Continuous Noninvasive Hemoglobin Monitor from Pulse Ox: Ready for Prime Time?

被引:23
|
作者
Joseph, Bellal [1 ]
Hadjizacharia, Pantelis [1 ]
Aziz, Hassan [1 ]
Snyder, Kara [1 ]
Wynne, Julie [1 ]
Kulvatunyou, Narong [1 ]
Tang, Andrew [1 ]
O'Keeffe, Terence [1 ]
Latifi, Rifat [1 ]
Friese, Randall [1 ]
Rhee, Peter [1 ]
机构
[1] Univ Arizona, Dept Surg, Tucson, AZ 85719 USA
关键词
INJURIES; TRAUMA; ACCURACY; ARTERIAL;
D O I
10.1007/s00268-012-1871-y
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background Advances in technology have allowed for continuous noninvasive hemoglobin monitoring (SpHb), which may enable earlier detection of hemorrhage and more efficient surgical and/or blood transfusion management. The use of SpHb has not been described in the trauma population. The purpose of the present study was to evaluate the accuracy of a SpHb measurement device in severely injured trauma patients. Methods We performed a prospective cohort analysis of severely injured trauma patients admitted to the intensive care unit (ICU) at our level I trauma center over a 6 month period. Serial IHb (invasive hemoglobin) levels and SpHb for the first 72 h were measured. Each SpHb measurement was matched with a corresponding IHb measurement. We defined normal Hgb as[8 mg/dL and low Hgb as <8 mg/dL. Data were then grouped based on Hgb level. Sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and Spearman correlation coefficient plot were calculated. Results A total of 23 trauma patients with 89 data pairs were reviewed. Eighty-six percent of the patients were male with a mean age of 32 years and a mean injury severity score (ISS) of 21.1 +/- 14. Invasive hemoglobin had a range of 7.2-16.9 and SpHb had a range of 3.3-15.2. The average mean and difference between IHb and SpHb were 10.7 and 1, respectively. Continuous noninvasive hemoglobin measurement did not record data points 13.5 % of the time. The Spearman correlation plot revealed a correlation of R = 0.670 (p < 0.001). After dichotomization with Hgb > 8, SpHb was found to have a sensitivity of 91 %, PPV 96 %, specificity 40 %, NPV 20 %, and an accuracy of 88 %. Conclusions The continuous noninvasive hemoglobin monitor does not appear to represent serum hemoglobin levels accurately in severely injured trauma patients. However, we were able to identify utility for this noninvasive tool when Hgb was dichotomized into normal or low levels.
引用
收藏
页码:525 / 529
页数:5
相关论文
共 50 条
  • [1] Continuous Noninvasive Hemoglobin Monitor from Pulse Ox: Ready for Prime Time?
    Bellal Joseph
    Pantelis Hadjizacharia
    Hassan Aziz
    Kara Snyder
    Julie Wynne
    Narong Kulvatunyou
    Andrew Tang
    Terence O’Keeffe
    Rifat Latifi
    Randall Friese
    Peter Rhee
    World Journal of Surgery, 2013, 37 : 525 - 529
  • [2] Continuous noninvasive hemoglobin monitoring: ready for prime time?
    Suehiro, Koichi
    Joosten, Alexandre
    Alexander, Brenton
    Cannesson, Maxime
    CURRENT OPINION IN CRITICAL CARE, 2015, 21 (03) : 265 - 270
  • [3] Pulse Waveform Analysis: Is It Ready for Prime Time?
    Bernhard Hametner
    Siegfried Wassertheurer
    Current Hypertension Reports, 2017, 19
  • [4] Noninvasive Electrocardiographic Mapping: Are We Ready for Prime Time?
    Aryana, Arash
    O'Neill, Padraig Gearoid
    d'Avila, Andre
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2015, 4 (10):
  • [5] Pulse Waveform Analysis: Is It Ready for Prime Time?
    Hametner, Bernhard
    Wassertheurer, Siegfried
    CURRENT HYPERTENSION REPORTS, 2017, 19 (09)
  • [6] Continuous EEG monitoring: is it ready for prime time?
    Kurtz, Pedro
    Hanafy, Khalid A.
    Claassen, Jan
    CURRENT OPINION IN CRITICAL CARE, 2009, 15 (02) : 99 - 109
  • [7] Continuous speech recognition: Ready for prime time?
    Finley, S
    TOWARD AN ELECTRONIC PATIENT RECORD '98 CONFERENCE AND EXPOSITION, PROCEEDINGS - VOLS 1-3, 1998, : A335 - A337
  • [8] Noninvasive prenatal testing for aneuploidy-ready for prime time?
    Chitty, Lyn S.
    Hill, Melissa
    White, Helen
    Wright, David
    Morris, Stephen
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2012, 206 (04) : 269 - 275
  • [9] Noninvasive Tools to Assess Hepatic Fibrosis: Ready for Prime Time?
    Schmeltzer, Paul A.
    Talwalkar, Jayant A.
    GASTROENTEROLOGY CLINICS OF NORTH AMERICA, 2011, 40 (03) : 507 - +
  • [10] PULSE OXIMETRY, A CONTINUOUS AND NONINVASIVE MONITOR - A COMPARISON OF 2 INSTRUMENTS
    ALTEMEYER, KH
    MAYER, J
    BERGSEITER, S
    FOSEL, T
    ANAESTHESIST, 1986, 35 (01): : 43 - 45