Visualization of flow structure in a centrifugal blood pump impeller

被引:2
|
作者
Tsukiya T. [1 ]
Taenaka Y. [1 ]
Tatsumi E. [1 ]
Takano H. [1 ]
机构
[1] Department of Artificial Organs, National Cardiovascular Center, Fujishiro-dai, Suita
关键词
Artificial heart; Centrifugal pump; Flow visualization; Oil film method; Shear stress;
D O I
10.1007/s100470200002
中图分类号
学科分类号
摘要
The flow in the impeller of a centrifugal blood pump developed at the National Cardiovascular Center of Japan was clearly visualized by the oil film method to illustrate the overall flow structure inside the open-type impeller. The results showed that the main stream went along the pressure surface with a backward flow at the inlet. Large-scale vortices were formed near both tips of the suction side, and a low-velocity area was found in the middle of the suction side that could be eliminated by increasing the flow rate. These findings were helpful for improving the design of the impeller to enhance the hydraulic performance and the antithrombogenicity of the pump.
引用
收藏
页码:13 / 17
页数:4
相关论文
共 50 条
  • [1] Visualization study of the transient flow in the centrifugal blood pump impeller
    Tsukiya, T
    Taenaka, Y
    Tatsumi, E
    Takano, H
    [J]. ASAIO JOURNAL, 2002, 48 (04) : 431 - 436
  • [2] Flow visualization analysis of relative stagnation areas on the impeller surface of a centrifugal blood pump
    Aouidef A.
    Yamane T.
    Nishida M.
    Toyoda M.
    Tsukamoto Y.
    Ito K.
    Konishi Y.
    Masuzawa T.
    Tsukiya T.
    Taenaka Y.
    [J]. Journal of Artificial Organs, 2001, 4 (4) : 262 - 268
  • [3] Flow visualization measurement for shear velocity distribution in the impeller-casing gap of a centrifugal blood pump
    Yamane, T
    Clarke, H
    Asztalos, B
    Nishida, M
    Kobayashi, T
    [J]. JSME INTERNATIONAL JOURNAL SERIES C-MECHANICAL SYSTEMS MACHINE ELEMENTS AND MANUFACTURING, 1999, 42 (03) : 621 - 627
  • [4] Flow visualization study to investigate the secondary flow behind the impeller in the gyro centrifugal pump
    Ichikawa, S
    Nonaka, K
    Linneweber, J
    Kawahito, S
    Motomura, T
    Ishitoya, H
    Mikami, M
    Nishimura, I
    Glueck, J
    Shinohara, T
    Nosé, Y
    [J]. ARTIFICIAL ORGANS, 2002, 26 (12) : 1050 - 1052
  • [5] Visualization of relative flow patterns in centrifugal blood pump
    Chan, WK
    Yu, SCM
    Chua, LP
    Wong, YW
    [J]. KSME INTERNATIONAL JOURNAL, 2001, 15 (12): : 1869 - 1875
  • [6] Visualization of relative flow patterns in centrifugal blood pump
    Weng Kong Chan
    S. C. M Yu
    L. P. Chua
    Y. W. Wong
    [J]. KSME International Journal, 2001, 15 : 1869 - 1875
  • [7] A FLOW VISUALIZATION STUDY OF THE NCVC CENTRIFUGAL BLOOD PUMP
    ARAKI, K
    TAENAKA, Y
    MASUZAWA, T
    TATSUMI, E
    WAKISAKA, Y
    WATARI, M
    NAKATANI, T
    AKAGI, H
    BABA, Y
    ANAI, H
    PARK, YH
    EYA, K
    TODA, K
    TAKANO, H
    [J]. ARTIFICIAL ORGANS, 1994, 18 (09) : 669 - 672
  • [8] The flow patterns within the impeller passages of a centrifugal blood pump model
    Yu, SCM
    Ng, BTH
    Chan, WK
    Chua, LP
    [J]. MEDICAL ENGINEERING & PHYSICS, 2000, 22 (06) : 381 - 393
  • [9] Visualization of gas-liquid flow pattern in a centrifugal pump impeller and its influence on the pump performance
    Zhao, Lin
    Chang, Zhuang
    Zhang, Zhenduo
    Huang, Rui
    He, Denghui
    [J]. Measurement: Sensors, 2021, 13
  • [10] Impeller design for a miniaturized centrifugal blood pump
    Takano, T
    Schulte-Eistrup, S
    Yoshikawa, M
    Nakata, K
    Kawahito, S
    Maeda, T
    Nonaka, K
    Linneweber, J
    Glueck, J
    Fujisawa, A
    Makinouchi, K
    Yokokawa, M
    Nosé, Y
    [J]. ARTIFICIAL ORGANS, 2000, 24 (10) : 821 - 825