Performance evaluation of a trace-moisture analyzer based on cavity ring-down spectroscopy: Direct comparison with the NMIJ trace-moisture standard

被引:34
|
作者
Abe, Hisashi [1 ]
Yamada, Koichi M. T. [2 ]
机构
[1] AIST, Natl Metrol Inst Japan NMIJ, Tsukuba, Ibaraki 3058563, Japan
[2] AIST, Inst Environm Management Technol EMTech, Tsukuba, Ibaraki 3058569, Japan
关键词
Diffusion tube; Humidity standard; Measurement performance; Metrological traceability; Sensor test; Standard gas; MASS-FLOW RATES; ABSORPTION-SPECTROSCOPY; H2O; SPECTROMETER; UNCERTAINTY; GENERATION; NOZZLES;
D O I
10.1016/j.sna.2010.11.005
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The performance of a moisture analyzer (MA) based on cavity ring-down spectroscopy (CRDS) was evaluated by direct comparison with the primary trace-moisture standard developed at the National Metrology Institute of Japan (NMIJ). The limit of detection for trace moisture in nitrogen gas by the CRDS-based MA, expressed in amount fraction (mole fraction), was estimated to be 1 nmol mol(-1) or less. The CRDS-based MA showed excellent performance in terms of linearity, accuracy, time response, stability, and reproducibility over four years. It was found for a moderate range of pressures and temperatures that we can measure trace moisture in nitrogen gas with a relative standard uncertainty of approximately 4% down to approximately 10 nmol mol(-1), even using a simple CRDS-based MA that measures only the peak intensity of the absorption line, provided that the MA is properly calibrated on the basis of a reliable trace-moisture standard. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:230 / 238
页数:9
相关论文
共 50 条
  • [31] Trace Detection of NO2 Using Cavity Ring-Down Spectroscopy and a Diode Laser
    Karpf, Andreas
    Rao, Gottipaty N.
    2013 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2013,
  • [32] Trace detection of volatile organic compounds by diode laser cavity ring-down spectroscopy
    Parkes, AM
    Fawcett, BL
    Austin, RE
    Nakamichi, S
    Shallcross, DE
    Orr-Ewing, AJ
    ANALYST, 2003, 128 (07) : 960 - 965
  • [33] Cavity Ring-Down Spectroscopy for Gaseous Fission Products Trace Measurements in Sodium Fast Reactors
    Jacquet, P.
    Pailloux, A.
    Aoust, G.
    Jeannot, J-P
    Doizi, D.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (04) : 2011 - 2016
  • [34] Cavity Ring-Down Spectroscopy for Gaseous Fission Products Trace Measurements in Sodium Fast Reactors
    Jacquet, P.
    Pailloux, A.
    Aoust, G.
    Jeannot, J. -P.
    Doizi, D.
    2013 3RD INTERNATIONAL CONFERENCE ON ADVANCEMENTS IN NUCLEAR INSTRUMENTATION, MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA), 2013,
  • [35] Open-path cavity ring-down spectroscopy for trace gas measurements in ambient air
    McHale, Laura E.
    Hecobian, Arsineh
    Yalin, Azer P.
    OPTICS EXPRESS, 2016, 24 (05): : 5523 - 5535
  • [36] Trace detection of methane using continuous wave cavity ring-down spectroscopy at 1.65 μm
    Fawcett, BL
    Parkes, AM
    Shallcross, DE
    Orr-Ewing, AJ
    PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 2002, 4 (24) : 5960 - 5965
  • [37] Wavelength-meter controlled cavity ring-down spectroscopy: high-sensitivity detection of trace moisture in N2 at sub-ppb levels
    Hashiguchi, Koji
    Lisak, Daniel
    Cygan, Agata
    Ciurylo, Roman
    Abe, Hisashi
    SENSORS AND ACTUATORS A-PHYSICAL, 2016, 241 : 152 - 160
  • [38] Using Cavity Ring-Down Spectroscopy to Measure Trace Impurities in Noble Gases for Dark Matter Detection
    Keeter, K. J.
    Mount, B. J.
    Serfling, G. L.
    WORKSHOP ON DARK MATTER, UNIFICATION AND NEUTRINO PHYSICS (CETUP 2012), 2013, 1534 : 96 - 101
  • [39] Detection of trace gases using frequency modulated off-axis cavity ring-down spectroscopy
    Karpf, Andreas
    Rao, Gottipaty N.
    CHEMICAL, BIOLOGICAL, RADIOLOGICAL, NUCLEAR, AND EXPLOSIVES (CBRNE) SENSING XIII, 2012, 8358
  • [40] Cavity ring-down spectroscopy for atmospheric trace gas detection:: application to the nitrate radical (NO3)
    Brown, SS
    Stark, H
    Ravishankara, AR
    APPLIED PHYSICS B-LASERS AND OPTICS, 2002, 75 (2-3): : 173 - 182