Performance of a New Test Strip for FreeStyle Blood Glucose Monitoring Systems

被引:8
|
作者
Lock, John Paul [2 ]
Brazg, Ronald [3 ]
Bernstein, Robert M. [4 ]
Taylor, Elizabeth [5 ]
Patel, Mona [1 ]
Ward, Jeanne [1 ]
Alva, Shridhara [1 ]
Chen, Ting [1 ]
Welsh, Zoe [1 ]
Amor, Walter [1 ]
Bhogal, Claire [1 ]
Ng, Ronald [1 ]
机构
[1] Abbott Diabet Care, Alameda, CA 94502 USA
[2] Internal Med & Endocrinol, Worcester, MA USA
[3] Rainier Clin Res, Renton, WA USA
[4] Reg Endocrinol Associates, Santa Fe, NM USA
[5] Mass Res, Waltham, MA USA
关键词
D O I
10.1089/dia.2010.0116
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: A new strip, designed to enhance the ease of use and minimize interference of non-glucose sugars, has been developed to replace the current FreeStyle (R) (Abbott Diabetes Care, Alameda, CA) blood glucose test strip. We evaluated the performance of this new strip. Methods: Laboratory evaluation included precision, linearity, dynamic range, effects of operating temperature, humidity, altitude, hematocrit, interferents, and blood reapplication. System accuracy, lay user performance, and ease of use for finger capillary blood testing and accuracy for venous blood testing were evaluated at clinics. Lay users also compared the speed and ease of use between the new strip and the current FreeStyle strip. Results: For glucose concentrations <75 mg/dL, 73%, 100%, and 100% of the individual capillary blood glucose results obtained by lay users fell within +/- 5, 10, and 15 mg/dL, respectively, of the reference. For glucose concentrations >= 75 mg/dL, 68%, 95%, 99%, and 99% of the lay user results fell within +/- 5%, 10%, 15%, and 20%, respectively, of the reference. Comparable accuracy was obtained in the venous blood study. Lay users found the new test strip easy to use and faster and easier to use than the current FreeStyle strip. The new strip maintained accuracy under various challenging conditions, including high concentrations of various interferents, sample reapplication up to 60 s, and extremes in hematocrit, altitude, and operating temperature and humidity. Conclusions: Our results demonstrated excellent accuracy of the new FreeStyle test strip and validated the improvements in minimizing interference and enhancing ease of use.
引用
收藏
页码:1 / 10
页数:10
相关论文
共 50 条
  • [1] Performance of a New Blood Glucose Test Strip
    Bailey, Timothy
    Chu, Amy
    Parkes, Joan Lee
    Wallace, Jane
    Pardo, Scott
    Schachner, Holly C.
    Castro, Rita
    Simmons, David A.
    DIABETES, 2011, 60 : A246 - A246
  • [2] Performance evaluation of FreeStyle for blood glucose monitoring system
    Kuwa, K
    EUROMEDLAB 2003: PROCEEDINGS OF THE 15TH IFCC-FESCC EUROPEAN CONGRESS OF CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, : 587 - 591
  • [3] Comparison of test strip stability for four blood glucose monitoring systems
    Harrison, BE
    Kueppers, LE
    Slough, CH
    Chiddister, SK
    DIABETES, 2004, 53 : A104 - A104
  • [4] ACCURACY EVALUATION OF A NEW PLATFORM OF BLOOD GLUCOSE MONITORING SYSTEMS WITH THE CONTOUR® NEXT TEST STRIP
    Harrison, B.
    Brown, D.
    Warchal-Windham, M. E.
    Schachner, H. C.
    Pardo, S.
    Goldy, A.
    Parkes, J. L.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 : A72 - A72
  • [5] ASSESSMENT OF A NEW VISUAL TEST STRIP FOR BLOOD-GLUCOSE MONITORING
    MARSHALL, SM
    ALBERTI, KGMM
    DIABETES CARE, 1983, 6 (06) : 543 - 547
  • [6] A comprehensive evaluation of strip performance in multiple blood glucose monitoring systems
    Katz, Laurence B.
    Macleod, Kirsty
    Grady, Mike
    Cameron, Hilary
    Pfuetzner, Andreas
    Setford, Steven
    EXPERT REVIEW OF MEDICAL DEVICES, 2015, 12 (03) : 263 - 271
  • [7] CLINICAL-PERFORMANCE OF A NEW TEST STRIP WITH THE MEDISENSE BLOOD-GLUCOSE SENSOR
    NG, R
    MARTIN, L
    HALPIN, M
    BERNSTEIN, R
    FISCHER, J
    TAYLOR, E
    SCHRODER, S
    CLINICAL CHEMISTRY, 1995, 41 (06) : S181 - S181
  • [8] Self-Monitoring Blood Glucose Test Strip Utilization in Canada
    Yeaw, Jason
    Christensen, Torsten E.
    Groleau, Danielle
    Wolden, Michael L.
    Lee, Won C.
    DIABETES, 2012, 61 : A35 - A35
  • [9] Ongoing Manufacturer-Led Clinical Surveillance of Two Blood Glucose Monitoring Test Strip Systems
    Setford, Steven J.
    Katz, Laurence B.
    Grady, Mike
    Macleod, Kirsty
    Cameron, Hilary
    DIABETES, 2020, 69
  • [10] Arterial blood testing using the Freestyle blood glucose monitoring system
    Smith, AC
    Wu, AH
    CLINICAL CHEMISTRY, 2005, 51 : A260 - A260