Influence of the Realistic Artery Geometry Parameters on a Coronary Stent Fatigue Life

被引:4
|
作者
Cui, Xinyang [1 ]
Ren, Qingshuai [1 ,2 ]
Li, Gaoyang [1 ]
Li, Zihao [1 ]
Qiao, Aike [1 ]
机构
[1] Beijing Univ Technol, Coll Life Sci & Bioengn, 100 Pingleyuan, Beijing, Peoples R China
[2] Chinese Acad Sci, Inst Mech, 15 BeiSiHuan West Rd, Beijing, Peoples R China
基金
北京市自然科学基金;
关键词
Biomechanics; artery geometry parameters; fatigue life prediction; PERIPHERAL STENTS; FRACTURE; PREDICTION; BEHAVIOR; PATIENT; DESIGN;
D O I
10.1142/S0219876218420069
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The stents' adaptability and safety in realistic and idealized stenotic coronary model were compared to investigate the influence of artery geometry parameter on stent fatigue life. The stents' fatigue resistance ability was calculated using Goodman diagram, and the cycle to failure, the fatigue life, and the fatigue safety factor (FSF) were analyzed. Although the peak top of the von Mises stress was located at the bending area of crowns, the stress distributions were different in the two models. Considering the safety and accuracy, it is necessary to use a realistic geometric model to calculate the stent fatigue performance.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Influence of surface mount lead end geometry on fatigue life
    Gupta, Vineet K.
    Barker, Donald B.
    Journal of Electronic Packaging, Transactions of the ASME, 1994, 116 (02): : 157 - 160
  • [22] Coronary artery stent infection
    Dieter, RS
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 62 (02) : 281 - 281
  • [23] Coronary artery stent infection
    Dieter, RS
    CLINICAL CARDIOLOGY, 2000, 23 (11) : 808 - 810
  • [24] Coronary artery stent embolism
    Lamarre, A
    Lacasse, J
    CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 1999, 50 (02): : 135 - 136
  • [25] Coronary artery stent fracture
    Makaryus, Amgad N.
    Lefkowitz, Lance
    Lee, Alexander D. K.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2007, 23 (03): : 305 - 309
  • [26] Coronary artery stent fracture
    Amgad N. Makaryus
    Lance Lefkowitz
    Alexander D. K. Lee
    The International Journal of Cardiovascular Imaging, 2007, 23 : 305 - 309
  • [27] Double Coronary Artery Stent Fracture With Coronary Artery Microaneurysms
    Lee, Kang Hyu
    Lee, Sang Rok
    Jin, Gong Yong
    Lee, Sun Hwa
    Rhee, Kyoung-Suk
    Chae, Jei Keon
    Kim, Won Ho
    Ko, Jae Ki
    INTERNATIONAL HEART JOURNAL, 2009, 50 (01) : 127 - 132
  • [28] Influence of curvature dynamics on pulsatile coronary artery flow in a realistic bifurcation model
    Prosi, M
    Perktold, K
    Ding, ZH
    Friedman, MH
    JOURNAL OF BIOMECHANICS, 2004, 37 (11) : 1767 - 1775
  • [29] Traumatic coronary artery dissection with total occlusion of left descending coronary artery: to stent or not to stent
    Abellas-Sequeiros, Maria
    Pardo Sanz, Ana
    Sanchez-Recalde, Angel
    Sanmartin-Fernandez, Marcelo
    EUROPEAN HEART JOURNAL, 2021, 42 (45) : 4701 - 4701
  • [30] Spontaneous coronary artery dissection: "To stent or not to stent, that is the question"
    Martins, Jose Luis
    Silva, Joana
    Paiva, Luis
    Costa, Marco
    Goncalves, Lino
    REVISTA PORTUGUESA DE CARDIOLOGIA, 2019, 38 (08) : 609.e1 - 609.e7