When recording images with an atomic force microscope using the resonant vibrating cantilever mode, surprising strange results are often achieved. Typical artifacts are strange contours, unexpected height shifts, and sudden changes of the apparent resolution in the acquired images. Such artifacts can be related to the dynamical properties of the cantilever under the influence of the force between the tip and the sample. The damping of the cantilever oscillation can be either due to attractive interaction between the tip and the sample or due to a combination of attractive and repulsive interaction. The oscillating cantilever will be in a specific swing mode according to which type of interaction is dominating, and it is the switching between these modes that is responsible for a range of artifacts observed during image acquisition. This includes the artifact often referred to as "contrast reversal".
引用
收藏
页码:S329 / S332
页数:4
相关论文
共 50 条
  • [1] Contrast artifacts in tapping tip atomic force microscopy
    Department of Physics, Bldg. 307, Technical University of Denmark, DK-2800 Lyngby, Denmark
    [J]. Appl Phys A, SUPPL. 1 (S329-S332):
  • [2] Contrast artifacts in tapping tip atomic force microscopy
    A. Kühle
    A.H. Sorensen
    J.B. Zandbergen
    J. Bohr
    [J]. Applied Physics A, 1998, 66 : S329 - S332
  • [3] TIP ARTIFACTS OF MICROFABRICATED FORCE SENSORS FOR ATOMIC FORCE MICROSCOPY
    GRUTTER, P
    ZIMMERMANNEDLING, W
    BRODBECK, D
    [J]. APPLIED PHYSICS LETTERS, 1992, 60 (22) : 2741 - 2743
  • [4] Study of tip-sample contact process and phase contrast in tapping mode atomic force microscopy
    Wang, XP
    Lei, L
    Hu, HL
    Kun, Z
    [J]. ACTA PHYSICA SINICA, 2004, 53 (04) : 1008 - 1014
  • [5] Studies of tip wear processes in tapping mode™ atomic force microscopy
    Su, CM
    Huang, L
    Kjoller, K
    Babcock, K
    [J]. ULTRAMICROSCOPY, 2003, 97 (1-4) : 135 - 144
  • [6] Mechanisms of formation of phase contrast in tapping mode atomic force microscopy
    Chizhik, SA
    Ahn, HS
    Suslov, AA
    Kovalev, AV
    Kim, CH
    [J]. PHYSICS OF LOW-DIMENSIONAL STRUCTURES, 2001, 3-4 : 39 - 46
  • [7] Artifacts in Atomic Force Microscopy
    Gainutdinov R.V.
    Arutyunov P.A.
    [J]. Russian Microelectronics, 2001, 30 (4) : 219 - 224
  • [8] Artifacts in atomic force microscopy
    Gajnutdinov, R.V.
    Arutyunov, P.A.
    [J]. Mikroelektronika, 2001, 30 (04): : 257 - 266
  • [9] Imaging stability and average tip-sample force in tapping mode atomic force microscopy
    Kowalewski, T
    Legleiter, J
    [J]. JOURNAL OF APPLIED PHYSICS, 2006, 99 (06)
  • [10] Imaging stability and average tip-sample force in tapping mode atomic force microscopy
    Kowalewski, Tomasz
    Legleiter, Justin
    [J]. Journal of Applied Physics, 2006, 99 (06):