Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients

被引:147
|
作者
Frasca, Denis [1 ,2 ,3 ]
Dahyot-Fizelier, Claire [1 ,2 ,3 ]
Catherine, Karen [1 ]
Levrat, Quentin [1 ]
Debaene, Bertrand [1 ,2 ,3 ]
Mimoz, Olivier [1 ,2 ,3 ]
机构
[1] Ctr Hosp Univ Poitiers, Poitiers, France
[2] INSERM, ERI 23, F-75654 Paris 13, France
[3] Univ Poitiers, UFR Med Pharm, Poitiers, France
关键词
anemia; blood transfusion; hemoglobinometry methods; intensive care; monitoring; point-of-care; BLOOD-TRANSFUSION; ANEMIA; ANALYZER; SCORE;
D O I
10.1097/CCM.0b013e3182227e2d
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To determine whether noninvasive hemoglobin measurement by Pulse CO-Oximetry could provide clinically acceptable absolute and trend accuracy in critically ill patients, compared to other invasive methods of hemoglobin assessment available at bedside and the gold standard, the laboratory analyzer. Design: Prospective study. Setting: Surgical intensive care unit of a university teaching hospital. Patients: Sixty-two patients continuously monitored with Pulse CO-Oximetry (Masimo Radical-7). Interventions: None. Measurements and Results: Four hundred seventy-one blood samples were analyzed by a point-of-care device (HemoCue 301), a satellite lab CO-Oximeter (Siemens RapidPoint 405), and a laboratory hematology analyzer (Sysmex XT-2000i), which was considered the reference device. Hemoglobin values reported from the invasive methods were compared to the values reported by the Pulse CO-Oximeter at the time of blood draw. When the case-to-case variation was assessed, the bias and limits of agreement were 0.0 +/- 1.0 g/dL for the Pulse CO-Oximeter, 0.3 +/- 1.3g/dL for the point-of-care device, and 0.9 +/- 0.6 g/dL for the satellite lab CO-Oximeter compared to the reference method. Pulse CO-Oximetry showed similar trend accuracy as satellite lab CO-Oximetry, whereas the point-of-care device did not appear to follow the trend of the laboratory analyzer as well as the other test devices. Conclusion: When compared to laboratory reference values, hemoglobin measurement with Pulse CO-Oximetry has absolute accuracy and trending accuracy similar to widely used, invasive methods of hemoglobin measurement at bedside. Hemoglobin measurement with pulse CO-Oximetry has the additional advantages of providing continuous measurements, noninvasively, which may facilitate hemoglobin monitoring in the intensive care unit. (Crit Care Med 2011; 39:2277-2282)
引用
收藏
页码:2277 / 2282
页数:6
相关论文
共 50 条
  • [41] CONTINUOUS INTRAARTERIAL OXYGEN MONITORING - ACCURACY AND RELIABILITY IN THE SURGICAL INTENSIVE-CARE UNIT
    LEMUS, JF
    KEARNEY, T
    MARGULIES, DR
    MACKENZIE, DJ
    LEYERLE, BJ
    SHABOT, MM
    AMERICAN SURGEON, 1992, 58 (12) : 740 - 742
  • [42] The relative trending accuracy of noninvasive continuous hemoglobin monitoring during hemodialysis in critically ill patients
    Yamada, Hiroshi
    Saeki, Minako
    Ito, Junko
    Kawada, Kazuhiro
    Higurashi, Aya
    Funakoshi, Hiromi
    Takeda, Kohji
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2015, 29 (01) : 107 - 112
  • [43] The Changing Landscape of Noninvasive Ventilation in the Intensive Care Unit
    Patel, Bhakti K.
    Kress, John P.
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2015, 314 (16): : 1697 - 1699
  • [44] Continuous Noninvasive Hemoglobin Monitor from Pulse Ox: Ready for Prime Time?
    Joseph, Bellal
    Hadjizacharia, Pantelis
    Aziz, Hassan
    Snyder, Kara
    Wynne, Julie
    Kulvatunyou, Narong
    Tang, Andrew
    O'Keeffe, Terence
    Latifi, Rifat
    Friese, Randall
    Rhee, Peter
    WORLD JOURNAL OF SURGERY, 2013, 37 (03) : 525 - 529
  • [45] Continuous Noninvasive Hemoglobin Monitor from Pulse Ox: Ready for Prime Time?
    Bellal Joseph
    Pantelis Hadjizacharia
    Hassan Aziz
    Kara Snyder
    Julie Wynne
    Narong Kulvatunyou
    Andrew Tang
    Terence O’Keeffe
    Rifat Latifi
    Randall Friese
    Peter Rhee
    World Journal of Surgery, 2013, 37 : 525 - 529
  • [46] Accuracy of pulse oximetry in the intensive care unit
    Van de Louw, A
    Cracco, C
    Cerf, C
    Harf, A
    Duvaldestin, P
    Lemaire, F
    Brochard, L
    INTENSIVE CARE MEDICINE, 2001, 27 (10) : 1606 - 1613
  • [47] Accuracy of pulse oximetry in the intensive care unit
    A. Louw
    C. Cracco
    C. Cerf
    A. Harf
    P. Duvaldestin
    F. Lemaire
    L. Brochard
    Intensive Care Medicine, 2001, 27 : 1606 - 1613
  • [48] Diagnostic accuracy in the medical intensive care unit
    M Podbregar
    G Voga
    B Krivec
    R Skale
    R Parežnik
    L Gabršček
    Critical Care, 5 (Suppl 1):
  • [49] Accuracy of noninvasive hemoglobin monitoring in patients at risk for hemorrhage
    Tsuei, Betty J.
    Hanseman, Dennis J.
    Blakeman, Michael J.
    Blakeman, Thomas C.
    Yang, Sung H.
    Branson, Richard D.
    Gerlach, Travis W.
    JOURNAL OF TRAUMA AND ACUTE CARE SURGERY, 2014, 77 : S134 - S139
  • [50] KETAMINE FOR CONTINUOUS SEDATION IN PATIENTS ADMITTED TO THE PEDIATRIC INTENSIVE CARE UNIT
    Judge, Pavan
    Bryce, Rhonda
    Pawha, Punam
    Holt, Tanya
    CRITICAL CARE MEDICINE, 2012, 40 (12) : U305 - U305