An intelligent control scheme for precise tip-motion control in atomic force microscopy

被引:2
|
作者
Wang, Yanyan [1 ]
Hu, Xiaodong [2 ]
Xu, Linyan [2 ]
机构
[1] Tianjin Univ Technol & Educ, Tianjin Key Lab Informat Sensing & Intelligent Co, Tianjin, Peoples R China
[2] Tianjin Univ, State Key Lab Precis Measuring Technol & Instrume, Tianjin 300072, Peoples R China
关键词
AFM; tip-motion control; intelligent controller; fuzzy controller; PI;
D O I
10.1002/sca.21244
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The paper proposes a new intelligent control method to precisely control the tip motion of the atomic force microscopy (AFM). The tip moves up and down at a high rate along the z direction during scanning, requiring the utilization of a rapid feedback controller. The standard proportional-integral (PI) feedback controller is commonly used in commercial AFMs to enable topography measurements. The controller's response performance is determined by the set of the proportional (P) parameter and the integral (I) parameter. However, the two parameters cannot be automatically altered simultaneously according to the scanning speed and the surface topography during continuors scanning, leading to an inaccurate measurement. Thus a new intelligent controller combining the fuzzy controller and the PI controller is put forward in the paper. The new controller automatically selects the most appropriate PI parameters to achieve a fast response rate on basis of the tracking errors. In the experimental setup, the new controller is realized with a digital signal process (DSP) system, implemented in a conventional AFM system. Experiments are carried out by comparing the new method with the standard PI controller. The results demonstrate that the new method is more robust and effective for the precise tip motion control, corresponding to the achievement of a highly qualified image by shortening the response time of the controller. SCANNING 38:93-99, 2016. (c) 2015 Wiley Periodicals, Inc.
引用
收藏
页码:93 / 99
页数:7
相关论文
共 50 条
  • [31] Stabilized atomic force microscopy imaging in liquids using second harmonic of cantilever motion for setpoint control
    Schiener, J
    Witt, S
    Stark, M
    Guckenberger, R
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (08): : 2564 - 2568
  • [32] Real-Time Reconstruction of Multimode Tip Motion of Microcantilevers in Dynamic Atomic Force Microscopy
    Liu, Zhen
    Jeong, Younkoo
    Menq, Chia-Hsiang
    [J]. IEEE-ASME TRANSACTIONS ON MECHATRONICS, 2016, 21 (02) : 825 - 837
  • [33] DYNAMICS OF TIP SUBSTRATE INTERACTIONS IN ATOMIC FORCE MICROSCOPY
    LANDMAN, U
    LUEDTKE, WD
    NITZAN, A
    [J]. SURFACE SCIENCE, 1989, 210 (03) : L177 - L184
  • [34] A New Kind of Atomic Force Microscopy Scan Control Enabled by Artificial Intelligence: Concept for Achieving Tip and Sample Safety Through Asymmetric Control
    Degenhardt J.
    Bounaim M.W.
    Deng N.
    Tutsch R.
    Dai G.
    [J]. Nanomanufacturing and Metrology, 2024, 7 (01)
  • [35] EFFECT OF THE TIP STRUCTURE ON ATOMIC-FORCE MICROSCOPY
    SASAKI, N
    TSUKADA, M
    [J]. PHYSICAL REVIEW B, 1995, 52 (11): : 8471 - 8482
  • [36] ATOMIC FORCE MICROSCOPY USING ZNO WHISKER TIP
    KADO, H
    YOKOYAMA, K
    TOHDA, T
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1992, 63 (06): : 3330 - 3332
  • [37] Contrast artifacts in tapping tip atomic force microscopy
    A. Kühle
    A.H. Sorensen
    J.B. Zandbergen
    J. Bohr
    [J]. Applied Physics A, 1998, 66 : S329 - S332
  • [38] A NOVEL ZNO WHISKER TIP FOR ATOMIC FORCE MICROSCOPY
    KADO, H
    YOKOYAMA, K
    TOHDA, T
    [J]. ULTRAMICROSCOPY, 1992, 42 : 1659 - 1663
  • [39] Contrast artifacts in tapping tip atomic force microscopy
    Kuhle, A
    Sorensen, AH
    Zandbergen, JB
    Bohr, J
    [J]. APPLIED PHYSICS A-MATERIALS SCIENCE & PROCESSING, 1998, 66 (Suppl 1): : S329 - S332
  • [40] Characterizing Atomic Force Microscopy Tip Shape in Use
    Wang, Chunmei
    Itoh, Hiroshi
    Sun, Jielin
    Hu, Jun
    Shen, Dianhong
    Ichimura, Shingo
    [J]. JOURNAL OF NANOSCIENCE AND NANOTECHNOLOGY, 2009, 9 (02) : 803 - 808