Influence of circumferential annular grooving design of impeller on suspended fluid force of axial flow blood pump

被引:2
|
作者
Wang, Liang [1 ,2 ]
Yun, Zhong [1 ]
Tang, Xiaoyan [1 ]
Xiang, Chuang [2 ]
机构
[1] Cent South Univ, Sch Mech & Elect Engn, 932 South Lushan Rd, Changsha 410083, Hunan, Peoples R China
[2] Hunan Univ Arts & Sci, Coll Mech Engn, Changde, Hunan, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
Axial flow blood pump; impeller; circumferential groove; depth; suspension;
D O I
10.1177/03913988211064912
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aiming at insufficient suspension force on the impeller when the hydraulic suspension axial flow blood pump is start at low speed, the impeller suspension stability is poor, and can't quickly enter the suspended working state. By establishing the mathematical model of the suspension force on the impeller, then the influence of the circumferential groove depth of the impeller on the suspension force is analyzed, and the annular groove depth on the impeller blade in the direction of fluid inlet and outlet was determined as (0.26, 0.02 mm). When the blood pump starts, there is an eccentricity between the impeller and the pump tube, the relationship between the suspension force and the speed of the impeller under different eccentricities is analyzed. Combined with the prototype experiment, the circumferential annular grooving design of the impeller can make the blood pump rotate at about 3500 rpm into the suspension state, when the impeller is at 8000 rpm, the impeller can basically achieve stable suspension at the eccentricity of 0.1 mm in the gravity direction, indicating that the reasonable circumferential annular grooving design of the impeller can effectively improve the suspension hydraulic force of the impeller and improve the stability of the hydraulic suspension axial flow blood pump.
引用
收藏
页码:360 / 370
页数:11
相关论文
共 50 条
  • [1] Influence of impeller design on hemolysis of an axial blood pump
    Liu, Lei
    Wang, Fangqun
    Wu, Qinlin
    Wu, Wenjue
    Qian, Kunxi
    Ji, Jinghua
    Zhao, Wenxiang
    He, Weixing
    Li, Tianbo
    MEDICAL MATERIALS AND ENGINEERING, 2012, 140 : 162 - 166
  • [2] Influence of rotor impeller structure on performance improvement of suspended axial flow blood pumps
    Wang, Liang
    Yun, Zhong
    Tang, Xiaoyan
    Xiang, Chuang
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2024, 47 (03): : 162 - 172
  • [3] Effect of Floating Impeller on the Inner Flow Characteristics and Axial Force in Regenerative Flow Pump
    Li, Qianqian
    Tang, Deli
    Ge, Jian
    Lu, Yi
    Bie, Fengfeng
    Zhu, Xiaoqu
    Jixie Gongcheng Xuebao/Journal of Mechanical Engineering, 2024, 60 (14): : 378 - 386
  • [4] Development of rear-impeller axial flow blood pump for realization of axial flow blood pump installed at aortic valve position
    Okamoto, Eiji
    Yano, Tetsuya
    Inoue, Yusuke
    Shiraishi, Yasuyuki
    Yambe, Tomoyuki
    Mitamura, Yoshinori
    ARTIFICIAL ORGANS, 2019, 43 (09) : 828 - 833
  • [5] Analysis of liquid flow and axial force calculation in axial clearance for floating impeller of centrifugal pump
    Liu, Zailun
    Xu, Lizhong
    Jia, Xiao
    Wu, Jiao
    Wang, Dongwei
    Nongye Gongcheng Xuebao/Transactions of the Chinese Society of Agricultural Engineering, 2013, 29 (12): : 79 - 85
  • [6] Discussion about design of front guide impeller of axial flow pump
    Li, Tong-Zhuo
    Chen, Jian
    Lu, Hong-Qi
    Xiang, Qing-Jiang
    Hedongli Gongcheng/Nuclear Power Engineering, 2004, 25 (03): : 270 - 274
  • [7] Hydraulic design of axial flow pump impeller by integral equation method
    Tsinghua Univ, Beijing, China
    Qinghua Daxue Xuebao, 8 (60-65):
  • [8] Development of a centrifugal blood pump with magnetically suspended impeller and the related fluid mechanical problems
    T Akamatsu
    T Tsukiya
    Sadhana, 1998, 23 : 597 - 603
  • [9] Development of a centrifugal blood pump with magnetically suspended impeller and the related fluid mechanical problems
    Akamatsu, T
    Tsukiya, T
    SADHANA-ACADEMY PROCEEDINGS IN ENGINEERING SCIENCES, 1998, 23 (5-6): : 597 - 603
  • [10] Hydrodynamic Optimization for Design of a Submersible Axial-Flow Pump with a Swept Impeller
    Kim, Youn-Sung
    Heo, Man-Woong
    Shim, Hyeon-Seok
    Lee, Bong-Soo
    Kim, Dong-Hwan
    Kim, Kwang-Yong
    ENERGIES, 2020, 13 (12)