Contributions of remote monitoring to the follow-up of implantable cardioverter-defibrillator leads under advisory

被引:51
|
作者
Guedon-Moreau, Laurence [1 ]
Chevalier, Philippe [2 ]
Marquie, Christelle [1 ]
Kouakam, Claude [1 ]
Klug, Didier [1 ]
Lacroix, Dominique [1 ]
Brigadeau, Francois [1 ]
Kacet, Salem [1 ]
机构
[1] Ctr Hosp Reg Univ, Hop Cardiol, F-59037 Lille, France
[2] Ctr Hosp Univ, Lyon, France
关键词
Telemedicine; Remote monitoring; Lead failure; Implantable cardioverter-defibrillator; Inappropriate shock; RESYNCHRONIZATION THERAPY SYSTEMS; REDUCE INAPPROPRIATE SHOCKS; FAILURE; HOME; PACEMAKERS; EXPERIENCE; FRACTURE; RECALLS; ALERTS;
D O I
10.1093/eurheartj/ehq203
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Automatic daily transmission of data from implantable cardioverter-defibrillators (ICDs) enables the remote monitoring of device status and leads function. We report on a 2-year experience with remote monitoring in 40 recipients of high-voltage ICD leads, prone to fracture and under advisory since October 2007. The ICDs were remotely monitored as well as systematically interrogated in the ambulatory department every 3 months. The patients were also seen in case of abnormal lead impedance, or other manifestations consistent with lead dysfunction. Over a mean follow-up of 22 +/- 4 months after ICD implantation, four lead dysfunctions were suspected because of remotely transmitted oversensing of noise artifacts, abrupt rise in pacing impedance, or both. A lead fracture needing lead replacement was confirmed in three patients (7.5%), two of them before any inappropriate therapy and one after the delivery of three inappropriate shocks. No lead failure was observed in the remaining 36 patients, either at the time of ambulatory visits or during remote monitoring. Remote monitoring allowed the early and reliable detection of ICD leads failure without requiring any patient intervention.
引用
收藏
页码:2246 / 2252
页数:7
相关论文
共 50 条
  • [1] Individualization of follow-up in implantable cardioverter-defibrillator patients with the help of automatic remote implantable cardioverter-defibrillator monitoring
    Neuser, H
    Schumacher, B
    Kottkamp, H
    Schmidinger, H
    Schuchert, A
    Brugada, P
    [J]. EUROPEAN HEART JOURNAL, 2004, 25 : 166 - 166
  • [2] Psychological Effects of Implantable Cardioverter-Defibrillator Leads under Advisory
    Heatherly, Steven J.
    Simmons, Tony
    Fitzgerald, David M.
    Mitchell, Mark
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2011, 34 (06): : 694 - 699
  • [3] Outcome of advisory implantable cardioverter-defibrillator replacement: One-year follow-up
    Gould, Paul A.
    Gula, Lorne J.
    Champagne, Jean
    Healyey, Jeffrey S.
    Cameron, Doug
    Simpson, Christophers
    Thibault, Bernard
    Pinter, Arnold
    Tung, Stanley
    Sterns, Laurence
    Birnie, David
    Exner, Derek
    Parkash, Ratika
    Skanes, Allan C.
    Yee, Raymond
    Klein, George J.
    Krahn, Andrew D.
    [J]. HEART RHYTHM, 2008, 5 (12) : 1675 - 1681
  • [4] Clinical follow-up of patients with implantable cardioverter-defibrillator
    da Fonseca, Silvia Martelo Souza
    Belo, Luiz Gustavo
    Carvalho, Hecio
    Araujo, Nilson
    Munhoz, Claudio
    Siqueira, Leonardo
    Maciel, Washington
    Andrea, Eduardo
    Atie, Jacob
    [J]. ARQUIVOS BRASILEIROS DE CARDIOLOGIA, 2007, 88 (01) : 8 - 16
  • [5] Longitudinal Follow-Up of Implantable Cardioverter Defibrillator Leads
    Liu, Jeffrey
    Brumberg, Genevieve
    Rattan, Rohit
    Patel, Divyang
    Adelstein, Evan
    Jain, Sandeep
    Saba, Samir
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2014, 113 (01): : 103 - 106
  • [6] Transvenous extraction of implantable cardioverter-defibrillator leads under advisory-A comparison of Riata, Sprint Fidelis, and non-recalled implantable cardioverter-defibrillator leads
    Brunner, Michael P.
    Cronin, Edmond M.
    Jacob, Jessen
    Duarte, Valeria E.
    Tarakji, Khaldoun G.
    Martin, David O.
    Callahan, Thomas
    Borek, P. Peter
    Cantillon, Daniel J.
    Niebauer, Mark J.
    Saliba, Walid I.
    Kanj, Mohamed
    Wazni, Oussama
    Baranowski, Bryan
    Wilkoff, Bruce L.
    [J]. HEART RHYTHM, 2013, 10 (10) : 1444 - 1450
  • [7] FOLLOW-UP OF 51 PATIENTS WITH THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    TROUP, PJ
    DUQUETTE, SE
    CHAPMAN, PD
    [J]. CLINICAL RESEARCH, 1986, 34 (04): : A903 - A903
  • [8] The Implantable Cardioverter-Defibrillator Minimalist An Approach to Patient Follow-Up and Management of Implantable Defibrillators
    Sweeney, Michael O.
    [J]. CIRCULATION, 2012, 126 (03) : 369 - 377
  • [9] Sophisticated remote implantable cardioverter defibrillator follow-up: A status report
    Schoenfeld, MH
    Reynolds, DW
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (03): : 235 - 240
  • [10] High failure rate for an epicardial implantable cardioverter-defibrillator lead: Implications for long-term follow-up of patients with an implantable cardioverter-defibrillator
    Brady, PA
    Friedman, PA
    Trusty, JM
    Grice, S
    Hammill, SC
    Stanton, MS
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (03) : 616 - 622