Assessment of peri-implant defects at titanium and zirconium dioxide implants by means of periapical radiographs and cone beam computed tomography: An in-vitro examination

被引:19
|
作者
Steiger-Ronay, Valerie [1 ]
Krcmaric, Zvonimir
Schmidlin, Patrick R. [1 ]
Sahrmann, Philipp [1 ]
Wiedemeier, Daniel B. [2 ]
Benic, Goran I. [3 ]
机构
[1] Univ Zurich, Clin Prevent Dent Periodontol & Cariol, Ctr Dent Med, Zurich, Switzerland
[2] Univ Zurich, Stat Serv, Ctr Dent Med, Zurich, Switzerland
[3] Univ Zurich, Clin Fixed & Removable Prosthodont & Dent Mat Sci, Ctr Dent Med, Zurich, Switzerland
关键词
bone; bone defect; computed tomography; cone beam computed tomography; dental implant; digital; periapical radiography; peri-implantitis; radiology; scan; titanium; titanium implant; X-ray; zirconium dioxide; zirconium dioxide implant; DENTAL IMPLANTS; BONE LEVEL; ARTIFACTS; ACCURACY; WORKSHOP; DISEASES; CBCT;
D O I
10.1111/clr.13383
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective To test the accuracy of measurement of interproximal peri-implant bone defects at titanium (Ti) and zirconium dioxide (ZrO2) implants by digital periapical radiography (PR) and cone beam computed tomography (CBCT). Material and methods A total of 18 models, each containing one Ti and one ZrO2 implant, were cast in dental stone. Six models each were allocated to following defect groups: A-no peri-implant defect, B-1 mm width defect, C-1.5 mm width defect. The defect width was measured with a digital sliding caliper. Subsequently, the models were scanned by means of PR and CBCT. Three examiners assessed the defect width on PR and CBCT. Wilcoxon signed-rank test and Wilcoxon rank sum test were applied to detect differences between imaging techniques and implant types. Results For PR, the deviation of the defect width measurement (mm) for groups A, B, and C amounted to 0.01 +/- 0.03, -0.02 +/- 0.06, and -0.00 +/- 0.04 at Ti and 0.05 +/- 0.02, 0.01 +/- 0.03, and 0.09 +/- 0.03 at ZrO2 implants. The corresponding values (mm) for CBCT reached 0.10 +/- 0.11, 0.26 +/- 0.05, and 0.24 +/- 0.08 at Ti and 1.07 +/- 0.06, 0.64 +/- 0.37, and 0.54 +/- 0.17 at ZrO2 implants. Except for Ti with defect A, measurements in PR were significantly more accurate in comparison to CBCT (p <= 0.05). Both methods generally yielded more accurate measurements for Ti than for ZrO2. Conclusions The assessment of interproximal peri-implant defect width at Ti and ZrO2 implants was more accurate in PR in comparison to CBCT. Measurements in CBCT always led to an overestimation of the defect width, reaching clinical relevance for ZrO2 implants.
引用
收藏
页码:1195 / 1201
页数:7
相关论文
共 50 条
  • [1] A comparison of cone beam computed tomography and conventional periapical radiography at detecting peri-implant bone defects
    Dave, Meetal
    Davies, Jonathan
    Wilson, Ron
    Palmer, Richard
    CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 (06) : 671 - 678
  • [2] In vitro assessment of artifacts induced by titanium, titanium-zirconium and zirconium dioxide implants in cone-beam computed tomography
    Sancho-Puchades, Manuel
    Haemmerle, Christoph H. F.
    Benic, Goran I.
    CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (10) : 1222 - 1228
  • [3] Peri-implant assessment via cone beam computed tomography and digital periapical radiography: an ex vivo study
    Silveira-Neto, Nicolau
    Flores, Mateus Ericson
    De Carli, Joao Paulo
    Costa, Max Doria
    Matos, Felipe de Souza
    Paranhos, Luiz Renato
    Sandini Linden, Maria Salete
    CLINICS, 2017, 72 (11) : 708 - 713
  • [4] Detection of peri-implant bone defects using cone-beam computed tomography and digital periapical radiography with parallel and oblique projection
    Saberi, Bardia Vadiati
    Khosravifard, Negar
    Ghandari, Farnaz
    Hadinezhad, Arash
    IMAGING SCIENCE IN DENTISTRY, 2019, 49 (04) : 265 - 272
  • [5] Cone beam computed tomography artefacts around dental implants with different materials influencing the detection of peri-implant bone defects
    Schriber, Martina
    Yeung, Andy Wai Kan
    Suter, Valerie G. A.
    Buser, Daniel
    Leung, Yiu Yan
    Bornstein, Michael M.
    CLINICAL ORAL IMPLANTS RESEARCH, 2020, 31 (07) : 595 - 606
  • [6] Effectiveness of Periapical Radiography Versus Cone Beam Computed Tomography with Different Kilovoltage Settings in the Detection of Chemically Created Peri-implant Bone Defects: An In Vitro Study
    Pinheiro, Lucas Rodrigues
    Scarfe, William C.
    de Oliveira Sales, Marcelo Augusto
    Gaia, Bruno Felipe
    Gonzalez Cortes, Arthur Rodriguez
    Paraiso Cavalcanti, Marcelo Gusmao
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (04) : 741 - 750
  • [7] Analysis of peri-implant bone defects by using cone beam computed tomography (CBCT): an integrative review
    Costa, J. A.
    Mendes, J. M.
    Salazar, F.
    Pacheco, J. J.
    Rompante, P.
    Camara, M. I.
    ORAL RADIOLOGY, 2023, 39 (03) : 455 - 466
  • [8] Analysis of peri-implant bone defects by using cone beam computed tomography (CBCT): an integrative review
    J. A. Costa
    J. M. Mendes
    F. Salazar
    J. J. Pacheco
    P. Rompante
    M. I. Câmara
    Oral Radiology, 2023, 39 : 455 - 466
  • [9] Evaluation of a metal artifact reduction algorithm and an optimization filter in the estimation of peri-implant dehiscence defects by using cone beam computed tomography: an in-vitro study
    Bayrak, Seval
    Orhan, Kaan
    Cakmak, Emine Sebnem Kursun
    Gorurgoz, Cansu
    Odabasi, Onur
    Yilmaz, Dervis
    Atakan, Cemal
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2020, 130 (02): : 209 - 216
  • [10] Efficacy of low-dose cone beam computed tomography and metal artifact reduction tool for assessment of peri-implant bone defects: an in vitro study
    Alaa Shafiek Nomier
    Yousria Salah El-Din Gaweesh
    Maha R. Taalab
    Shaimaa Abu El Sadat
    BMC Oral Health, 22