Study of the response of plastic scintillation detectors in small-field 6 MV photon beams by Monte Carlo simulations

被引:33
|
作者
Wang, Lilie L. W. [1 ]
Beddar, Sam [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
plastic scintillation detectors; small-field photon beams; Monte Carlo simulations; CERENKOV RADIATION; NONSTANDARD FIELDS; DOSIMETRY; ENERGY; ELECTRON; SYSTEM;
D O I
10.1118/1.3554644
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the response of plastic scintillation detectors (PSDs) in a 6 MV photon beam of various field sizes using Monte Carlo simulations. Methods: Three PSDs were simulated: A BC-400 and a BCF-12, each attached to a plastic-core optical fiber, and a BC-400 attached to an air-core optical fiber. PSD response was calculated as the detector dose per unit water dose for field sizes ranging from 10 x 10 down to 0.5 x 0.5 cm(2) for both perpendicular and parallel orientations of the detectors to an incident beam. Similar calculations were performed for a CC01 compact chamber. The off-axis dose profiles were calculated in the 0.5 x 0.5 cm(2) photon beam and were compared to the dose profile calculated for the CC01 chamber and that calculated in water without any detector. The angular dependence of the PSDs' responses in a small photon beam was studied. Results: In the perpendicular orientation, the response of the BCF-12 PSD varied by only 0.5% as the field size decreased from 10 x 10 to 0.5 x 0.5 cm(2), while the response of BC-400 PSD attached to a plastic-core fiber varied by more than 3% at the smallest field size because of its longer sensitive region. In the parallel orientation, the response of both PSDs attached to a plastic-core fiber varied by less than 0.4% for the same range of field sizes. For the PSD attached to an air-core fiber, the response varied, at most, by 2% for both orientations. Conclusions: The responses of all the PSDs investigated in this work can have a variation of only 1%-2% irrespective of field size and orientation of the detector if the length of the sensitive region is not more than 2 mm long and the optical fiber stems are prevented from pointing directly to the incident source. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3554644]
引用
收藏
页码:1596 / 1599
页数:4
相关论文
共 50 条
  • [41] COMMISSIONING OF A MONTE CARLO DOSE CALCULATION ALGORITHM FOR 6 MV AND 15 MV PHOTON BEAMS FROM A 160 MLC SIEMENS ACCELERATOR
    Zucca Aparicio, D.
    Perez Moreno, J. M.
    Fernandez-Leton, P.
    Garcia Ruiz-Zorrilla, J.
    Minambres, A.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S521 - S522
  • [42] Monte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry
    Yani, S.
    Rhani, M. F.
    Soh, R. C. X.
    Haryanto, F.
    Arif, I.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2017, 15 (03): : 275 - +
  • [43] Monte Carlo study of small-field dosimetry for an ELEKTA Unity MR-Linac system
    Yano, Masayuki
    Araki, Fujio
    Ohno, Takeshi
    RADIATION PHYSICS AND CHEMISTRY, 2022, 194
  • [44] A new reference-type ionization chamber with direction-independent response for use in small-field photon-beam dosimetry - An experimental and Monte Carlo study
    Delfs, Bjoern
    Kapsch, Ralf-Peter
    Chofor, Ndimofor
    Looe, Hui Khee
    Harder, Dietrich
    Poppe, Bjoern
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (01): : 39 - 48
  • [45] Feasibility study of using Stereotactic Field Diode for field output factors measurement and evaluating three new detectors for small field relative dosimetry of 6 and 10 MV photon beams
    Gul, Attia
    Fukuda, Shigekazu
    Mizuno, Hideyuki
    Taku, Nakaji
    Kakakhel, M. Basim
    Mirza, Sikander M.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11): : 23 - 36
  • [46] A comparison of small-field tissue phantom ratio data generation methods for an Elekta Agility 6 MV photon beam
    Richmond, Neil
    Brackenridge, Robert
    MEDICAL DOSIMETRY, 2014, 39 (01) : 60 - 63
  • [47] Study of detectors response on the scatter radiation of 6 MV flattening filter free beams
    Pratiwi, R. D. T.
    Ryangga, D.
    Pawiro, S. A.
    18TH ASIA-OCEANIA CONGRESS OF MEDICAL PHYSICS (AOCMP) & 16TH SOUTH-EAST ASIA CONGRESS OF MEDICAL PHYSICS (SEACOMP), 2019, 1248
  • [48] Comparison of synthetic Diamond Detector TM 60019 with other PTW Small-field Detector with 6 MV Photon Energy
    Scherf, C.
    Henrich, N.
    Koehn, J.
    Bluemer, N.
    Heine, B.
    Loutfi-Krauss, B.
    Mielke, M.
    Zink, K.
    Ramm, U.
    Licher, J.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 : 93 - 93
  • [49] Monte Carlo study of conversion factors for ionization chamber dosimetry in solid slab phantoms for MV photon beams
    Park, Dong-wook
    Lee, Jai-ki
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 69 (04) : 647 - 652
  • [50] Elaboration and experimental validation of a Monte Carlo source model for linac 6 MV photon beams with and without Flattening Filter
    Aboulbanine, Zakaria
    Bahhous, Karim
    Radiation Physics and Chemistry, 2022, 201