In vitro biomechanical comparison of pedicle screws, sublaminar hooks, and sublaminar cables

被引:46
|
作者
Hitchon, PW
Brenton, MD
Black, AG
From, A
Harrod, JS
Barry, C
Serhan, H
Torner, JC
机构
[1] Univ Iowa, Roy J & Lucille A Carver Coll Med, Coll Publ Hlth, Dept Neurosurg, Iowa City, IA USA
[2] Univ Iowa, Roy J & Lucille A Carver Coll Med, Coll Publ Hlth, Dept Biomed Engn, Iowa City, IA USA
[3] Univ Iowa, Roy J & Lucille A Carver Coll Med, Coll Publ Hlth, Dept Epidemiol, Iowa City, IA USA
[4] Vet Adm Med Ctr, Iowa City, IA USA
关键词
pedicle screw; sublaminar hook; sublaminar cable; bone mineral density; biomechanical study;
D O I
10.3171/spi.2003.99.1.0104
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Object. Three types of posterior thoracolumbar implants are in use today: pedicle screws, sublaminar titanium cables, and sublaminar hooks. The authors conducted a biomechanical comparison of these three implants in human cadaveric spines. Methods. Spine specimens (T5-12) were harvested, radiographically assessed for fractures or metastases, and their bone mineral density (BMD) was measured. Individual vertebrae were disarticulated and fitted with either pedicle screws, sublaminar cables, or bilateral claw hooks. The longitudinal component of each construct consisted of bilateral 10-cm rods connected with two cross-connectors. The vertebral body was embedded in cement, and the rods were affixed to a ball-and-socket apparatus for the application of a distraction force. The authors analyzed 1) 20 vertebrae implanted with screws; 2) 20 with hooks, and 3) 20 with cables. The maximum pullout (MPO) forces prior to failure (mean +/- standard deviation) for the screw, hook, and cable implants were 972 +/- 330, 802 +/- 356, and 654 +/- 248 N, respectively (p = 0.0375). Cables allowed significantly greater displacement (6.80 +/- 3.95 mm) prior to reaching the MPO force than hooks (3.73 +/- 1.42 mm) and screws (4.42 +/- 2.15 mm [p = 0.0108]). Eleven screw-implanted vertebrae failed because of screw pullout. All hook-and-cable-implanted vertebrae failed because of pedicle, middle column, or laminar fracture. Conclusions. These findings suggest that screws possess the greatest pullout strength of the three fixation systems. Sublaminar cables are the least rigid of the three. When screw failure occurred, the mechanism was generally screw backout, without vertebral fractures.
引用
收藏
页码:104 / 109
页数:6
相关论文
共 50 条
  • [41] The effectiveness of short hybrid stabilization with sublaminar bands and transpedicular screws in the treatment of thoracolumbar spine fractures
    Brembilla, Carlo
    Fanti, Andrea
    Rampini, Angela D.
    Dorelli, Gianluigi
    Sicignano, Angelo M.
    Cracchiolo, Giorgio
    Bernucci, Claudio
    JOURNAL OF NEUROSURGICAL SCIENCES, 2024, 68 (04) : 412 - 421
  • [42] Comparison of 2 Kinds of Pedicle Screws in Primary Spinal Instrumentation Biomechanical and Interfacial Evaluations in Sheep Vertebrae In Vitro
    Liu, Da
    Zhang, Yi
    Lei, Wei
    Wang, Cai-ru
    Xie, Qing-yun
    Liao, Dong-fa
    Jiang, Kai
    Zhou, Jin-song
    Zhang, Bo
    Pan, Xian-ming
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2014, 27 (02): : E72 - E80
  • [43] A biomechanical study of 3 different types of sublaminar wire used for constructs in the thoracic spine
    Murakami, Hideki
    Yamazaki, Ken
    Attallah-Wasif, Emad S.
    Tsai, Kai-Jow
    Shimamura, Tadashi
    Hutton, William C.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (06): : 442 - 446
  • [44] Pull out Strength of Dual Outer Diameter Pedicle Screws Compared to Uncemented and Cemented Standard Pedicle Screws: A Biomechanical in vitro Study
    Lorenz, Andrea
    Leichtle, Carmen I.
    Frantz, Sandra
    Bumann, Marte
    Tsiflikas, Ilias
    Shiozawa, Thomas
    Leichtle, Ulf G.
    ORTHOPAEDIC SURGERY, 2017, 9 (02) : 229 - 236
  • [45] Radiopaque UHMWPE sublaminar cables for spinal deformity correction: Preclinical mechanical and radiopacifier leaching assessment
    Roth, Alex K.
    Boon-Ceelen, Karlien
    Smelt, Harold
    van Rietbergen, Bert
    Willems, Paul C.
    van Rhijn, Lodewijk W.
    Arts, Jacobus J.
    JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2018, 106 (02) : 771 - 779
  • [46] Biomechanical comparison of pullout strengths of six pedicle screws with different thread designs
    Karakasli, Ahmet
    Acar, Nihat
    Husemoglu, R. Bugra
    JOINT DISEASES AND RELATED SURGERY, 2021, 32 (01): : 192 - 197
  • [47] Biomechanical comparison of alternative densities of pedicle screws for the treatment of adolescent idiopathic scoliosis
    Wang, Xiaoyu
    Aubin, Carl-Eric
    Robitaille, Isabelle
    Labelle, Hubert
    EUROPEAN SPINE JOURNAL, 2012, 21 (06) : 1082 - 1090
  • [48] Biomechanical comparison of alternative densities of pedicle screws for the treatment of adolescent idiopathic scoliosis
    Xiaoyu Wang
    Carl-Eric Aubin
    Isabelle Robitaille
    Hubert Labelle
    European Spine Journal, 2012, 21 : 1082 - 1090
  • [49] Biomechanical Comparison of Multilevel Stand-Alone Lumbar Lateral Interbody Fusion With Posterior Pedicle Screws: An In Vitro Study
    Mok, James M.
    Lin, Ye
    Tafur, Julio Castillo
    Diaz, Roberto Leonardo
    Amirouche, Farid
    NEUROSPINE, 2023, 20 (02) : 478 - 486
  • [50] Repair of the defect in spondylolysis - Durable fixation with pedicle screws and laminar hooks
    Kakiuchi, M
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (06): : 818 - 825