A comparison of electron beam dose calculation accuracy between treatment planning systems using either a pencil beam or a Monte Carlo algorithm

被引:58
|
作者
Ding, GX
Cygler, JE
Yu, CW
Kalach, NI
Daskalov, G
机构
[1] Vanderbilt Univ, Med Ctr, Dept Radiat Oncol, TVC, Nashville, TN 37232 USA
[2] Ottawa Hosp, Reg Canc Ctr, Dept Phys Med, Ottawa, ON, Canada
[3] British Columbia Canc Agcy, Fraser Valley Ctr, Surrey, England
关键词
Monte Carlo; electron pencil beam; radiotherapy; inhomogeneous phantoms; monitor unit calculations;
D O I
10.1016/j.ijrobp.2005.06.016
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To present a comparison of the accuracy of two commercial electron beam treatment planning systems: one uses a Monte Carlo algorithm and the other uses a pencil beam model for dose calculations. Methods and Materials: For the same inhomogeneous phantoms and incident beams, measured dose distributions are compared with those predicted by the commercial treatment planning systems at different source-to-surface distances (SSDs). The accuracy of the pencil beam system for monitor unit calculations is also tested at various SSDs. Beam energies of 6-20 MeV are used. Results: The pencil beam model shows some serious limitations in predicting hot and cold spots in inhomogeneous phantoms for small low- or high-density inhomogeneities, especially for low-energy electron beams, such as 9 MeV. Errors (> 10%) are seen in predicting high- and low-dose variations for three-dimensional inhomogeneous phantoms. The Monte Carlo calculated results generally agree much better with measurements. Conclusions: The accuracy of the pencil beam calculations is difficult to predict because it depends on both the inhomogeneity geometry and location. The pencil beam calculations using CADPLAN result in large errors in phantoms containing three-dimensional type inhomogeneities. The Monte Carlo method in Theraplan Plus dose calculation module is shown to be more robust in accurately predicting dose distributions and monitor units under the tested conditions. (c) 2005 Elsevier Inc.
引用
收藏
页码:622 / 633
页数:12
相关论文
共 50 条
  • [21] USE OF MONTE-CARLO GENERATED PENCIL BEAM DATA FOR ELECTRON TREATMENT PLANNING
    RILEY, R
    MOHAN, R
    CHUI, C
    [J]. MEDICAL PHYSICS, 1985, 12 (04) : 516 - 516
  • [22] Electron fields in clinical application. A comparison of pencil beam and Monte Carlo algorithm
    Treutwein, Marius
    Bogner, Ludwig
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 (08) : 454 - 458
  • [23] PLANNING COMPARISON BETWEEN IPLAN RT DOSE MONTE CARLO AND PENCIL BEAM ALGORITHMS FOR HEAD AND NECK PATIENTS
    Harmsen-Jongerden, R.
    Smink, F.
    Bos, P.
    Petoukhova, A.
    Wiggenraad, R.
    van Santvoort, J.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S585 - S585
  • [24] Comparison of Dose Calculation for Brain, Lung, Pelvis and Abdomen Targets with Monte Carlo Simulation and Eclipse Pencil Beam Algorithm
    Prusator, M.
    Muller, L.
    Ahmad, S.
    Chen, Y.
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E261 - E262
  • [25] A Monte Carlo approach to the validation of a pencil beam algorithm used in treatment planning for static conformal beam radiosurgery
    Belec, J
    Patrocinio, H
    Verhaegen, F
    [J]. MEDICAL PHYSICS, 2004, 31 (06) : 1714 - 1714
  • [26] Quantitative assessment of the accuracy of dose calculation using pencil beam and Monte Carlo algorithms and requirements for clinical quality assurance
    Ali, Imad
    Ahmad, Salahuddin
    [J]. MEDICAL DOSIMETRY, 2013, 38 (03) : 255 - 261
  • [27] Improvements in pencil beam scanning proton therapy dose calculation accuracy in brain tumor cases with a commercial Monte Carlo algorithm
    Widesott, Lamberto
    Lorentini, Stefano
    Fracchiolla, Francesco
    Farace, Paolo
    Schwarz, Marco
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (14):
  • [28] Effect of inhomogeneity in a patient's body on the accuracy of the pencil beam algorithm in comparison to Monte Carlo
    Yamashita, T.
    Akagi, T.
    Aso, T.
    Kimura, A.
    Sasaki, T.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (22): : 7673 - 7688
  • [29] Using electron beam modifiers in the DPM Monte Carlo code for electron beam treatment planning
    Tyagi, N.
    Tien, C.
    Curran, B.
    Fraass, B.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2502 - 2503
  • [30] A virtual-accelerator-based verification of a Monte Carlo dose calculation algorithm for electron beam treatment planning in homogeneous phantoms
    Wieslander, E
    Knöös, T
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (06): : 1533 - 1544