Influence of rotor impeller structure on performance improvement of suspended axial flow blood pumps

被引:0
|
作者
Wang, Liang [1 ,2 ]
Yun, Zhong [1 ,3 ]
Tang, Xiaoyan [1 ]
Xiang, Chuang [2 ]
机构
[1] Cent South Univ, Sch Mech & Elect Engn, Changsha, Peoples R China
[2] Hunan Univ Arts & Sci, Coll Mech Engn, Changde, Peoples R China
[3] Cent South Univ, Sch Mech & Elect Engn, 932 South Lushan Rd, Changsha 410083, Hunan, Peoples R China
来源
关键词
Axial flow blood pump; impeller hub taper; number of blades; hemolysis; performance improvement; SHEAR-STRESS; HEMOLYSIS; DAMAGE; PREDICTION;
D O I
10.1177/03913988231225128
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: The hydrodynamic suspension structure design of the axial blood pump impeller can avoid the problems associated with using mechanical bearings. However, the particular impeller structure will impact the hydraulic performance and hemolysis of the blood pump.Method: This article combines computational fluid dynamics (CFD) with the Lagrange particle tracking method, aiming to improve the blood pump's hydraulic and hemolysis performance. It analyzes the flow characteristics and hemolysis performance inside the pump. It optimizes the taper of the impeller hub, the number of blades, and the inclination angle of the circumferential groove at the top of the blade.Results: Under certain rotational speed conditions, an increase in the taper of the impeller hub or the number of blades can increase the pumping pressure of a blood pump, but an increase in the number of blades will reduce the flow rate. The design of circumferential grooves at the top of the blade can increase the pumping pressure to a certain extent, with little impact on the hemolysis performance. The impeller structure is optimized based on the estimated hemolysis of each impeller model blood pump. It could be seen that when the pump blood pressure and flow rate were reached, the optimized impeller speed was reduced by 11.4%, and the estimated hemolysis value was reduced by 10.5%.Conclusion: In this paper, the rotor impeller structure of the blood pump was optimized to improve the hydraulic and hemolytic performance effectively, which can provide a reference for the related research of the axial flow blood pump using hydraulic suspension.
引用
收藏
页码:162 / 172
页数:11
相关论文
共 50 条
  • [31] Axial flow artificial heart blood pumps: A brief review
    Kafagy, Dhyaa
    Gitano-Briggs, Horizon
    Trends in Biomaterials and Artificial Organs, 2013, 27 (03): : 124 - 130
  • [32] Influence of Double-layer Flow Passage Structure after Impeller on External Characteristics of Axial-flow Pump
    Zhou Y.
    Chen B.
    Zhang D.
    Zhang H.
    Yang C.
    Nongye Jixie Xuebao/Transactions of the Chinese Society for Agricultural Machinery, 2022, 53 (02): : 149 - 157and176
  • [33] Flow performance analysis in an axial compressor rotor with splitter
    Yan, Ming
    Chen, Mao-Zhang
    Tuijin Jishu/Journal of Propulsion Technology, 2002, 23 (04): : 280 - 282
  • [34] Performance simulation of a radial flow type impeller of centrifugal pumps using CFD
    Lopez, R.
    Vaca, M.
    Terres, H.
    Lizardi, A.
    Chavez, S.
    Garcia, M.
    VIII INTERNATIONAL CONGRESS OF ENGINEERING PHYSICS, 2017, 792
  • [35] Experimental investigation of the influence of the hydraulic performance of an axial blood pump on intraventricular blood flow
    Liu, Guang-Mao
    Jiang, Fu-Qing
    Yang, Xiao-Han
    Wei, Run-Jie
    Hu, Sheng-Shou
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2021, 44 (12): : 980 - 989
  • [36] Development of rear-impeller axial flow blood pump for realization of axial flow blood pump installed at aortic valve position
    Okamoto, Eiji
    Yano, Tetsuya
    Inoue, Yusuke
    Shiraishi, Yasuyuki
    Yambe, Tomoyuki
    Mitamura, Yoshinori
    ARTIFICIAL ORGANS, 2019, 43 (09) : 828 - 833
  • [37] Visualization of flow structure in a centrifugal blood pump impeller
    Tsukiya T.
    Taenaka Y.
    Tatsumi E.
    Takano H.
    Journal of Artificial Organs, 2002, 5 (1) : 13 - 17
  • [38] Numerical investigation of impeller trimming effect on performance of an axial flow fan
    Li, Chunxi
    Li, Xinying
    Li, Pengmin
    Ye, Xuemin
    ENERGY, 2014, 75 : 534 - 548
  • [39] INFLUENCE OF THE SUCTION SIDE DIFFUSER GEOMETRY ON THE PERFORMANCE OF RADIAL IMPELLER PUMPS.
    Amann, P.
    Hergt, P.
    Jaberg, H.
    KSB Technische Berichte, 1988, (23 e): : 3 - 8
  • [40] HEMOLYSIS ANALYSIS OF AXIAL BLOOD PUMPS WITH VARIOUS STRUCTURE IMPELLERS
    Chen, Zengsheng
    Yao, Zhaohui
    Zhu, Lailai
    Zhang, Xiwen
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2013, 13 (04)