Development and Validation of Hematocrit Level Measurement in Dried Blood Spots Using Near-Infrared Spectroscopy

被引:18
|
作者
van de Velde, Daan [1 ]
van der Graaf, Jordy L. [1 ]
Boussaidi, Mariam [1 ]
Huisman, Ruud [1 ]
Hesselink, Dennis A. [2 ]
Russcher, Henk [3 ]
Kooij-Egas, Annelies C. [4 ]
van Maarseveen, Erik [4 ]
de Winter, Brenda C. M. [1 ]
机构
[1] Erasmus MC, Dept Hosp Pharm, Wytemaweg 80,POB 2040, NL-3015 CN Rotterdam, Netherlands
[2] Erasmus MC, Div Nephrol & Transplantat, Dept Internal Med, Rotterdam, Netherlands
[3] Erasmus MC, Dept Clin Chem, Univ Med Ctr Rotterdam, Rotterdam, Netherlands
[4] Univ Med Ctr Utrecht, Dept Hosp Pharm, Utrecht, Netherlands
关键词
PLS; spectroscopy; near-infrared; hematocrit; dried blood spot; PREDICTION; TACROLIMUS; LIPEMIA; HEMOLYSIS; ICTERUS;
D O I
10.1097/FTD.0000000000000834
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Dried blood spots (DBSs) have gained recent popularity as a sampling method for therapeutic drug monitoring. For patients, DBS sampling has several advantages over venous blood sampling. However, technical issues primarily influenced by hematocrit levels, interfere with the implementation of this method in daily clinical practice. The results of concentration measurements of drugs that are influenced by hematocrit should be corrected for hematocrit levels. In this article, we developed a fast, nondestructive, near-infrared (NIR)-based method for measuring the hematocrit in DBSs. Method: Using a partial least squares algorithm, an NIR-based quantification method was developed for measuring hematocrit levels of 0.19-0.49 L/L. Residual venous blood of 522 patients was used to build this partial least squares model. The validity of the method was evaluated using 40 patient samples. DBSs were created by adding a small amount (50 mu L) of blood on a Whatman filter paper and drying for 24 hours in a desiccator cabinet. The robustness was evaluated by measuring 24 additional samples with a high hemolysis, icterus, and lipemia (HIL) index. The hematocrit values obtained using a Sysmex XN hemocytometry analyzer were used as reference. Results: The difference between hematocrit measurements obtained with NIR spectroscopy and a hemocytometry analyzer was <15% for the 40 samples. The accuracy (<= 9%) and precision (<= 7%) for all the quality control samples were within the acceptance criteria of <15%. The intraassay and interassay coefficient of variability was <= 3% and <= 6%, respectively, for the different quality control levels. There were no deviations in the measurements for the samples with high HIL indices. The stability of hematocrit in DBS was up to 14 days for all levels. Conclusions: We developed and validated a hematocrit model using NIR spectroscopy. This nondestructive, accurate, and reproducible method has a short analysis time (51 seconds), and can be used to analyze DBS samples stored for up to 2 weeks in a desiccator cabinet.
引用
收藏
页码:351 / 357
页数:7
相关论文
共 50 条
  • [1] Measurement of Hematocrit in Dried Blood Spots Using Near-Infrared Spectroscopy: Robust, Fast, and Nondestructive
    Oostendorp, Marlies
    El Amrani, Mohsin
    Diemel, Eric C.
    Hekman, Dennis
    van Maarseveen, Erik M.
    CLINICAL CHEMISTRY, 2016, 62 (11) : 1534 - 1536
  • [2] Near-infrared-based hematocrit prediction of dried blood spots: An in-depth evaluation
    Delahaye, Lisa
    Heughebaert, Liesl
    Luhr, Christoph
    Lambrecht, Stijn
    Stove, Christophe P.
    CLINICA CHIMICA ACTA, 2021, 523 : 239 - 246
  • [3] DEVELOPMENT AND VALIDATION OF A RELATIVE BLOOD FLOW CHANGE MEASUREMENT ALGORITHM USING NEAR-INFRARED LIGHT
    Park, Jieun
    Kim, Junghun
    Choi, Sang-il
    Lee, Jongmin
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2024, 24 (08)
  • [4] Near-infrared spectroscopy for the measurement of cerebral blood flow
    Luo, QM
    Deng, H
    Gong, H
    Zeng, SQ
    Chen, WG
    JOURNAL OF INFRARED AND MILLIMETER WAVES, 1999, 18 (02) : 138 - 144
  • [5] Measurement of pH in whole blood by near-infrared spectroscopy
    Alam, MK
    Rohrscheib, MR
    Franke, JE
    Niemczyk, TM
    Maynard, JD
    Robinson, MR
    APPLIED SPECTROSCOPY, 1999, 53 (03) : 316 - 324
  • [6] Near-infrared spectroscopy for the measurement of cerebral blood flow
    Huazhong Univ of Science and, Technology, Wuhan, China
    Hongwai Yu Haomibo Xuebao, 2 (138-144):
  • [7] MEASUREMENT OF HEMOGLOBIN IN UNLYSED BLOOD BY NEAR-INFRARED SPECTROSCOPY
    KUENSTNER, JT
    NORRIS, KH
    MCCARTHY, WF
    APPLIED SPECTROSCOPY, 1994, 48 (04) : 484 - 488
  • [8] Fully Automated Optical Hematocrit Measurement from Dried Blood Spots
    Luginbuhl, Marc
    Fischer, Yannick
    Gaugler, Stefan
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2022, 46 (02) : 187 - 193
  • [9] Comparison of near-infrared and UV-vis-based non-contact hematocrit prediction of dried blood spots from patients on immunosuppressants
    Deprez, Sigrid
    Heughebaert, Liesl
    Boffel, Laura
    Stove, Christophe P. P.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2023, 61 (06) : E87 - E90
  • [10] Cerebrovascular reactivity measurement using near-infrared spectroscopy
    Grosset, D
    CLINICAL SCIENCE, 1998, 95 (04) : 395 - 395