Volumetric ultrasound Imaging using 2-D CMUT arrays

被引:140
|
作者
Oralkan, Ö
Ergun, AS
Cheng, CH
Johnson, JA
Karaman, M
Lee, TH
Khuri-Yakub, BT
机构
[1] Stanford Univ, Edward L Ginzton Lab, Stanford, CA 94305 USA
[2] Stanford Univ, Ctr Integrated Syst, Stanford, CA 94305 USA
[3] Isik Univ, Dept Elect Engn, TR-80670 Istanbul, Turkey
[4] Stanford Univ, Sch Med, Dept Neurosurg, Image Guidance Lab, Stanford, CA 94305 USA
关键词
D O I
10.1109/TUFFC.2003.1251142
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Recently, capacitive micromachined ultrasonic transducers (CMUTs) have emerged as a candidate to overcome the difficulties in the realization of 2-D arrays for real-time 3-D imaging. In this paper, we present the first volumetric images obtained using a 2-D CMUT array. We have fabricated a 128 x 128-element 2-D CMUT array with through-wafer via interconnects and a 420-mum element pitch. As an experimental prototype, a 32 x 64-element portion of the 128 X 128-element array was diced and flip-chip bonded onto a glass fanout chip. This chip provides individual leads from a central 16 X 16-element portion of the array to surrounding bondpads. An 8 x 16-element poition of the array was used in the experiments along with a 128-channel data acquisition system. For imaging phantoms, we used a 2.37-mm diameter steel sphere located 10 mm from the array center and two 12-mm-thick Plexiglas plates located 20 mm and 60 mm from the array. A 4 X 4 group of elements in the middle of the 8 X 16-element array was used in transmit, and the remaining elements were used to receive the echo signals. The echo signal obtained from the spherical target presented a frequency spectrum centered at 4.37 MHz with a 100% fractional bandwidth, whereas the frequency spectrum for the echo signal from the parallel plate phantom was centered at 3.44 MHz with a 91% fractional bandwidth. The images were reconstructed by using RF beamforming and synthetic phased array approaches and visualized by surface rendering and multiplanar slicing techniques. The image of the spherical target has been used to approximate the point spread function of the system and is compared with theoretical expectations. This study experimentally demonstrates that 2-D CMUT arrays can be fabricated with high yield using silicon IC-fabrication processes, individual electrical connections can be provided using through-wafer vias, and flip-chip bonding can be used to integrate these dense 2-D arrays with electronic circuits for practical 3-D imaging applications.
引用
收藏
页码:1581 / 1594
页数:14
相关论文
共 50 条
  • [41] Low-Cost Scalable PCB-Based 2-D Transducer Arrays for Volumetric Photoacoustic Imaging
    Mitra, Mahaan
    Kumar, Akshay
    Khandare, Shubham
    Gaddale, Prameth
    Anandan, Yashoda
    Pedibhotla, Srian
    Roy, Kaustav
    Chen, Haoyang
    Pratap, Rudra
    Kothapalli, Sri-Rajasekhar
    [J]. IEEE SENSORS JOURNAL, 2024, 24 (04) : 4380 - 4386
  • [42] Integration of a Dual-Mode Catheter for Ultrasound Image Guidance and HIFU Ablation using a 2-D CMUT Array
    Jang, Ji Hoon
    Chang, Chienliu
    Rasmussen, Morten Fischer
    Moini, Azadeh
    Brenner, Kevin
    Stephens, Douglas N.
    Oralkan, Omer
    Khuri-Yakub, Butrus
    [J]. 2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [43] Nonlinear ultrasound imaging experiments using a CMUT probe
    Savoia, Alessandro Stuart
    Caliano, Giosue
    Matrone, Giulia
    Ramalli, Alessandro
    Boni, Enrico
    Tortoli, Piero
    [J]. 2016 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2016,
  • [44] Ultrasonic manipulation of particles on fluid surface using 2-D CMUT array
    Lee, Chang Hoon
    Park, Bum Hoon
    Park, Kwan Kyu
    [J]. 2018 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2018,
  • [45] 2-D arrays
    von Ramm, OT
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2000, 26 : S10 - S12
  • [46] 2-D NIR imaging reconstruction with ultrasound guidance
    Huang, MM
    Xie, TQ
    Chen, NG
    Piao, DQ
    Zhu, Q
    [J]. 2002 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING, PROCEEDINGS, 2002, : 1031 - 1034
  • [47] SPECTRAL ESTIMATION FOR 2-D DOPPLER ULTRASOUND IMAGING
    COSTA, A
    DEFRANCISCIS, A
    DEGLORIA, A
    FARABOSCHI, P
    OLIVIERI, M
    [J]. ELECTRONICS LETTERS, 1992, 28 (23) : 2177 - 2179
  • [48] High Resolution 2-D Electromagnetic Vortex Imaging Using Uniform Circular Arrays
    Guo, Shaoqing
    He, Zi
    Chen, Rushan
    [J]. IEEE ACCESS, 2019, 7 : 132430 - 132437
  • [49] Volumetric Real-Time Imaging Using a CMUT Ring Array
    Choe, Jung Woo
    Oralkan, Oemer
    Nikoozadeh, Amin
    Gencel, Mustafa
    Stephens, Douglas N.
    O'Donnell, Matthew
    Sahn, David J.
    Khuri-Yakub, Butrus T.
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2012, 59 (06) : 1201 - 1211
  • [50] 2-D army for 3-D ultrasound imaging using synthetic aperture techniques
    Daher, NM
    Yen, JT
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2006, 53 (05) : 912 - 924