Influence of circumferential annular grooving design of impeller on suspended fluid force of axial flow blood pump

被引:2
|
作者
Wang, Liang [1 ,2 ]
Yun, Zhong [1 ]
Tang, Xiaoyan [1 ]
Xiang, Chuang [2 ]
机构
[1] Cent South Univ, Sch Mech & Elect Engn, 932 South Lushan Rd, Changsha 410083, Hunan, Peoples R China
[2] Hunan Univ Arts & Sci, Coll Mech Engn, Changde, Hunan, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
Axial flow blood pump; impeller; circumferential groove; depth; suspension;
D O I
10.1177/03913988211064912
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aiming at insufficient suspension force on the impeller when the hydraulic suspension axial flow blood pump is start at low speed, the impeller suspension stability is poor, and can't quickly enter the suspended working state. By establishing the mathematical model of the suspension force on the impeller, then the influence of the circumferential groove depth of the impeller on the suspension force is analyzed, and the annular groove depth on the impeller blade in the direction of fluid inlet and outlet was determined as (0.26, 0.02 mm). When the blood pump starts, there is an eccentricity between the impeller and the pump tube, the relationship between the suspension force and the speed of the impeller under different eccentricities is analyzed. Combined with the prototype experiment, the circumferential annular grooving design of the impeller can make the blood pump rotate at about 3500 rpm into the suspension state, when the impeller is at 8000 rpm, the impeller can basically achieve stable suspension at the eccentricity of 0.1 mm in the gravity direction, indicating that the reasonable circumferential annular grooving design of the impeller can effectively improve the suspension hydraulic force of the impeller and improve the stability of the hydraulic suspension axial flow blood pump.
引用
收藏
页码:360 / 370
页数:11
相关论文
共 50 条
  • [21] Effect of guide vane outlet angle on pump performance and impeller radial force in an axial-flow heart pump
    刘复飞
    Sang Xiaohu
    HighTechnologyLetters, 2018, 24 (02) : 219 - 226
  • [22] Numerical study on a coupled viscous and potential flow design method of axial-flow pump impeller
    Qiu, Ji-Tao
    Li, Ning
    Yang, Chen-Jun
    Cai, You-Lin
    Meng, Kun-Yu
    OCEAN ENGINEERING, 2022, 266
  • [23] Influence of leading-edge cavitation on impeller blade axial force in the pump mode of reversible pump-turbine
    Zhu, Di
    Xiao, Ruofu
    Liu, Weichao
    RENEWABLE ENERGY, 2021, 163 : 939 - 949
  • [24] Relationship between pump speed design and hemolysis in an axial flow blood pump
    Anai, H
    Wakisaka, Y
    Nakatani, T
    Taenaka, Y
    Takano, H
    Hadama, T
    ARTIFICIAL ORGANS, 1996, 20 (06) : 564 - 567
  • [25] Relationship between pump speed design and hemolysis in an axial flow blood pump
    Texas Heart Institute, Houston, TX, United States
    不详
    不详
    不详
    ARTIF. ORGANS, 5 (564-567):
  • [26] Comparison of flow characteristics of enlarged blood pump models with different impeller design
    Chua, LP
    Yu, SCM
    Leo, HL
    Chan, WK
    INTERNATIONAL COMMUNICATIONS IN HEAT AND MASS TRANSFER, 1999, 26 (03) : 369 - 378
  • [27] Development of an axial flow blood pump: Characteristics of a magnetic fluid seal
    Sekine K.
    Asakawa M.
    Mitamura Y.
    Journal of Artificial Organs, 2001, 4 (3) : 245 - 251
  • [28] Fluid dynamics aspects of miniaturized axial-flow blood pump
    Kang, Can
    Huang, Qifeng
    Li, Yunxiao
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2014, 24 (01) : 723 - 729
  • [29] Impeller (straight blade) design variations and their influence on the performance of a centrifugal blood pump
    Fang, Peng
    Du, Jianjun
    Yu, Shunzhou
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2020, 43 (12): : 782 - 795
  • [30] Impeller (straight blade) design variations and their influence on the performance of a centrifugal blood pump
    Fang, Peng
    Du, Jianjun
    Yu, Shunzhou
    International Journal of Artificial Organs, 2020, 43 (12): : 782 - 795