Performance evaluation of a new small-volume coagulation monitor - The SmartCheck INR system

被引:4
|
作者
Gardiner, Chris [1 ]
Longair, Ian [1 ]
Hills, Jennifer [1 ]
Cohen, Hannah [1 ]
Mackie, Ian J. [2 ]
Machin, Samuel J. [1 ]
机构
[1] UCL Hosp, Dept Haematol, London W1T 4EU, England
[2] UCL, London, England
关键词
coagulation; anticoagulation; point of care; international normalized ratio;
D O I
10.1309/K87FWL7UA8H8ANHR
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The SmartCheck INR (Unipath, Bedford, England) is a point-of-care device for professional and patient self-monitoring of oral anticoagulant therapy. It measures the international normalized ratio (INR) and assesses test strip integrity, temperature, and sample quality. No significant differences were found in SmartCheck INR results from 3 different instruments or in 3 test-strip lots. Within run imprecision was 0.89% and 6.36% for low and high control samples, respectively (mean INR, 0.99 and 4.08, respectively). Comparability was assessed in 68 patients receiving warfarin by using PTHS Plus (Instrumentation Laboratory, Lexington, MA) and Innovin (Dade Behring, Marburg, Germany) thromboplastins. Good correlations were observed between the methods (r = 0.89 and r = 0.90, respectively) with no significant differences in means: SmartCheck INR, 2.82; Innovin, 2.75; PTHS Plus, 2.74. Clinical agreement with laboratory methods was 88% for Innovin and 97% for PTHS Plus. The SmartCheck INR was easy to use with no mechanical failures during the evaluation and a low test-strip failure rate of 4.7%. The SmartCheck INR provides accurate and reproducible results and is suitable for routine monitoring of oral anticoagulant therapy in the majority of patients.
引用
收藏
页码:500 / 504
页数:5
相关论文
共 50 条
  • [21] THE RECIRCULATING SCREW MIXER - A NEW SMALL-VOLUME MIXER FOR THE POLYMER LABORATORY
    SCOTT, CE
    MACOSKO, CW
    POLYMER ENGINEERING AND SCIENCE, 1993, 33 (16): : 1065 - 1078
  • [22] Design and Evaluation of a New Bladder Volume Monitor
    Wang, Jianhuo
    Hou, Chunlin
    Zheng, Xianyou
    Zhang, Wei
    Chen, Aimin
    Xu, Zhen
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2009, 90 (11): : 1944 - 1947
  • [23] Evaluation of the clinical effect of small-volume resuscitation on uncontrolled hemorrhagic shock in emergency
    Zhao, Gang
    Wu, Wei
    Feng, Qi-ming
    Sun, Jian
    THERAPEUTICS AND CLINICAL RISK MANAGEMENT, 2017, 13 : 387 - 392
  • [24] Systematic evaluation of the performance of a portable monitor of prothrombin time in patients with supratherapeutic INR levels
    de Paula, E.
    Colella, M. P.
    Fiusa, M. M. L.
    Machado, T. F. G. S.
    Custodio, S.
    Orsi, F. A.
    Annichino-Bizzacchi, J. M.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 : 448 - 448
  • [25] Development and application of an automated, small-volume chromatography system for resin and condition screening
    Teeters, Mark
    Bezila, Daniel
    Alred, Patricia
    Velayudhan, Ajoy
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2007, 234
  • [26] Testing a new small-volume technique for determining 234Th in seawater
    C. R. Benitez-Nelson
    K. O. Buesseler
    M. Rutgers van der Loeff
    J. Andrews
    L. Ball
    G. Crossin
    M. A. Charette
    Journal of Radioanalytical and Nuclear Chemistry, 2001, 248 : 795 - 799
  • [27] A small-volume PVTX system for broadband spectroscopic calibration of downhole optical sensors
    Jones, Christopher Michael
    Pelletier, Michael T.
    Atkinson, Robert
    Shen, Jing
    Moore, Jeff
    Anders, Jimmy
    Perkins, David L.
    Myrick, Michael L.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2017, 88 (07):
  • [28] NEW SMALL-VOLUME CELL FOR STUDYING INFRARED-SPECTRA OF CHEMISORBED SPECIES
    ERKELENS, J
    WOSTEN, WJ
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1974, 7 (08): : 607 - 608
  • [29] R-mask: a new concept and its application for small-volume production
    Hayano, K
    Hotta, S
    Hasegawa, N
    Hosono, K
    Tanaka, T
    Suko, K
    Sasaki, S
    Mohri, H
    Hoga, M
    Hayashi, N
    24TH ANNUAL BACUS SYMPOSIUM ON PHOTOMASK TECHNOLOGY, PT 1 AND 2, 2004, 5567 : 257 - 268
  • [30] Testing a new small-volume technique for determining 234Th in seawater
    Benitez-Nelson, CR
    Buesseler, KO
    van der Loeff, MR
    Andrews, J
    Ball, L
    Crossin, G
    Charette, MA
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2001, 248 (03) : 795 - 799