Influence of the Realistic Artery Geometry Parameters on a Coronary Stent Fatigue Life

被引:4
|
作者
Cui, Xinyang [1 ]
Ren, Qingshuai [1 ,2 ]
Li, Gaoyang [1 ]
Li, Zihao [1 ]
Qiao, Aike [1 ]
机构
[1] Beijing Univ Technol, Coll Life Sci & Bioengn, 100 Pingleyuan, Beijing, Peoples R China
[2] Chinese Acad Sci, Inst Mech, 15 BeiSiHuan West Rd, Beijing, Peoples R China
基金
北京市自然科学基金;
关键词
Biomechanics; artery geometry parameters; fatigue life prediction; PERIPHERAL STENTS; FRACTURE; PREDICTION; BEHAVIOR; PATIENT; DESIGN;
D O I
10.1142/S0219876218420069
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The stents' adaptability and safety in realistic and idealized stenotic coronary model were compared to investigate the influence of artery geometry parameter on stent fatigue life. The stents' fatigue resistance ability was calculated using Goodman diagram, and the cycle to failure, the fatigue life, and the fatigue safety factor (FSF) were analyzed. Although the peak top of the von Mises stress was located at the bending area of crowns, the stress distributions were different in the two models. Considering the safety and accuracy, it is necessary to use a realistic geometric model to calculate the stent fatigue performance.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Spontaneous coronary artery dissection - "To stent or not to stent that is the question"
    Martins, J. L.
    Afreixo, V.
    Santos, L.
    Viana, J.
    Santos, J.
    EUROPEAN HEART JOURNAL, 2017, 38 : 396 - 397
  • [32] Spontaneous coronary artery dissection: To stent or not to stent? That is the question
    Buccheri, Dario
    Piraino, Davide
    Andolina, Giuseppe
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2016, 215 : 399 - 399
  • [33] Influence of coronary artery bypass surgery on thyroid hormone parameters
    Reinhardt, W
    Mocker, V
    Jockenhovel, F
    Olbricht, T
    Reinwein, D
    Mann, K
    Sadony, V
    HORMONE RESEARCH, 1997, 47 (01) : 1 - 8
  • [34] A new type of aortic valved stent with good stability and no influence on coronary artery
    Jianzhi Cai
    Haitao Huang
    Yongxin Zhou
    Yunqing Mei
    Jie Shao
    Yongwu Wang
    Journal of Cardiothoracic Surgery, 8
  • [35] A new type of aortic valved stent with good stability and no influence on coronary artery
    Cai, Jianzhi
    Huang, Haitao
    Zhou, Yongxin
    Mei, Yunqing
    Shao, Jie
    Wang, Yongwu
    JOURNAL OF CARDIOTHORACIC SURGERY, 2013, 8
  • [36] Influence of machining parameters on fretting fatigue life of Inconel 718
    Infante-Garcia, D.
    Diaz-Alvarez, A.
    Belda, R.
    Diaz-Alvarez, J.
    Cantero, J. L.
    Giner, E.
    Miguelez, M. H.
    INTERNATIONAL JOURNAL OF FATIGUE, 2022, 162
  • [37] Influence of Structural Parameters of Deformable Mirror on Its Fatigue Life
    Luo S.
    Chen L.
    Zhang B.
    Guangxue Xuebao/Acta Optica Sinica, 2017, 37 (10):
  • [38] Coronary Stent Embolization of Peroneal Artery
    Marjot, Thomas
    Lim, Chung S.
    Spannagl, Erwin
    Hussain, Tahir
    VASCULAR AND ENDOVASCULAR SURGERY, 2013, 47 (05) : 400 - 401
  • [39] Coronary artery stent migration in aorta
    de Roux, Quentin
    Provenchere, Sophie
    Longrois, Dan
    de Tymowski, Christia
    ANAESTHESIA CRITICAL CARE & PAIN MEDICINE, 2021, 40 (04)
  • [40] A new type of coronary artery stent
    Evans, A.
    Davies, R. G.
    BRITISH JOURNAL OF ANAESTHESIA, 2009, 102 (03) : 435 - 436