A comparison of electron beam dose calculation accuracy between treatment planning systems using either a pencil beam or a Monte Carlo algorithm

被引:58
|
作者
Ding, GX
Cygler, JE
Yu, CW
Kalach, NI
Daskalov, G
机构
[1] Vanderbilt Univ, Med Ctr, Dept Radiat Oncol, TVC, Nashville, TN 37232 USA
[2] Ottawa Hosp, Reg Canc Ctr, Dept Phys Med, Ottawa, ON, Canada
[3] British Columbia Canc Agcy, Fraser Valley Ctr, Surrey, England
关键词
Monte Carlo; electron pencil beam; radiotherapy; inhomogeneous phantoms; monitor unit calculations;
D O I
10.1016/j.ijrobp.2005.06.016
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To present a comparison of the accuracy of two commercial electron beam treatment planning systems: one uses a Monte Carlo algorithm and the other uses a pencil beam model for dose calculations. Methods and Materials: For the same inhomogeneous phantoms and incident beams, measured dose distributions are compared with those predicted by the commercial treatment planning systems at different source-to-surface distances (SSDs). The accuracy of the pencil beam system for monitor unit calculations is also tested at various SSDs. Beam energies of 6-20 MeV are used. Results: The pencil beam model shows some serious limitations in predicting hot and cold spots in inhomogeneous phantoms for small low- or high-density inhomogeneities, especially for low-energy electron beams, such as 9 MeV. Errors (> 10%) are seen in predicting high- and low-dose variations for three-dimensional inhomogeneous phantoms. The Monte Carlo calculated results generally agree much better with measurements. Conclusions: The accuracy of the pencil beam calculations is difficult to predict because it depends on both the inhomogeneity geometry and location. The pencil beam calculations using CADPLAN result in large errors in phantoms containing three-dimensional type inhomogeneities. The Monte Carlo method in Theraplan Plus dose calculation module is shown to be more robust in accurately predicting dose distributions and monitor units under the tested conditions. (c) 2005 Elsevier Inc.
引用
收藏
页码:622 / 633
页数:12
相关论文
共 50 条
  • [1] SRS Dose Calculation Accuracy Comparison Between Pencil Beam and Monte Carlo Algorithms
    Stathakis, S.
    Defoor, D.
    Saenz, D.
    Kirby, N.
    Mavroidis, P.
    Papanikolaou, N.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3655 - 3655
  • [2] Evaluation of a Commercial Monte Carlo Dose Calculation Algorithm for Electron Beam Treatment Planning
    Chamberland, E.
    Lachance, B.
    Beaulieu, L.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [3] Comparison of Monte Carlo and pencil beam electron planning algorithms
    Perrett, T
    Gray, K
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S92 - S92
  • [4] Dose Comparison between Proton Pencil Beam and Monte Carlo Dose Calculation Algorithm in Lung Cancer Patients
    Maes, D.
    Bowen, S. R.
    Fung, A.
    Saini, J.
    Bloch, C.
    Egan, A.
    Zeng, J.
    Rengan, R.
    Wong, T. P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E694 - E694
  • [5] Pseudo Proton Radiography Beam Validation of Monte Carlo Dose Calculation in Two Pencil Beam Scanning Treatment Planning Systems
    Chang, C.
    Zhou, J.
    Yang, X.
    Dhabaan, A.
    Zhang, R.
    Liu, T.
    McDonald, M.
    Langen, K.
    Lin, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E592 - E592
  • [6] Verification of a proton treatment-planning pencil-beam dose algorithm with Monte Carlo
    Titt, U.
    Zheng, Y.
    Zhu, X.
    Mohan, R.
    Newhauser, W.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2149 - 2149
  • [7] A standardized commissioning framework of Monte Carlo dose calculation algorithms for proton pencil beam scanning treatment planning systems
    Chang, Chih-Wei
    Huang, Sheng
    Harms, Joseph
    Zhou, Jun
    Zhang, Rongxiao
    Dhabaan, Anees
    Slopsema, Roelf
    Kang, Minglei
    Liu, Tian
    McDonald, Mark
    Langen, Katja
    Lin, Liyong
    [J]. MEDICAL PHYSICS, 2020, 47 (04) : 1545 - 1557
  • [8] Comparison of pencil beam algorithm and Monte Carlo dose calculation for proton therapy of paranasal sinus cancer
    Jiang, H
    Adams, J
    Rosenthal, S
    Kollipara, S
    Paganetti, H
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2028 - 2028
  • [9] Proton dose calculation using Monte-Carlo-validated pencil beam database for KonRad treatment planning system
    Trofimov, A
    Knopf, A
    Jiang, H
    Bortfeld, T
    Paganetti, H
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2030 - 2030
  • [10] Evaluation of the AAA Treatment Planning Algorithm for SBRT Lung Treatment: Comparison with Monte Carlo and Homogeneous Pencil Beam Dose Calculations
    Gete, Ermias
    Teke, Tony
    Kwa, William
    [J]. JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2012, 43 (01) : 26 - 33