Verification of IMRT dose distributions using a water beam imaging system

被引:0
|
作者
Li, JS [1 ]
Ozhasoglu, C [1 ]
Deng, J [1 ]
Boyer, AL [1 ]
Ma, CM [1 ]
机构
[1] Stanford Univ, Dept Radiat Oncol, Sch Med, Stanford, CA 94305 USA
关键词
IMRT; quality assurance (QA); dosimetry; Monte Carlo; water beam imaging system (WBIS);
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A water beam imaging system (WBIS) was developed and used to verify dose distributions for intensity modulated radiotherapy (IMRT) using dynamic multileaf collimator (MLC). The WBIS consists of a water container, a scintillator screen, a charge-coupled device (CCD) camera, and a portable personal computer (Wellhofer Dosimetrie, Schwarzenbruck, Germany). The scintillation image is captured by the CCD camera. The pixel value in this image indicates the dose value in the scintillation screen. The verification is performed by comparing the WBIS image achieved from the measurement with dose distribution from the IMRT plan. Because of Light scattering in the WBIS the image is blurred. An iterative reconstruction algorithm is proposed to remove the blurring effect of light scattering. From the measured image of a 10 cm x 10 cm x-ray beam and the simulation result of the dose distribution using the Monte Carlo method, the blur function can be achieved. Based on this function, the proposed algorithm is applied to reconstruct the true dose distribution for an IMRT plan from the measured WBIS image. The reconstructed dose distributions are compared with Monte Carlo simulation results. Reasonable agreement can be observed from the comparison. The proposed approach makes it possible to carry out real-time quality assurance tasks for IMRT dose verification.
引用
收藏
页码:1158 / 1161
页数:4
相关论文
共 50 条
  • [1] Verification of IMRT dose distributions using a water beam imaging system
    Li, JS
    Boyer, AL
    Ma, CM
    [J]. MEDICAL PHYSICS, 2001, 28 (12) : 2466 - 2474
  • [2] Pretreatment dosimetric verification of IMRT plans with a beam imaging system
    Bindoni, L.
    Fontan, L.
    Mantovani, L.
    Princivalli, M.
    Tonetto, V.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S307 - S308
  • [3] Verification of IMRT dose distributions using a tissue equivalent plastic scintillator based dosimetry system
    Petric, MP
    Robar, JL
    Clark, BG
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1891 - 1891
  • [4] Clinical experience in using EPID for quantitative verification of IMRT dose distributions
    Wang, J. Z.
    Rice, R.
    Marcus, L.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2167 - 2167
  • [5] Verification of the absolute dose in IMRT:: beam set and/or beam by beam
    Marcié, S
    Charpiot, E
    Bensadoun, RJ
    Serrano, B
    Hérault, J
    Costa, A
    Gérard, JP
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S336 - S336
  • [6] Verification of IMRT dose distributions with EDR2 films
    Smulders, B
    [J]. RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S38 - S38
  • [7] Pretreatment verification of IMRT absolute dose distributions using a commercial a-Si EPID
    Talamonti, C.
    Casati, M.
    Bucciolini, M.
    [J]. MEDICAL PHYSICS, 2006, 33 (11) : 4367 - 4378
  • [8] Monte Carlo verification of IMRT dose distributions from a commercial treatment planning optimization system
    Ma, CM
    Pawlicki, T
    Jiang, SB
    Li, JS
    Deng, J
    Mok, E
    Kapur, A
    Xing, L
    Ma, L
    Boyer, AL
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (09): : 2483 - 2495
  • [9] Direct dose to water dosimetry for pretreatment IMRT verification using a modified EPID
    Gustafsson, Helen
    Vial, Philip
    Kuncic, Zdenka
    Baldock, Clive
    Denham, James W.
    Greer, Peter B.
    [J]. MEDICAL PHYSICS, 2011, 38 (11) : 6257 - 6264
  • [10] APPLICATION OF GAFCHROMIC EBT FILMS TO DOSIMETRIC VERIFICATION OF DOSE DISTRIBUTIONS IN IMRT
    Chelminski, K.
    Bulski, W.
    Georg, D.
    Bodzak, D.
    Maniakowski, Z.
    Oborska, D.
    Rostkowska, J.
    Kania, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : XXI - XXI