Integration of a Dual-Mode Catheter for Ultrasound Image Guidance and HIFU Ablation using a 2-D CMUT Array

被引:0
|
作者
Jang, Ji Hoon [1 ]
Chang, Chienliu [1 ]
Rasmussen, Morten Fischer [1 ]
Moini, Azadeh [1 ]
Brenner, Kevin [1 ]
Stephens, Douglas N. [2 ]
Oralkan, Omer [3 ]
Khuri-Yakub, Butrus [1 ]
机构
[1] Stanford Univ, Elect Engn, Stanford, CA 94305 USA
[2] Univ Calif Davis, Biomed Engn, Davis, CA 95616 USA
[3] North Carolina State Univ, Elect & Comp Engn, Raleigh, NC USA
基金
美国国家卫生研究院;
关键词
Dual-Mode Catheter; Integration; flip-chip bonding; Flexible PCB; Ultrasound Image-guided HIFU; INTENSITY FOCUSED ULTRASOUND; SURGERY;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Image-guided high-intensity focused ultrasound (HIFU) is widely used not only for non-invasive therapy but also for a precise approach for tissue ablation. Most HIFU systems use piezoelectric transducers, which are typically bulky due to active cooling, and separate imaging and HIFU transducers, and are therefore impractical for catheter-based applications. Taking advantage of a single 2-D capacitive micromachined ultrasonic transducer (CMUT) array, we developed a dual-mode catheter that can switch between ultrasound imaging mode and HIFU ablation mode. The catheter is equipped with an application-specific integrated circuit (ASIC) and a 32 x 32-element 2-D CMUT array. Both ASIC and CMUT are flip-chip bonded to a custom-designed flexible printed circuit board (flex PCB) via 100-mu m and 80-mu m solder balls. Then, the flex legs are folded and terminated with pads for a micro zero insertion force (mu ZIF) connector, allowing easy assembly replacement without the extra cost of coaxial cable assembly. Next, the micro-coaxial cables are assembled at the end of the mu ZIF connectors. After integration with a 3-D printed tip and encapsulating with polydimethylsiloxane (PDMS), the catheter is finalized in a 22-mm diameter shaft. We successfully validated the functionality of both modes of the dual-mode catheter in oil. We are currently preparing the test for an animal study.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Dual-Mode Integrated Circuit for Imaging and HIFU With 2-D CMUT Arrays
    Jang, Ji Hoon
    Rasmussen, Morten Fischer
    Bhuyan, Anshuman
    Yoon, Hyo-Seon
    Moini, Azadeh
    Chang, Chienliu
    Watkins, Ronald D.
    Choe, Jung Woo
    Nikoozadeh, Amin
    Stephens, Douglas
    Oralkan, Oemer
    Pauly, Kim Butts
    Khuri-Yakub, Butrus
    [J]. 2015 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2015,
  • [2] A 1-MHz 2-D CMUT Array for HIFU Thermal Ablation
    Yoon, Hyo-Seon
    Vaithilingam, Srikant
    Park, Kwan Kyu
    Nikoozadeh, Amin
    Firouzi, Kamyar
    Choe, Jung Woo
    Watkins, Ronald D.
    Oguz, Huseyin Kagan
    Kupnik, Mario
    Pauly, Kim Butts
    Khuri-Yakub, Pierre
    [J]. PROCEEDINGS FROM THE 14TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2017, 1821
  • [3] Monitoring and Guidance of HIFU Beams with Dual-Mode Ultrasound Arrays
    Ballard, John R.
    Casper, Andrew J.
    Ebbini, Emad S.
    [J]. 2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 137 - 140
  • [4] Continuous Monitoring of Pulsed HIFU Beams Using Dual-mode Ultrasound Array Systems
    Casper, Andrew
    Ballard, John
    Liu, Dalong
    Haritonova, Alyona
    Ebbini, Emad
    [J]. 11TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2012, 1481 : 37 - 40
  • [5] Compact Fast-Switching DC and Resonant RF Drivers for a Dual-Mode Imaging and HIFU 2D CMUT Array
    Stedman, Quintin
    Gu, Lei
    Pai, Chi-Nan
    Rasmussen, Morten
    Brenner, Kevin
    Ma, Bo
    Ergun, Arif Sanli
    Davila, Juan Rivas
    Khuri-Yakub, Butrus
    [J]. 2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1951 - 1954
  • [6] Enhanced image resolution of dual-mode ultrasound array using coded excitation
    Wan, Yayun
    Shrestha, Ajay
    Ebbini, Emad S.
    [J]. 2007 4TH IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING : MACRO TO NANO, VOLS 1-3, 2007, : 1128 - 1131
  • [7] A Dual-Mode 2D Matrix Array for Ultrasound Image-Guided Noninvasive Therapy
    Zhang, Zhiqiang
    Liu, Rong
    Li, Guofeng
    Su, Min
    Li, Fei
    Zheng, Hairong
    Qiu, Weibao
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2021, 68 (12) : 3482 - 3490
  • [8] RETRACTED: Preliminary investigation of dual mode CMUT probe for ultrasound image guided HIFU therapy (Retracted Article)
    Bawiec, Christopher R.
    N'Djin, W. Apoutou
    Bouchoux, Guillaume
    Senegond, Nicolas
    Guillen, Nicolas
    Chapelon, Jean-Yves
    [J]. 2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [9] Volumetric ultrasound Imaging using 2-D CMUT arrays
    Oralkan, Ö
    Ergun, AS
    Cheng, CH
    Johnson, JA
    Karaman, M
    Lee, TH
    Khuri-Yakub, BT
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2003, 50 (11) : 1581 - 1594
  • [10] An Efficient VLSI Architecture For 2-D Dual-Mode SMDWT
    Hsia, Chih-Hsien
    Chiang, Jen-Shiun
    Chang, Shih-Hao
    [J]. 2013 10TH IEEE INTERNATIONAL CONFERENCE ON NETWORKING, SENSING AND CONTROL (ICNSC), 2013, : 775 - 779