Evaluation of a deformable image registration quality assurance tool for head and neck cancer patients

被引:7
|
作者
Mee, Molly [1 ]
Stewart, Kate [1 ,2 ]
Lathouras, Marika [2 ]
Truong, Helen [2 ]
Hargrave, Catriona [1 ,3 ]
机构
[1] Queensland Univ Technol, Sch Clin Sci, Fac Hlth, Brisbane, Qld, Australia
[2] Royal Brisbane & Womens Hosp, Dept Radiat Oncol, Metro North Hlth Serv Dist, Herston, Qld 4029, Australia
[3] Metro South Hlth Serv Dist, Div Canc Serv, Radiat Oncol PAH Raymond Terrace Campus, South Brisbane, Qld, Australia
关键词
deformable image registration; quality assurance; radiation therapy; treatment planning; RADIATION-THERAPY; CONTOUR PROPAGATION; PLANNING CT; ACCURACY;
D O I
10.1002/jmrs.428
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: A challenge in implementing deformable image registration (DIR) in radiation therapy planning is effectively communicating registration accuracy to the radiation oncologist. This study aimed to evaluate the MIM (R) quality assurance (QA) tool for rating DIR accuracy. Methods: Retrospective DIR was performed on CT images for 35 head and neck cancer patients. The QA tool was used to rate DIR accuracy as good, fair or bad. Thirty registered patient images were assessed independently by three RTs and a further five patients assessed by five RTs. Ratings were evaluated by comparison of Hausdorff Distance (HD), Mean Distance to Agreement (MDA), Dice Similarity Coefficients (DSC) and Jacobian determinants for parotid and mandible subregions on the two CTs post-DIR. Inter-operator reliability was assessed using Krippendorff's alpha coefficient (KALPA). Rating time and volume measures for each rating were also calculated.Results: Quantitative metrics calculated for most anatomical subregions reflected the expected trend by registration accuracy, with good obtaining the most ideal values on average (HD = 7.50 +/- 3.18, MDA = 0.64 +/- 0.47, DSC = 0.90 +/- 0.07, Jacobian = 0.95 +/- 0.06). Highest inter-operator reliability was observed for good ratings and within the parotids (KALPA 0.66-0.93), whilst ratings varied the most in regions of dental artefact. Overall, average rating time was 33 minutes and the least commonly applied rating by volume was fair.:Conclusion:Results from qualitative and quantitative data, operator rating differences and rating time suggest highlighting only bad regions of DIR accuracy and implementing clinical guidelines and RT training for consistent and efficient use of the QA tool.
引用
收藏
页码:284 / 293
页数:10
相关论文
共 50 条
  • [1] Quality assurance in head and neck cancer registration
    Holt, DJ
    Williams, EMI
    Jones, AS
    Lennon, MA
    [J]. JOURNAL OF DENTAL RESEARCH, 1998, 77 : 862 - 862
  • [2] Deformable image registration for IMRT of head and neck cancer (WIP)
    Meyer, J
    Ford, E
    Kinahan, PE
    Lewellen, B
    Rajendran, J
    Schwartz, DL
    Phillips, M
    [J]. MEDICAL PHYSICS, 2003, 30 (06) : 1364 - 1364
  • [3] EVALUATION OF DEFORMABLE IMAGE CO-REGISTRATION IN ADAPTIVE IMRT FOR HEAD AND NECK CANCER
    Olteanu, A. M. L.
    Madani, I.
    De Gersem, W.
    Vercauteren, T.
    De Neve, W.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S443 - S444
  • [4] Quality Assurance of Deformable Image Registration in Radiotherapy
    Watkins, W. T.
    Bzdusek, K.
    Siebers, J. V.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 399 - 399
  • [5] Identifying Representative Examples of Clinically Relevant Deformable Image Registration Biases for Head and Neck Cancer Patients Prior to TG-132 Quality Assurance
    Weppler, Sarah
    Schinkel, Colleen
    Smith, Wendy
    [J]. MEDICAL PHYSICS, 2019, 46 (11) : 5379 - 5380
  • [6] A finite element head and neck model as a supportive tool for deformable image registration
    Jihun Kim
    Kazuhiro Saitou
    Martha M. Matuszak
    James M. Balter
    [J]. International Journal of Computer Assisted Radiology and Surgery, 2016, 11 : 1311 - 1317
  • [7] A finite element head and neck model as a supportive tool for deformable image registration
    Kim, Jihun
    Saitou, Kazuhiro
    Matuszak, Martha M.
    Balter, James M.
    [J]. INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2016, 11 (07) : 1311 - 1317
  • [8] Verification of a Deformable Image Registration Algorithm for Head and Neck Cancer Therapy
    Dogan, N.
    Sleeman, W.
    Fatyga, M.
    Lehman, W.
    Christensen, G.
    Wu, J.
    Weiss, E.
    Zhang, B.
    Williamson, J.
    [J]. MEDICAL PHYSICS, 2010, 37 (06)
  • [9] Utilizing Deformable Image Registration and Quality Assurance to Automate Analysis of Anatomic Changes After Head and Neck Radiation Therapy
    Watkins, W.
    Sim, A. J.
    Chen, Q.
    Wilson, D. D.
    Read, P. W.
    Siebers, J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : S98 - S99
  • [10] Evaluation of Deformable Image Registration Methods for Dose Monitoring in Head and Neck Radiotherapy
    Rigaud, Bastien
    Simon, Antoine
    Castelli, Joel
    Gobeli, Maxime
    Arango, Juan-David Ospina
    Cazoulat, Guillaume
    Henry, Olivier
    Haigron, Pascal
    De Crevoisier, Renaud
    [J]. BIOMED RESEARCH INTERNATIONAL, 2015, 2015