Influence of rotor impeller structure on performance improvement of suspended axial flow blood pumps

被引:0
|
作者
Wang, Liang [1 ,2 ]
Yun, Zhong [1 ,3 ]
Tang, Xiaoyan [1 ]
Xiang, Chuang [2 ]
机构
[1] Cent South Univ, Sch Mech & Elect Engn, Changsha, Peoples R China
[2] Hunan Univ Arts & Sci, Coll Mech Engn, Changde, Peoples R China
[3] Cent South Univ, Sch Mech & Elect Engn, 932 South Lushan Rd, Changsha 410083, Hunan, Peoples R China
来源
关键词
Axial flow blood pump; impeller hub taper; number of blades; hemolysis; performance improvement; SHEAR-STRESS; HEMOLYSIS; DAMAGE; PREDICTION;
D O I
10.1177/03913988231225128
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: The hydrodynamic suspension structure design of the axial blood pump impeller can avoid the problems associated with using mechanical bearings. However, the particular impeller structure will impact the hydraulic performance and hemolysis of the blood pump.Method: This article combines computational fluid dynamics (CFD) with the Lagrange particle tracking method, aiming to improve the blood pump's hydraulic and hemolysis performance. It analyzes the flow characteristics and hemolysis performance inside the pump. It optimizes the taper of the impeller hub, the number of blades, and the inclination angle of the circumferential groove at the top of the blade.Results: Under certain rotational speed conditions, an increase in the taper of the impeller hub or the number of blades can increase the pumping pressure of a blood pump, but an increase in the number of blades will reduce the flow rate. The design of circumferential grooves at the top of the blade can increase the pumping pressure to a certain extent, with little impact on the hemolysis performance. The impeller structure is optimized based on the estimated hemolysis of each impeller model blood pump. It could be seen that when the pump blood pressure and flow rate were reached, the optimized impeller speed was reduced by 11.4%, and the estimated hemolysis value was reduced by 10.5%.Conclusion: In this paper, the rotor impeller structure of the blood pump was optimized to improve the hydraulic and hemolytic performance effectively, which can provide a reference for the related research of the axial flow blood pump using hydraulic suspension.
引用
收藏
页码:162 / 172
页数:11
相关论文
共 50 条
  • [1] Influence of circumferential annular grooving design of impeller on suspended fluid force of axial flow blood pump
    Wang, Liang
    Yun, Zhong
    Tang, Xiaoyan
    Xiang, Chuang
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2022, 45 (04): : 360 - 370
  • [2] Axial flow blood pumps
    Song, XW
    Throckmorton, AL
    Untaroiu, A
    Patel, S
    Allaire, PE
    Wood, HG
    Olsen, DB
    ASAIO JOURNAL, 2003, 49 (04) : 355 - 364
  • [3] Influence of Impeller Diameter on the Performance of Centrifugal pumps
    Matlakala, M. E.
    Kallon, D. V. V.
    Mogapi, K. E.
    Mabelane, I. M.
    Makgopa, D. M.
    CONFERENCE OF THE SOUTH AFRICAN ADVANCED MATERIALS INITIATIVE (COSAAMI 2019), 2019, 655
  • [4] Influence of rotor eccentricity on the aerodynamic performance of an axial flow turbine
    Kang, S. (kangs@ncepu.edu.cn), 1600, Science Press (34):
  • [5] Impeller inlet geometry effect on performance Improvement for centrifugal pumps
    Luo, Xianwu
    Zhang, Yao
    Peng, Junqi
    Xu, Hongyuan
    Yu, Weiping
    JOURNAL OF MECHANICAL SCIENCE AND TECHNOLOGY, 2008, 22 (10) : 1971 - 1976
  • [6] Impeller inlet geometry effect on performance improvement for centrifugal pumps
    Xianwu Luo
    Yao Zhang
    Junqi Peng
    Hongyuan Xu
    Weiping Yu
    Journal of Mechanical Science and Technology, 2008, 22 : 1971 - 1976
  • [7] Influence of tip clearance inclination on the performance of axial flow impeller in storage mode
    Zhu, Di
    Jin, Faye
    Hu, Zilong
    Luo, Yongyao
    Tao, Ran
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART A-JOURNAL OF POWER AND ENERGY, 2023, 237 (04) : 831 - 842
  • [8] Hydrodynamic Analysis of the Flow in an Axial Rotor and Impeller for Large Storage Pump
    Bosioc, A. I.
    Muntean, S.
    Draghici, I.
    Anton, L. E.
    28TH IAHR SYMPOSIUM ON HYDRAULIC MACHINERY AND SYSTEMS (IAHR2016), PTS 1-12, 2016, 49
  • [9] Influence of Impeller Structure Parameters on the Hydraulic Performance and Casting Molding of Spiral Centrifugal Pumps
    Wang, Chao
    Luo, Yin
    Li, Zihan
    Shen, Zhenhua
    Ye, Daoxing
    WATER, 2024, 16 (11)
  • [10] Influence of impeller top clearance structure on the wear characteristics of the impeller of an axial flow lead-bismuth pump
    Chang J.
    Li Y.
    Ma W.
    Zhang R.
    Liu X.
    Niu T.
    He Jishu/Nuclear Techniques, 2023, 46 (10):