Determination of aldosterone in serum by liquid chromatography-tandem mass spectrometry

被引:69
|
作者
Turpeinen, Ursula [1 ]
Hamalainen, Esa [1 ,2 ]
Stenman, Ulf-Hakan [2 ]
机构
[1] HUSLAB, Lab Womens Clin, Helsinki 00029, Finland
[2] Univ Helsinki, Dept Clin Chem, SF-00100 Helsinki, Finland
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2008年 / 862卷 / 1-2期
关键词
serum aldosterone; LC-MS/MS; ESI; Electrospray ionization mass spectrometry; API; 3000; RIA;
D O I
10.1016/j.jchromb.2007.11.005
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Measurement of serum aldosterone is clinically important in the diagnosis of hypertension.)While isotope dilution gas chromatography-mass spectrometry (ID-GC-MS) provides reliable results, it requires derivatization and is lengthy and time-consuming. Detection by liquid chromatography-mass spectrometry (LC-MS) is a potentially superior method. The analysis utilizes 0.5 mL of serum. The samples were extracted with dichloromethane-ether. The extract was evaporated to dryness and aldosterone was analyzed by LC-MS/MS operating in the negative mode ESI after separation on a reversed-phase column. Aldosterone was also measured by RIA. The calibration curves for analysis of serum aldosterone exhibited consistent linearity and reproducibility in the range of 60-3000 pmol/L. Interassay CVs were 4.3-7.5% at aldosterone concentrations of 97-993 pmol/L. The lower limit of quantitation (LOQ) was 30 pmol/L (signal to noise ratio = 10). The mean recovery of the analyte added to serum ranged from 95 to 102%. The regression equation by LC-MS/MS (x) and RIA (y) method was: y = 1.33x + 185 (r = 0.95; n = 124). Sensitivity and specificity of the LC-MS/MS method for serum aldosterone offer advantages over GC-MS by eliminating derivatization. The novel method is rapid, reliable and simple to perform with a routine LC-MS/MS spectrometer. The sensitivity is adequate for patient samples. Aldosterone concentrations reported by nonextraction RIA were consistently higher than those produced by LC-MS/MS. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:113 / 118
页数:6
相关论文
共 50 条
  • [11] Determination of flumethasone in calf urine and serum by liquid chromatography-tandem mass spectrometry
    Brambilla, G
    Buiarelli, F
    Cartoni, GP
    Coccioli, F
    Colamonici, C
    Fagiolo, A
    Giannini, C
    Neri, B
    JOURNAL OF CHROMATOGRAPHY B, 2001, 755 (1-2): : 265 - 278
  • [12] Determination of Bedaquiline in Human Serum Using Liquid Chromatography-Tandem Mass Spectrometry
    Alffenaar, Jan-Willem C.
    Bolhuis, Mathieu
    van Hateren, Kai
    Sturkenboom, Marieke
    Akkerman, Onno
    de Lange, Wiel
    Greijdanus, Ben
    van der Werf, Tjip
    Touw, Daan
    ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 2015, 59 (09) : 5675 - 5680
  • [13] Determination of evodiamine and rutecarpine in human serum by liquid chromatography-tandem mass spectrometry
    Wen, Dawei
    Li, Chenchen
    Liu, Yuping
    Liao, Yiping
    Liu, Huwei
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2006, 385 (06) : 1075 - 1081
  • [14] Determination of alprazolam by liquid chromatography-tandem mass spectrometry
    Svinarov, DA
    Kotzev, C
    Kassabova, LV
    CLINICA CHIMICA ACTA, 2005, 355 : S421 - S421
  • [15] Mucopolysaccharides quantitation in serum by liquid chromatography-tandem mass spectrometry
    Raymond, Kimiyo
    Lacey, Jean M.
    Dimitar, Gavrilov
    Matern, Dietrich
    Oglesbee, Devin
    Rinaldo, Piero
    Tortorelli, Silvia
    MOLECULAR GENETICS AND METABOLISM, 2018, 123 (02) : S123 - S123
  • [16] Liquid chromatography-tandem mass spectrometry
    Petrovic, Mira
    Barcelo, Damia
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2013, 405 (18) : 5857 - 5858
  • [17] Quantitative determination of methylnaltrexone in human serum using liquid chromatography-tandem mass spectrometry
    Oswald, Stefan
    Schumacher, Gitta
    Siegmund, Werner
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2011, 56 (05) : 1079 - 1084
  • [18] Liquid chromatography-tandem mass spectrometry for the determination of lithospermic acid B in rat serum
    Hui-Hyun Kim
    Hye Young Ji
    Hye Won Lee
    Youn-Chul Kim
    Dong Hwan Sohn
    Hye Suk Lee
    Archives of Pharmacal Research, 2004, 27 : 1202 - 1206
  • [19] Determination of free vitamin D in serum by ultrafiltration with liquid chromatography-tandem mass spectrometry
    Wang, D.
    CLINICA CHIMICA ACTA, 2022, 530 : S357 - S357
  • [20] Determination of parabens in serum by liquid chromatography-tandem mass spectrometry: Correlation with lipstick use
    Tahan, Gabriella Padovani
    Souza Santos, Nayara de Kassia
    Albuquerque, Ana Carolina
    Martins, Isarita
    REGULATORY TOXICOLOGY AND PHARMACOLOGY, 2016, 79 : 42 - 48