Influence of circumferential annular grooving design of impeller on suspended fluid force of axial flow blood pump

被引:2
|
作者
Wang, Liang [1 ,2 ]
Yun, Zhong [1 ]
Tang, Xiaoyan [1 ]
Xiang, Chuang [2 ]
机构
[1] Cent South Univ, Sch Mech & Elect Engn, 932 South Lushan Rd, Changsha 410083, Hunan, Peoples R China
[2] Hunan Univ Arts & Sci, Coll Mech Engn, Changde, Hunan, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
Axial flow blood pump; impeller; circumferential groove; depth; suspension;
D O I
10.1177/03913988211064912
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aiming at insufficient suspension force on the impeller when the hydraulic suspension axial flow blood pump is start at low speed, the impeller suspension stability is poor, and can't quickly enter the suspended working state. By establishing the mathematical model of the suspension force on the impeller, then the influence of the circumferential groove depth of the impeller on the suspension force is analyzed, and the annular groove depth on the impeller blade in the direction of fluid inlet and outlet was determined as (0.26, 0.02 mm). When the blood pump starts, there is an eccentricity between the impeller and the pump tube, the relationship between the suspension force and the speed of the impeller under different eccentricities is analyzed. Combined with the prototype experiment, the circumferential annular grooving design of the impeller can make the blood pump rotate at about 3500 rpm into the suspension state, when the impeller is at 8000 rpm, the impeller can basically achieve stable suspension at the eccentricity of 0.1 mm in the gravity direction, indicating that the reasonable circumferential annular grooving design of the impeller can effectively improve the suspension hydraulic force of the impeller and improve the stability of the hydraulic suspension axial flow blood pump.
引用
收藏
页码:360 / 370
页数:11
相关论文
共 50 条
  • [41] Fluid-structure coupling analysis of impeller in unstable region for a reversible axial-flow pump device
    Meng, Fan
    Yuan, Shouqi
    Li, Yanjun
    ADVANCES IN MECHANICAL ENGINEERING, 2018, 10 (03)
  • [42] An advanced quasi-three-dimensional inverse computation model for axial flow pump impeller design
    Peng, G
    Fujikawa, S
    Cao, SL
    PROCEEDINGS OF THE XIX IAHR SYMPOSIUM ON HYDRAULIC MACHINERY AND CAVITATION, VOLS 1 AND 2, 1998, : 722 - 733
  • [43] Influence of Double-layer Flow Passage Structure after Impeller on External Characteristics of Axial-flow Pump
    Zhou Y.
    Chen B.
    Zhang D.
    Zhang H.
    Yang C.
    Nongye Jixie Xuebao/Transactions of the Chinese Society for Agricultural Machinery, 2022, 53 (02): : 149 - 157and176
  • [44] INFLUENCE OF IMPELLER'S GEOMETRIC PARAMETERS ON REFRIGERANT FLOW AND THERMAL PERFORMANCE OF A WORKING FLUID PUMP
    Zheng, Huifan
    Ma, Yuanqing
    Wang, Yongchang
    Yin, Yonggao
    Dai, Yanjun
    Zhang, Xuelai
    Xin, Xin
    THERMAL SCIENCE, 2020, 24 (04): : 2393 - 2399
  • [45] Experimental investigation of the influence of the hydraulic performance of an axial blood pump on intraventricular blood flow
    Liu, Guang-Mao
    Jiang, Fu-Qing
    Yang, Xiao-Han
    Wei, Run-Jie
    Hu, Sheng-Shou
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2021, 44 (12): : 980 - 989
  • [46] A new design and computational fluid dynamics study of an implantable axial blood pump
    Mojtaba Koochaki
    Hanieh Niroomand-Oscuii
    Australasian Physical & Engineering Sciences in Medicine, 2013, 36 : 417 - 422
  • [47] A new design and computational fluid dynamics study of an implantable axial blood pump
    Koochaki, Mojtaba
    Niroomand-Oscuii, Hanieh
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2013, 36 (04) : 417 - 422
  • [48] Applications of CFD Technique in the Design and Flow Analysis of Implantable Axial Flow Blood Pump
    Hui-min Fan
    Fang-wen Hong
    Guo-ping Zhang
    Liang Ye
    Zhong-Min Liu
    Journal of Hydrodynamics, 2010, 22 : 518 - 525
  • [49] APPLICATIONS OF CFD TECHNIQUE IN THE DESIGN AND FLOW ANALYSIS OF IMPLANTABLE AXIAL FLOW BLOOD PUMP
    Fan Hui-min
    Hong Fang-wen
    Zhang Guo-ping
    Ye Liang
    Liu Zhong-min
    JOURNAL OF HYDRODYNAMICS, 2010, 22 (04) : 518 - 525
  • [50] APPLICATIONS OF CFD TECHNIQUE IN THE DESIGN AND FLOW ANALYSIS OF IMPLANTABLE AXIAL FLOW BLOOD PUMP
    FAN Hui-min Shanghai East Hospital Affiliated to Tongji University
    JournalofHydrodynamics, 2010, 22 (04) : 518 - 525