A new device to measure upper limb circumferences: validity and reliability

被引:0
|
作者
Devoogdt, N. [1 ,2 ,3 ]
Lemkens, H. [1 ,2 ]
Geraerts, I. [1 ,2 ]
Van Nuland, J. [1 ,2 ]
Flour, M. [4 ]
Coremans, T. [3 ]
Christiaens, M. R. [5 ]
Van Kampen, M. [1 ,2 ]
机构
[1] Univ Hosp Leuven, Dept Physiotherapy, B-3000 Louvain, Belgium
[2] Katholieke Univ Leuven, Fac Kinesiol & Rehabil Sci, Leuven, Belgium
[3] Univ Coll Antwerp, Dept Hlth Care Sci, Antwerp, Belgium
[4] Univ Hosp Leuven, Vasc Ctr, Louvain, Belgium
[5] Univ Hosp Leuven, Multidisciplinary Breast Ctr, Louvain, Belgium
关键词
Arm circumference; Measurements; Validation study; UPPER-EXTREMITY VOLUME; WATER DISPLACEMENT; LYMPHEDEMA; AGREEMENT; WOMEN;
D O I
暂无
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Aim. The aim of this paper was to examine the reliability and validity of a new measurement device that counters the disadvantages of the traditional method of arm circumference measurements. Methods. We measured the arm on the non-operated side of breast cancer patients. Sixty-four patients were measured twice by the same assessor and 48 patients were measured twice by two different assessors. The arm circumferences were measured at the olecranon and each 4 cm proximal and distal of the olecranon. The measurements were performed with a self-developed device consisting of a stainless steel bar on which a tapeline was fixed at every 4 cm distance. The arm volume was calculated from the circumference measurements with the frustrum formula and was also measured directly with the water displacement method. Results. For the circumference measurements, intrarater and interrater ICCs ranged between 0.942 and 0.998. ICCs for the calculated arm volume were also very high. No systematic changes between the first to the second assessment could be found. The standard error of measurement for the circumference measurements as well as for the calculated arm volume was low (between 0.8% and 2.0%). An increase of 1.0 cm or more of the arm circumference at any measurement side and of 55 ml or more of the calculated arm volume was clinically significant. Conclusion. Arm circumferences and also the calculated arm volume can be measured quickly and accurately with a simple and inexpensive device. [Int Angiol 2010;29:401-7]
引用
收藏
页码:401 / 407
页数:7
相关论文
共 50 条
  • [31] Reliability and validity of the upper limb DASH questionnaire in patients with distal radius fractures
    Westphal, T
    Piatek, S
    Schubert, S
    Schuschke, Y
    Winckler, S
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2002, 140 (04): : 447 - 451
  • [32] Validity and Reliability of upper limb muscles strength measurements by handheld dynamometer in COPD
    Beaumont, M.
    Peran, L.
    Berriet, A.
    Keller, R.
    Le Ber, C.
    EUROPEAN RESPIRATORY JOURNAL, 2022, 60
  • [33] THE VALIDITY AND RELIABILITY OF UNSUPPORTED UPPER LIMB EXERCISE TEST IN INDIVIDUALS WITH RHEUMATOID ARTHRITIS
    Cetin, S. Yaprak
    Calik, Bilge Basakci
    Ayan, Ayse
    Cavlak, Ugur
    ANNALS OF THE RHEUMATIC DISEASES, 2019, 78 : 2155 - 2156
  • [34] Reliability and validity of modified upper limb neurodynamic tests in patients with cervical radiculopathy
    Zanwar, Rupa
    Wani, Surendra
    HONG KONG PHYSIOTHERAPY JOURNAL, 2024, 44 (02) : 137 - 146
  • [35] VALIDITY AND RELIABILITY OF A NEW TEST OF UPPER BODY POWER
    Clemons, James M.
    Campbell, Brian
    Jeansonne, Chris
    JOURNAL OF STRENGTH AND CONDITIONING RESEARCH, 2010, 24 (06) : 1559 - 1565
  • [36] Validity and reliability of a new ankle dorsiflexion measurement device
    Gatt, Alfred
    Chockalingam, Nachiappan
    PROSTHETICS AND ORTHOTICS INTERNATIONAL, 2013, 37 (04) : 289 - 297
  • [37] The capabilities of upper extremity instrument: Reliability and validity of a measure of functional limitation in tetraplegia
    Marino, RJ
    Shea, JA
    Stineman, MG
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 1998, 79 (12): : 1512 - 1521
  • [38] Reliability testing of a new device to measure tibial rotation
    Olaf Lorbach
    P. Wilmes
    D. Theisen
    M. Brockmeyer
    S. Maas
    D. Kohn
    R. Seil
    Knee Surgery, Sports Traumatology, Arthroscopy, 2009, 17 : 920 - 926
  • [39] RELIABILITY OF A NEW DEVICE USED TO MEASURE SHOULDER SUBLUXATION
    HAYES, KW
    SULLIVAN, JE
    PHYSICAL THERAPY, 1989, 69 (09): : 762 - 767
  • [40] Reliability testing of a new device to measure tibial rotation
    Lorbach, Olaf
    Wilmes, P.
    Theisen, D.
    Brockmeyer, M.
    Maas, S.
    Kohn, D.
    Seil, R.
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2009, 17 (08) : 920 - 926