Performance of a New Test Strip for FreeStyle Blood Glucose Monitoring Systems

被引:8
|
作者
Lock, John Paul [2 ]
Brazg, Ronald [3 ]
Bernstein, Robert M. [4 ]
Taylor, Elizabeth [5 ]
Patel, Mona [1 ]
Ward, Jeanne [1 ]
Alva, Shridhara [1 ]
Chen, Ting [1 ]
Welsh, Zoe [1 ]
Amor, Walter [1 ]
Bhogal, Claire [1 ]
Ng, Ronald [1 ]
机构
[1] Abbott Diabet Care, Alameda, CA 94502 USA
[2] Internal Med & Endocrinol, Worcester, MA USA
[3] Rainier Clin Res, Renton, WA USA
[4] Reg Endocrinol Associates, Santa Fe, NM USA
[5] Mass Res, Waltham, MA USA
关键词
D O I
10.1089/dia.2010.0116
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: A new strip, designed to enhance the ease of use and minimize interference of non-glucose sugars, has been developed to replace the current FreeStyle (R) (Abbott Diabetes Care, Alameda, CA) blood glucose test strip. We evaluated the performance of this new strip. Methods: Laboratory evaluation included precision, linearity, dynamic range, effects of operating temperature, humidity, altitude, hematocrit, interferents, and blood reapplication. System accuracy, lay user performance, and ease of use for finger capillary blood testing and accuracy for venous blood testing were evaluated at clinics. Lay users also compared the speed and ease of use between the new strip and the current FreeStyle strip. Results: For glucose concentrations <75 mg/dL, 73%, 100%, and 100% of the individual capillary blood glucose results obtained by lay users fell within +/- 5, 10, and 15 mg/dL, respectively, of the reference. For glucose concentrations >= 75 mg/dL, 68%, 95%, 99%, and 99% of the lay user results fell within +/- 5%, 10%, 15%, and 20%, respectively, of the reference. Comparable accuracy was obtained in the venous blood study. Lay users found the new test strip easy to use and faster and easier to use than the current FreeStyle strip. The new strip maintained accuracy under various challenging conditions, including high concentrations of various interferents, sample reapplication up to 60 s, and extremes in hematocrit, altitude, and operating temperature and humidity. Conclusions: Our results demonstrated excellent accuracy of the new FreeStyle test strip and validated the improvements in minimizing interference and enhancing ease of use.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 50 条
  • [11] Performance of FreeStyle navigator against standards for continuous glucose monitoring systems and ISO15197 system accuracy blood glucose meter guidelines
    Bugler, J. R.
    Pryor, H.
    Harris, C.
    DIABETOLOGIA, 2007, 50 : S419 - S419
  • [12] Performance of the FreeStyle Libre Flash Glucose Monitoring System during an Oral Glucose Tolerance Test and Exercise in Healthy Adolescents
    Afeef, Sahar
    Tolfrey, Keith
    Zakrzewski-Fruer, Julia K.
    Barrett, Laura A.
    SENSORS, 2023, 23 (09)
  • [13] Nanoplasmonic Strip Test for Salivary Glucose Monitoring
    Torne-Morato, Helena
    Donati, Paolo
    Pompa, Pier Paolo
    NANOMATERIALS, 2022, 12 (01)
  • [14] Performance of a New Blood Glucose Monitoring System Using OneTouch® Verio™ Test Strips
    Katz, Laurence B.
    Stewart, Lorna S.
    Cameron, Hilary
    Salgado, Anna
    Setford, Steven
    DIABETES, 2015, 64 : A639 - A639
  • [15] DIABETIC HOME BLOOD-GLUCOSE MONITORING USING A DIRECT READING GLUCOSE SENSITIVE TEST STRIP
    BAKER, JR
    MULVEY, TT
    VIVIAN, JA
    NEW ZEALAND MEDICAL JOURNAL, 1981, 94 (693) : 252 - 254
  • [16] Clinical evaluation of a novel test strip technology for blood glucose monitoring: Accuracy at hypoglycaemic glucose levels
    Bellary, Srikanth
    Cameron, Hillary
    Macleod, Kirsty
    Malecha, Michael
    Koria, Krisna
    Raja, Praveen
    Diago Cabezudo, Jesus
    Ellison, John
    DIABETES RESEARCH AND CLINICAL PRACTICE, 2012, 98 (03) : 430 - 435
  • [17] PERFORMANCE EVALUATION OF FOUR BLOOD GLUCOSE MONITORING SYSTEMS
    Link, M.
    Freckmann, G.
    Baumstark, A.
    Schmid, C.
    Pleus, S.
    Haug, C.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2014, 16 : A74 - A74
  • [18] ACCURACY, PRECISION, AND USER PERFORMANCE EVALUATION OF THE CONTOUR® NEXT BLOOD GLUCOSE MONITORING SYSTEM WITH THE CONTOUR® NEXT TEST STRIP
    Bernstein, R. M.
    Caswell, M.
    Frank, J.
    Fullam, J.
    Pardo, S.
    Harrison, B.
    Dunne, N.
    Wallace, J. F.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 : A55 - A55
  • [19] Performance evaluation of "Advantage Test Strip S" type blood glucose test paper.
    Kuwa, K
    Tsubaki, A
    Ogura, K
    CLINICAL CHEMISTRY, 2002, 48 (06) : A197 - A197
  • [20] Performance of the FreeStyle Navigator Continuous Glucose Monitoring System with the TRUstart Algorithm
    Weinstein, Richard
    Brazg, Ronald
    DIABETES, 2009, 58 : A109 - A109