBILITY OF THE WEARABLE ARTIFICIAL ENDOCRINE PANCREAS WITH A NEEDLE-TYPE GLUCOSE SENSOR IN CLINICAL-APPLICATION
    SHICHIRI, M
    KAWAMORI, R
    YAMASAKI, Y
    HAKUI, N
    ABE, H
    DIABETES, 1983, 32 : A36 - A36
  • [50] EVALUATING INVITRO AND INVIVO, THE INTERFERENCE OF ASCORBATE AND ACETAMINOPHEN ON GLUCOSE DETECTION BY A NEEDLE-TYPE GLUCOSE SENSOR
    MOATTISIRAT, D
    VELHO, G
    REACH, G
    BIOSENSORS & BIOELECTRONICS, 1992, 7 (05): : 345 - 352