Performance evaluation of a trace-moisture analyzer based on cavity ring-down spectroscopy: Direct comparison with the NMIJ trace-moisture standard

被引:34
|
作者
Abe, Hisashi [1 ]
Yamada, Koichi M. T. [2 ]
机构
[1] AIST, Natl Metrol Inst Japan NMIJ, Tsukuba, Ibaraki 3058563, Japan
[2] AIST, Inst Environm Management Technol EMTech, Tsukuba, Ibaraki 3058569, Japan
关键词
Diffusion tube; Humidity standard; Measurement performance; Metrological traceability; Sensor test; Standard gas; MASS-FLOW RATES; ABSORPTION-SPECTROSCOPY; H2O; SPECTROMETER; UNCERTAINTY; GENERATION; NOZZLES;
D O I
10.1016/j.sna.2010.11.005
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The performance of a moisture analyzer (MA) based on cavity ring-down spectroscopy (CRDS) was evaluated by direct comparison with the primary trace-moisture standard developed at the National Metrology Institute of Japan (NMIJ). The limit of detection for trace moisture in nitrogen gas by the CRDS-based MA, expressed in amount fraction (mole fraction), was estimated to be 1 nmol mol(-1) or less. The CRDS-based MA showed excellent performance in terms of linearity, accuracy, time response, stability, and reproducibility over four years. It was found for a moderate range of pressures and temperatures that we can measure trace moisture in nitrogen gas with a relative standard uncertainty of approximately 4% down to approximately 10 nmol mol(-1), even using a simple CRDS-based MA that measures only the peak intensity of the absorption line, provided that the MA is properly calibrated on the basis of a reliable trace-moisture standard. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:230 / 238
页数:9
相关论文
共 50 条
  • [41] Cavity ring-down spectroscopy for atmospheric trace gas detection: application to the nitrate radical (NO3)
    S.S. Brown
    H. Stark
    A.R. Ravishankara
    Applied Physics B, 2002, 75 : 173 - 182
  • [42] 410-nm diode laser cavity ring-down spectroscopy for trace detection of NO2
    Mazurenka, MI
    Fawcett, BL
    Elks, JMF
    Shallcross, DE
    Orr-Ewing, AJ
    CHEMICAL PHYSICS LETTERS, 2003, 367 (1-2) : 1 - 9
  • [43] Detection of trace water vapor in high-purity phosphine using cavity ring-down spectroscopy
    Funke, Hans H.
    Raynor, Mark W.
    Bertness, Kris A.
    Chen, Yu
    APPLIED SPECTROSCOPY, 2007, 61 (04) : 419 - 423
  • [44] Combining preconcentration of air samples with cavity ring-down spectroscopy for detection of trace volatile organic compounds in the atmosphere
    Parkes, AM
    Lindley, RE
    Orr-Ewing, AJ
    ANALYTICAL CHEMISTRY, 2004, 76 (24) : 7329 - 7335
  • [45] A DFG-based cavity ring-down spectrometer for trace gas sensing in the mid-infrared
    Whittaker, K. E.
    Ciaffoni, L.
    Hancock, G.
    Peverall, R.
    Ritchie, G. A. D.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2012, 109 (02): : 333 - 343
  • [46] A DFG-based cavity ring-down spectrometer for trace gas sensing in the mid-infrared
    K. E. Whittaker
    L. Ciaffoni
    G. Hancock
    R. Peverall
    G. A. D. Ritchie
    Applied Physics B, 2012, 109 : 333 - 343
  • [47] Cavity Ring-Down Spectroscopy for Molecular Trace Gas Detection Using A Pulsed DFB QCL Emitting at 6.8 μm
    Gadedjisso-Tossou, Komlan S.
    Stoychev, Lyubomir, I
    Mohou, Messanh A.
    Cabrera, Humberto
    Niemela, Joseph
    Danailov, Miltcho B.
    Vacchi, Andrea
    PHOTONICS, 2020, 7 (03)
  • [48] Detection of Trace Amounts of NO2 Using Frequency Modulated Off-Axis Cavity Ring-Down Spectroscopy
    Karpf, Andreas
    Rao, Gottipaty N.
    2012 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2012,
  • [49] Evaluation of local heating in Doppler-broadening thermometry based on cavity ring-down spectroscopy
    Moretti, Luigi
    Castrillo, Antonio
    Gianfrani, Livio
    PHYSICAL REVIEW A, 2019, 100 (04)
  • [50] Quantitative nonlinear spectroscopy: A direct comparison of degenerate four-wave mixing with cavity ring-down spectroscopy applied to NaH
    Lehr, L
    Hering, P
    IEEE JOURNAL OF QUANTUM ELECTRONICS, 1997, 33 (09) : 1465 - 1473