Study of the response of plastic scintillation detectors in small-field 6 MV photon beams by Monte Carlo simulations

被引:33
|
作者
Wang, Lilie L. W. [1 ]
Beddar, Sam [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
关键词
plastic scintillation detectors; small-field photon beams; Monte Carlo simulations; CERENKOV RADIATION; NONSTANDARD FIELDS; DOSIMETRY; ENERGY; ELECTRON; SYSTEM;
D O I
10.1118/1.3554644
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the response of plastic scintillation detectors (PSDs) in a 6 MV photon beam of various field sizes using Monte Carlo simulations. Methods: Three PSDs were simulated: A BC-400 and a BCF-12, each attached to a plastic-core optical fiber, and a BC-400 attached to an air-core optical fiber. PSD response was calculated as the detector dose per unit water dose for field sizes ranging from 10 x 10 down to 0.5 x 0.5 cm(2) for both perpendicular and parallel orientations of the detectors to an incident beam. Similar calculations were performed for a CC01 compact chamber. The off-axis dose profiles were calculated in the 0.5 x 0.5 cm(2) photon beam and were compared to the dose profile calculated for the CC01 chamber and that calculated in water without any detector. The angular dependence of the PSDs' responses in a small photon beam was studied. Results: In the perpendicular orientation, the response of the BCF-12 PSD varied by only 0.5% as the field size decreased from 10 x 10 to 0.5 x 0.5 cm(2), while the response of BC-400 PSD attached to a plastic-core fiber varied by more than 3% at the smallest field size because of its longer sensitive region. In the parallel orientation, the response of both PSDs attached to a plastic-core fiber varied by less than 0.4% for the same range of field sizes. For the PSD attached to an air-core fiber, the response varied, at most, by 2% for both orientations. Conclusions: The responses of all the PSDs investigated in this work can have a variation of only 1%-2% irrespective of field size and orientation of the detector if the length of the sensitive region is not more than 2 mm long and the optical fiber stems are prevented from pointing directly to the incident source. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3554644]
引用
收藏
页码:1596 / 1599
页数:4
相关论文
共 50 条
  • [21] Modelling 6 MV photon beams of a stereotactic radiosurgery system for Monte Carlo treatment planning
    Deng, J
    Guerrero, T
    Ma, CM
    Nath, R
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (09): : 1689 - 1704
  • [22] Measured and Monte Carlo calculated output ratios in air for 6 and 18 MV photon beams
    Ding, G
    Yu, C
    MEDICAL PHYSICS, 2004, 31 (06) : 1900 - 1900
  • [23] Commissioning 6 MV photon beams of a stereotactic radiosurgery system for Monte Carlo treatment planning
    Deng, J
    Ma, CM
    Hai, J
    Nath, R
    MEDICAL PHYSICS, 2003, 30 (12) : 3124 - 3134
  • [24] Out-of-field beam characteristics of a 6 MV photon beam: Results of a Monte Carlo study
    Atarod, M.
    Shokrani, P.
    Azarnoosh, A.
    APPLIED RADIATION AND ISOTOPES, 2013, 72 : 182 - 194
  • [25] Dosimetry of interface region near closed air cavities for Co-60, 6 MV and 15 MV photon beams using Monte Carlo simulations
    Joshi, Chandra P.
    Darko, Johnson
    Vidyasagar, P. B.
    Schreiner, L. John
    JOURNAL OF MEDICAL PHYSICS, 2010, 35 (02) : 73 - 80
  • [26] Characterization of small field electron beams for radiotherapy using Monte Carlo simulations
    Kapur, A
    Ma, CM
    Mok, E
    Findley, D
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 157 - 158
  • [27] Accuracy of Monte Carlo software PRIMO against a reference dosimetry dataset for 6 MV photon beams
    Hermida-Lopez, M.
    Sanchez-Artunedo, D.
    Calvo-Ortega, J. F.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S938 - S939
  • [28] Monte Carlo simulation approaches to dose distributions for 6 MV photon beams in clinical linear accelerator
    Tartar, Ahmet
    BIOCYBERNETICS AND BIOMEDICAL ENGINEERING, 2014, 34 (02) : 90 - 100
  • [29] A Monte Carlo study of the variation of electron fluence in water from a 6 MV photon beam outside of the field
    Kirkby, C.
    Field, C.
    MacKenzie, M.
    Syme, A.
    Fallone, B. G.
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (12): : 3563 - 3578
  • [30] Monte Carlo study on mucosal dose in oral and nasal cavity using photon beams with small field
    Chow, James C. L.
    Owrangi, Amir M.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2010, 10 (04) : 261 - 271