Validation of mammographic x-ray spectra generated using Particle and Heavy Ion Transport code System

被引:3
|
作者
Chusin, Thunyarat [1 ,2 ]
Matsubara, Kosuke [3 ]
Takemura, Akihiro [3 ]
Okubo, Rena [4 ]
Ogawa, Yoshinori [1 ]
机构
[1] Kanazawa Univ, Dept Quantum Med Technol, Div Hlth Sci, Grad Sch Med Sci, Kanazawa, Ishikawa, Japan
[2] Naresuan Univ, Dept Radiol Technol, Fac Allied Hlth Sci, Phitsanulok, Thailand
[3] Kanazawa Univ, Fac Hlth Sci, Inst Med Pharmaceut & Hlth Sci, Dept Quantum Med Technol, Kanazawa, Ishikawa, Japan
[4] Daiyukai Hlth Syst, Dept Radiol Technol, Ichinomiya, Japan
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2020年 / 65卷 / 06期
关键词
x-ray spectra; Monte Carlo simulation; PHITS; breast imaging; mammography; MONTE-CARLO-SIMULATION; DIAGNOSTIC-RADIOLOGY; TUNGSTEN-ANODE; MOLYBDENUM; MODELS; PHITS; RHODIUM;
D O I
10.1088/1361-6560/ab735c
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A Monte Carlo (MC) code is a robust method to generate a mammographic x-ray spectrum because the geometry of a mammography system can be flexible and directly modeled in MC simulation. However, simulations from MC code need to be validated before it can be reliably used for specific applications. This study aimed to generate and validate the x-ray spectra of relevant anodes used in mammography and breast tomosynthesis using Particle and Heavy Ion Transport code System (PHITS). PHITS version 3.08 was used to generate the x-ray spectra of molybdenum (Mo), rhodium (Rh), and tungsten (W) anodes. The Mo anode spectrum derived using PHITS was compared with those obtained using other MC codes. The generated spectra of all anodes were compared with the literature. Parameters including spectral shape, K characteristic x-ray yield, heel effect, and half-value layer (HVL) were used for a comparative assessment. The differences in these assessment parameters conducted by PHITS and PHITSEGS5 simulations were studied. Regarding the comparative parameters, PHITSEGS5 simulation improved the accuracy of mammographic x-ray generation compared to PHITS simulation; K x-ray and bremsstrahlung yields of the Mo anode spectrum generated by PHITSEGS5 simulation were a better agreement with those generated by other MC code simulations. The PHITSEGS5 spectra overestimated K x-ray and low-energy bremsstrahlung photons in comparison with measured spectra. Subsequently, HVLs calculated from PHITSEGS5 spectra were 1.0% (Mo/Mo) and 7.0% (W/Al) lower than those derived from measured spectra. For Mo and Rh anodes, relative difference of HVLs calculated from PHITSEGS5 spectra and those obtained from literature and measurement were within the TRS 457 acceptance criteria (+/- 0.02 mm Al). The observed difference exceeded the acceptance criteria for W anode. Regarding existed consistency in HVL between simulation and measurement, PHITSEGS5 simulation can be reliably used to generate x-ray spectra of Mo and Rh anodes. However, its accuracy should be improved for generating W anode spectrum.
引用
收藏
页数:13
相关论文
共 50 条
  • [21] X-RAY ABSORPTION SPECTRA OF ION CRYSTALS
    NEFEDOV, VI
    ZHURNAL STRUKTURNOI KHIMII, 1970, 11 (02): : 299 - &
  • [22] Simulations of an accelerator-based shielding experiment using the particle and heavy-ion transport code system PHITS
    Sato, T
    Sihver, L
    Iwase, H
    Nakashima, H
    Niita, K
    SPACE LIFE SCIENCES: GROUND-BASED IRON-ION BIOLOGY AND PHYSICS, INCLUDING SHIELDING, 2005, 35 (02): : 208 - 213
  • [23] Very soft x-ray spectra of heavy elements
    del Rosario, C
    PHYSICAL REVIEW, 1932, 41 (02): : 136 - 140
  • [24] ISOTOPE SHIFT IN THE X-RAY SPECTRA OF HEAVY ELEMENTS
    WERTHEIM, MS
    IGO, G
    PHYSICAL REVIEW, 1955, 98 (01): : 1 - 5
  • [25] ISOTOPE SHIFT IN X-RAY SPECTRA OF HEAVY ELEMENTS
    IGO, G
    WERTHEIM, MS
    PHYSICAL REVIEW, 1954, 95 (04): : 1097 - 1097
  • [26] LASER-BASED ULTRAMICROFOCUSED MAMMOGRAPHIC X-RAY SYSTEM
    KROL, A
    KIEFFER, JC
    BASSANO, DA
    CHAMBERLAIN, CC
    IKHLEF, A
    PRASAD, SC
    RADIOLOGY, 1995, 197 : 357 - 357
  • [27] Portable CdTe detection system for mammographic X-ray spectroscopy
    La Manna, A.
    Abbene, L.
    Fauci, F.
    Gerardi, G.
    Raso, G.
    Del Sordo, S.
    Barone Tonghi, L.
    Stumbo, S.
    NUOVO CIMENTO DELLA SOCIETA ITALIANA DI FISICA C-COLLOQUIA ON PHYSICS, 2006, 29 (03): : 361 - 367
  • [28] A Comparison of Mammographic X-ray Spectra: Simulation with EGSnrc and Experiment with CdTe Detector
    Nigapruke, Khatayut
    Puwanich, Patana
    Phaisangittisakul, Nakorn
    Youngdee, Wiwat
    JOURNAL OF RADIATION RESEARCH, 2009, 50 (06) : 507 - 512
  • [29] Benchmarking the MEKAL spectral code with solar X-ray spectra
    Phillips, KJH
    Mewe, R
    Harra-Murnion, LK
    Kaastra, JS
    Beiersdorfer, P
    Brown, GV
    Liedahl, DA
    ASTRONOMY & ASTROPHYSICS SUPPLEMENT SERIES, 1999, 138 (02): : 381 - 393
  • [30] Simulation of x-ray spectra generated by electron impact on solids
    Acosta, E
    Llovet, X
    Coleoni, E
    Salvat, F
    Riveros, JA
    X-RAY SPECTROMETRY, 1999, 28 (02) : 121 - 127