Determination of chlorine ions in raw milk by pulsed amperometric detection in a flow injection system

被引:13
|
作者
Chen, Xingguang [1 ]
Chu, Beibei [1 ]
Xi, Huiting [1 ]
Xu, Junjun [1 ]
Lai, Lizhi [1 ]
Peng, Huifeng [1 ]
Deng, Danwen [1 ]
Huang, Ganhui [1 ]
机构
[1] Nanchang Univ, State Key Lab Food Sci & Technol, Nanchang 330047, Jiangxi, Peoples R China
基金
中国国家自然科学基金;
关键词
chloride ion; pulsed amperometric detection; flow injection system; milk adulteration; ADULTERATION; CHROMATOGRAPHY; GOLD;
D O I
10.3168/jds.2018-14395
中图分类号
S8 [畜牧、 动物医学、狩猎、蚕、蜂];
学科分类号
0905 ;
摘要
Chloride ion concentration in milk was determined by pulsed amperometric detection in a flow injection system. Results showed that the Au electrode lost 3 electrons at 1.10 V and formed chloroaurate ions (AuCl4-) by combining with chloride ions, after which AuCl4- was partly reduced to Au at 0.6 V. Based on the electrochemical process, a triple waveform with detection potential of 1.15 V, detection time of 150 ms, oxidation potential of 1.4 V, oxidation time of 550 ms, reduction potential of 0 V, and reduction time of 400 ms was applied to the Au electrode for detecting chloride ion concentration in milk. The approach is rapid and automatic and features a detection limit of 0.005 g/L. The relative standard deviation obtained by 60 repetitive injections reached 1.48% at 2 g/L of NaCl. The method developed using the Au electrode without modification was used to analyze the chloride ion concentration in raw milk without preprocessing. The method showed good agreement with potentiometric titration.
引用
收藏
页码:9647 / 9658
页数:12
相关论文
共 50 条
  • [1] Pulsed amperometric detection of metal ions complexing with EDTA in a flow injection system
    Lee, JW
    Yeo, IH
    Pyun, CH
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 1997, 18 (03) : 316 - 318
  • [2] Optimization of a flow injection system with amperometric detection for arsenic determination
    Lolic, Aleksandar
    Nikolic, Snezana
    Mutic, Jelena
    ANALYTICAL SCIENCES, 2008, 24 (07) : 877 - 880
  • [3] Optimization of a Flow Injection System with Amperometric Detection for Arsenic Determination
    Aleksandar Lolic
    Snezana Nikolic
    Jelena Mutic
    Analytical Sciences, 2008, 24 : 877 - 880
  • [4] Flow injection and HPLC determination of furosemide using pulsed amperometric detection at microelectrodes
    Guzmán, A
    Agüí, L
    Pedrero, M
    Yáñez-Sedeño, P
    Pingarrón, JM
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2003, 33 (05) : 923 - 933
  • [5] FLOW-INJECTION AMPEROMETRIC DETERMINATION OF CHLORINE AT A GOLD ELECTRODE
    TSAOUSIS, AN
    HUBER, CO
    ANALYTICA CHIMICA ACTA, 1985, 178 (02) : 319 - 323
  • [6] Quadruple-pulsed amperometric detection for simultaneous flow injection determination of glucose and fructose
    Surareungchai, W
    Deepunya, W
    Tasakorn, P
    ANALYTICA CHIMICA ACTA, 2001, 448 (1-2) : 215 - 220
  • [7] INDIRECT DETERMINATION OF IRON IN A FLOW-INJECTION SYSTEM WITH AMPEROMETRIC DETECTION
    SILVA, L
    LEON, LE
    CALVO, A
    ANALYTICA CHIMICA ACTA, 1987, 192 (02) : 349 - 353
  • [8] Determination of 2-furaldehyde in transformer oil using flow injection with pulsed amperometric detection
    Dilleen, JW
    Lawrence, CM
    Slater, JM
    ANALYST, 1996, 121 (06) : 755 - 759
  • [9] Flow Injection System with Amperometric Detection for Determination of Iodine Value of Vegetable Oils
    Kruatian, Thidarat
    Jitmanee, Kritsana
    Jakmunee, Jaroon
    CHIANG MAI JOURNAL OF SCIENCE, 2021, 48 (02): : 557 - 567
  • [10] PULSED AMPEROMETRIC DETECTION OF ELECTROINACTIVE ADSORBATES AT PLATINUM-ELECTRODES IN A FLOW-INJECTION SYSTEM
    POLTA, JA
    JOHNSON, DC
    ANALYTICAL CHEMISTRY, 1985, 57 (07) : 1373 - 1376