Monte Carlo dosimetry for forthcoming clinical trials in x-ray microbeam radiation therapy

被引:41
|
作者
Martinez-Rovira, I. [1 ,2 ]
Sempau, J. [2 ]
Fernandez-Varea, J. M. [3 ]
Bravin, A. [1 ]
Prezado, Y. [1 ]
机构
[1] European Synchrotron Radiat Facil, Biomed Beamline ID17, F-38043 Grenoble, France
[2] Univ Politecn Cataluna, Inst Tecn Energet, E-08028 Barcelona, Spain
[3] Univ Barcelona, Fac Fis ECM & ICC, E-08028 Barcelona, Spain
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2010年 / 55卷 / 15期
关键词
LINEAR-QUADRATIC MODEL; PRIMARY BRAIN-TUMORS; MICROPLANAR BEAM; STEREOTACTIC RADIOTHERAPY; ADJUVANT TEMOZOLOMIDE; 9L GLIOSARCOMA; SYNCHROTRON; SIMULATION; CONCOMITANT; GLIOBLASTOMA;
D O I
10.1088/0031-9155/55/15/012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The purpose of this work is to define safe irradiation protocols in microbeam radiation therapy. The intense synchrotron-generated x-ray beam used for the treatment is collimated and delivered in an array of 50 mu m-sized rectangular fields with a centre-to-centre distance between microplanes of 400 mu m. The absorbed doses received by the tumour and the healthy tissues in a human head phantom have been assessed by means of Monte Carlo simulations. The identification of safe dose limits is carried out by evaluating the maximum peak and valley doses achievable in the tumour while keeping the valley doses in the healthy tissues under tolerances. As the skull receives a significant fraction of the dose, the dose limits are referred to this tissue. Dose distributions with high spatial resolution are presented for various tumour positions, skull thicknesses and interbeam separations. Considering a unidirectional irradiation (field size of 2x2 cm(2)) and a centrally located tumour, the largest peak and valley doses achievable in the tumour are 55 Gy and 2.6 Gy, respectively. The corresponding maximum valley doses received by the skin, bone and healthy brain are 4 Gy, 14 Gy and 7 Gy (doses in one fraction), respectively, i.e. within tolerances (5% probability of complication within 5 years).
引用
收藏
页码:4375 / 4388
页数:14
相关论文
共 50 条
  • [1] MONTE CARLO DOSIMETRY FOR THE CLINICAL TRIALS IN MICROBEAM RADIATION THERAPY
    Martinez-Rovira, I.
    Alonso, Y. Prezado
    Sempau, J.
    Fernandez-Varea, J. M.
    Bravin, A.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S71 - S72
  • [2] Development and commissioning of a Monte Carlo photon beam model for the forthcoming clinical trials in microbeam radiation therapy
    Martinez-Rovira, I.
    Sempau, J.
    Prezado, Y.
    MEDICAL PHYSICS, 2012, 39 (01) : 119 - 131
  • [3] Monte Carlo study of a mosfet detector response applied to x-ray Microbeam Radiation Therapy
    De Felici, M
    Felici, R
    Del Rio, MS
    Dilmanian, A
    Ferreo, C
    ASTROPARTICLE, PARTICLE AND SPACE PHYSICS , DETECTORS AND MEDICAL PHYSICS APPLICATIONS, 2004, 2 : 366 - 373
  • [4] Monte Carlo Study of a Compact Distributed X-Ray Source Microbeam Radiation Therapy Research Irradiator
    Schreiber, E.
    Chang, S.
    MEDICAL PHYSICS, 2012, 39 (06) : 3899 - 3899
  • [5] Monte Carlo simulation of stereotactic microbeam radiation therapy: evaluation of the usage of a linear accelerator as the x-ray source
    Gokeri, Gurdal
    Kocar, Cemil
    Tombakoglu, Mehmet
    Cecen, Yigit
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (13): : 4621 - 4642
  • [6] MONTE CARLO DOSIMETRY TO PREPARE THE PRE-CLINICAL TRIALS IN MINIBEAM RADIATION THERAPY
    Alonso, Y. Prezado
    Thengumpallil, S.
    Martinez-Rovira, I.
    Renier, M.
    Bravin, A.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S221 - S221
  • [7] Optimization of a compact x-ray source for clinical microbeam radiation therapy
    Winter, J.
    Wilkens, J. J.
    Combs, S. E.
    Bartzsch, S.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S264 - S265
  • [8] X-ray microbeam radiation therapy calculations, including polarisation effects, with the Monte Carlo code EGS5
    Hugtenburg, Richard P.
    Adegunloye, A. S.
    Bradley, David A.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3): : 221 - 224
  • [9] MOSFET dosimetry of an X-ray microbeam
    Rosenfeld, A.B.
    Kaplan, G.I.
    Kron, T.
    Allen, B.J.
    Dilmanian, A.
    Orion, I.
    Ren, B.
    Lerch, M.L.F.
    Holmes-Siedle, A.
    IEEE Transactions on Nuclear Science, 1999, 46 (6 I): : 1774 - 1780
  • [10] Diagnostic x-ray dosimetry using Monte Carlo simulation
    Ioppolo, JL
    Price, RI
    Tuchyna, T
    Buckley, CE
    PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (10): : 1707 - 1720