A deformable head and neck phantom with in-vivo dosimetry for adaptive radiotherapy quality assurance

被引:28
|
作者
Graves, Yan Jiang [1 ,2 ,3 ]
Smith, Arthur-Allen [4 ]
Mcilvena, David [4 ]
Manilay, Zherrina [4 ]
Lai, Yuet Kong [4 ]
Rice, Roger [1 ,2 ]
Mell, Loren [1 ,2 ]
Jia, Xun [1 ,2 ,5 ]
Jiang, Steve B. [1 ,2 ,5 ]
Cervino, Laura [1 ,2 ]
机构
[1] Univ Calif San Diego, Ctr Adv Radiotherapy Technol, La Jolla, CA 92037 USA
[2] Univ Calif San Diego, Dept Radiat Med & Appl Sci, La Jolla, CA 92037 USA
[3] Univ Calif San Diego, Dept Phys, La Jolla, CA 92093 USA
[4] Univ Calif San Diego, Dept Mech & Aerosp Engn, La Jolla, CA 92093 USA
[5] Univ Texas SW Med Ctr Dallas, Dept Radiat Oncol, Dallas, TX 75235 USA
关键词
deformable registration; adaptive radiotherapy; deformable phantom; deformable registration verification; ART verification; INTERFRACTIONAL ANATOMIC CHANGES; IMAGE REGISTRATION ALGORITHMS; RADIATION-THERAPY; RE-OPTIMIZATION; IMRT; GPU; DELINEATION; ACCURACY; CANCER; CT;
D O I
10.1118/1.4908205
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Patients' interfractional anatomic changes can compromise the initial treatment plan quality. To overcome this issue, adaptive radiotherapy (ART) has been introduced. Deformable image registration (DIR) is an important tool for ART and several deformable phantoms have been built to evaluate the algorithms' accuracy. However, there is a lack of deformable phantoms that can also provide dosimetric information to verify the accuracy of the whole ART process. The goal of this work is to design and construct a deformable head and neck (HN) ART quality assurance (QA) phantom with in vivo dosimetry. Methods: An axial slice of a HN patient is taken as a model for the phantom construction. Six anatomic materials are considered, with HU numbers similar to a real patient. A filled balloon inside the phantom tissue is inserted to simulate tumor. Deflation of the balloon simulates tumor shrinkage. Nonradiopaque surface markers, which do not influence DIR algorithms, provide the deformation ground truth. Fixed and movable holders are built in the phantom to hold a diode for dosimetric measurements. Results: The measured deformations at the surface marker positions can be compared with deformations calculated by a DIR algorithm to evaluate its accuracy. In this study, the authors selected a Demons algorithm as a DIR algorithm example for demonstration purposes. The average error magnitude is 2.1 mm. The point dose measurements from the in vivo diode dosimeters show a good agreement with the calculated doses from the treatment planning system with a maximum difference of 3.1% of prescription dose, when the treatment plans are delivered to the phantom with original or deformed geometry. Conclusions: In this study, the authors have presented the functionality of this deformable HN phantom for testing the accuracy of DIR algorithms and verifying the ART dosimetric accuracy. The authors' experiments demonstrate the feasibility of this phantom serving as an end-to-end ART QA phantom. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:1490 / 1497
页数:8
相关论文
共 50 条
  • [31] In vivo estimation of midline dose maps by transit dosimetry in head and neck radiotherapy
    Broggi, S
    Fiorino, C
    Calandrino, R
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2002, 75 (900): : 974 - 981
  • [32] Quality Assurance of Deformable Image Registration in Radiotherapy
    Watkins, W. T.
    Bzdusek, K.
    Siebers, J. V.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 399 - 399
  • [33] Dosimetry audit in advanced radiotherapy using in-house developed anthropomorphic head & neck phantom
    Kakade, Nitin R.
    Kumar, Rajesh
    Sharma, S. D.
    Sapra, B. K.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2024, 10 (02)
  • [34] Evaluation of a deformable image registration quality assurance tool for head and neck cancer patients
    Mee, Molly
    Stewart, Kate
    Lathouras, Marika
    Truong, Helen
    Hargrave, Catriona
    [J]. JOURNAL OF MEDICAL RADIATION SCIENCES, 2020, 67 (04) : 284 - 293
  • [35] Head and neck adaptive radiotherapy
    Graff, P.
    Huger, S.
    Kirby, N.
    Pouliot, J.
    [J]. CANCER RADIOTHERAPIE, 2013, 17 (5-6): : 513 - 522
  • [36] Development of a modified head and neck quality assurance phantom for use in stereotactic radiosurgery trials
    Faught, Austin M.
    Kry, Stephen F.
    Luo, Dershan
    Molineu, Andrea
    Bellezza, David
    Gerber, Russell L.
    Davidson, Scott E.
    Bosch, Walter
    Drzymala, Robert E.
    Galvin, Jim
    Timmerman, Robert
    Sheehan, Jason
    Gillin, Michael T.
    Ibbott, Geoffrey S.
    Followill, David S.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (04): : 206 - 215
  • [37] The Impact of Clinical Trial Quality Assurance on Outcome in Head and Neck Radiotherapy Treatment
    Zhong, Haoyu
    Men, Kuo
    Wang, Jiazhou
    van Soest, Johan
    Rosenthal, David
    Dekker, Andre
    Zhang, Zhen
    Xiao, Ying
    [J]. FRONTIERS IN ONCOLOGY, 2019, 9
  • [38] HEAD-NECK ADAPTIVE IMRT TREATMENTS USING AN IN VIVO PORTAL DOSIMETRY
    Cilia, S.
    Grimaldi, L.
    Augelli, B.
    D'Onofrio, G.
    Digesu, C.
    Macchia, G.
    Fidanzio, A.
    Morganti, A. G.
    Piermattei, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S34 - S34
  • [39] Evaluation of deformable image registration methods for dose monitoring in head and neck adaptive radiotherapy
    Rigaud, B.
    Simon, A.
    Castelli, J.
    Gobeli, M.
    Arango, J. D. Ospina
    Cazoulat, G.
    Henry, O.
    Haigron, P.
    De Crevoisier, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S488 - S489
  • [40] 4D in-vivo dosimetry in radiotherapy
    Cherpak, A. J.
    Buckley, L.
    Ding, W.
    Ashton, R.
    Cygler, J. E.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S52 - S52