In Vitro Fatigue and Fracture Load of Monolithic Ceramic Crowns Supported by Hybrid Abutment

被引:0
|
作者
Nawafleh, Noor [1 ]
Elshiyab, Shareen [1 ]
Ochsner, Andreas [2 ,3 ]
George, Roy [4 ]
机构
[1] Jordan Univ Sci & Technol, Fac Appl Med Sci, Irbid 22110, Jordan
[2] Esslingen Univ Appl Sci, Fac Mech Engn, Esslingen, Germany
[3] Univ Newcastle, Fac Engn & Built Environm, Newcastle, NSW, Australia
[4] Griffith Univ, Sch Dent & Oral Hlth, Gold Coast, Qld, Australia
来源
OPEN DENTISTRY JOURNAL | 2021年 / 15卷
关键词
Chewing simulation; Fatigue; Fracture resistance; Lithium disilicate; Zirconia; Hybrid abutment; Dental implant; FIXED DENTAL PROSTHESES; LITHIUM DISILICATE CROWNS; ZIRCONIA-BASED CROWNS; MEAN FOLLOW-UP; SINGLE CROWNS; CLINICAL-EVALUATION; PREPARATION DESIGNS; CAD/CAM-MATERIALS; SURVIVAL RATES; MOLAR CROWNS;
D O I
10.2174/1874210602115010664
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: This study evaluated the performance of zirconia and lithium disilicate crowns supported by implants or cemented to epoxy resin dies. Methods: Eigthy zirconia and lithium disilicate crowns each were prepared and assigned in four groups according to the crown material and supporting structure combinations (implant-supported zirconia, die-supported zirconia, implant-supported lithium disilicate, and die-supported lithium disilicate). Ten crowns in each group acted as control while the rest (n=10) underwent thermocycling and fatigue with 100 N loading force for 1.5 million cycles. Specimens were then loaded to fracture in a universal testing machine. Data were analysed using one-way ANOVA and Tukey multiple comparison test with a 95% level of significance. Results: No implants or crown failure occurred during fatigue. The mean fracture load values (control, fatigued) in newton Were as follows: (4054, 3344) for implant-supported zirconia, (3783, 3477) for die-supported zirconia, (2506, 2207) for implant-supported lithium disilicate, and (2159, 1806) for die-supported lithium disilicate. Comparing the control with the corresponding fatigued subgroup showed a significantly higher fracture load mean of the control group in all cases. Zirconia showed a significantly higher fracture load mean than lithium disilicate (P=0.001, P<0.001). However, comparing crowns made from the same material according to the supporting structure showed no significant difference (P=0.923, P=0.337). Conclusion: Zirconia and lithium disilicate posterior crowns have adequate fatigue and fracture resistance required for posterior crowns. However, when heavy fatigue forces are expected, zirconia material is preferable over lithium disilicate. Zirconia and lithium disilicate implant-supported crowns cemented to hybrid abutments should have satisfactory clinical performance.
引用
收藏
页码:664 / 671
页数:8
相关论文
共 50 条
  • [41] How does the occlusal contact region influence the mechanical fatigue performance and fracture region of monolithic lithium disilicate ceramic crowns?
    Velho, Helder Callegaro
    da Rosa, Lucas Saldanha
    Dapieve, Kiara Serafini
    Borges, Alexandre Luiz Souto
    Pereira, Gabriel Kalil Rocha
    Venturini, Andressa Borin
    Valandro, Luiz Felipe
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2023, 140
  • [42] Fracture load of feldspar ceramic crowns: effects of surface treatments and aging
    Coldea, Andrea
    Stawarczyk, Bogna
    Meinen, John
    Lankes, Valerie
    Swain, Michael V.
    Roos, Malgorzata
    CLINICAL ORAL INVESTIGATIONS, 2025, 29 (01)
  • [43] Fracture load of all-ceramic CAD/CAM molar crowns
    Chen, HY
    Kunzelmann, KH
    Hickel, R
    JOURNAL OF DENTAL RESEARCH, 1998, 77 (05) : 1257 - 1257
  • [44] In-vitro fatigue and fracture testing of CAD/CAM-materials in implant-supported molar crowns
    Preis, Verena
    Hahnel, Sebastian
    Behr, Michael
    Bein, Laila
    RosentrittDepartment, Martin
    DENTAL MATERIALS, 2017, 33 (04) : 427 - 433
  • [45] Fracture strength of implant-supported hybrid abutment crowns in premolar region fabricated using different restorative CAD/CAM materials
    Takano, Ryoki
    Honda, Junichi
    Kobayashi, Tatsuro
    Kubochi, Kei
    Takata, Hiroki
    Komine, Futoshi
    DENTAL MATERIALS JOURNAL, 2023, 42 (02) : 187 - 192
  • [46] Fracture resistance and survival of implant-supported, zirconia-based hybrid-abutment crowns: Influence of aging and crown structure
    Elshiyab, Shareen H.
    Nawafleh, Noor
    Walsh, Laurence
    George, Roy
    JOURNAL OF INVESTIGATIVE AND CLINICAL DENTISTRY, 2018, 9 (04) : e12355
  • [47] Mechanical Stability of Screw-Retained Monolithic and Bi-Layer Posterior Hybrid Abutment Crowns after Thermomechanical Loading: An In Vitro Study
    Spitznagel, Frank A.
    Bonfante, Estevam A.
    Campos, Tiago M. B.
    Vollmer, Maximilian A.
    Boldt, Johannes
    Doerken, Sam
    Gierthmuehlen, Petra C.
    MATERIALS, 2021, 14 (24)
  • [48] Fracture strength of implant- supported monolithic crowns after aging in a mastication simulator
    Salvarli, Kaan Cevat
    Cevlik, Esra Talay
    INTERNATIONAL JOURNAL OF ESTHETIC DENTISTRY, 2023, 18 (01): : 38 - 48
  • [49] Monolithic zirconia dental crowns. Internal fit, margin quality, fracture mode and load at fracture
    Schriwer, Christian
    Skjold, Anneli
    Gjerdet, Nils Roar
    Oilo, Marit
    DENTAL MATERIALS, 2017, 33 (09) : 1012 - 1020
  • [50] Fracture load of ceramic restorations after fatigue loading
    Baladhandayutham, Balasudha
    Lawson, Nathaniel C.
    Burgess, John O.
    JOURNAL OF PROSTHETIC DENTISTRY, 2015, 114 (02): : 266 - 271