Variation in the total lengths of abutment/implant assemblies generated with a function of applied tightening torque in external and internal implant-abutment connection

被引:57
|
作者
Kim, Ki-Seong [1 ,2 ]
Lim, Young-Jun [1 ,2 ]
Kim, Myung-Joo [1 ,2 ]
Kwon, Ho-Beom [1 ,2 ]
Yang, Jae-Ho [1 ,2 ]
Lee, Jai-Bong [1 ,2 ]
Yim, Soon-Ho [3 ]
机构
[1] Seoul Natl Univ, Dept Prosthodont, Sch Dent, Seoul 110749, South Korea
[2] Seoul Natl Univ, Dent Res Inst, Sch Dent, Seoul 110749, South Korea
[3] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Prosthodont,Inst Oral Hlth Sci, Seoul, South Korea
关键词
dental implants; implant-abutment interface; repeated tightening; settling effect; tightening torque; SCREW JOINT STABILITY; OSSEOINTEGRATED IMPLANTS; COMPLICATIONS; MISFIT;
D O I
10.1111/j.1600-0501.2010.02063.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim: Settling (embedment relaxation), which is the main cause for screw loosening, is developed by microroughness between implant and abutment metal surface. The objective of this study was to evaluate and compare the relationship between the level of applied torque and the settling of abutments into implants in external and internal implant-abutment connection. Material and methods: Five different implant-abutment connections were used (Ext, External butt joint + two-piece abutment; Int-H2, Internal hexagon + two-piece abutment; Int-H1, Internal hexagon + one-piece abutment; Int-O2, Internal octagon + two-piece abutment; Int-O1, Internal octagon + one-piece abutment). All abutments of each group were assembled and tightened with corresponding implants by a digital torque gauge. The total lengths of implant-abutment samples were measured at each torque (5, 10, 30 N cm and repeated 30 N cm with 10-min interval) by an electronic digital micrometer. The settling values were calculated by changes between the total lengths of implant-abutment samples. Results: All groups developed settling with repeated tightening. The Int-H2 group showed markedly higher settling for all instances of tightening torque and the Ext group was the lowest. Statistically significant differences were found in settling values between the groups and statistically significant increases were observed within each group at different tightening torques (P<0.05). After the second tightening of 30 N cm, repeated tightening showed almost constant settling values. Conclusions: Results from the present study suggested that to minimize the settling effect, abutment screws should be retightened at least twice at 30 N cm torque at a 10-min interval in all laboratory and clinical procedures.
引用
收藏
页码:834 / 839
页数:6
相关论文
共 50 条
  • [41] Load fatigue performance of conical implant-abutment connection: effect of torque level and interface junction
    Gehrke, S. A.
    Serra, R. De Carvalho
    MINERVA STOMATOLOGICA, 2015, 64 (01): : 1 - 7
  • [42] Bacterial Leakage and Microgap along Implant-Abutment Connection in Three Different Implant Systems
    Bajoghli, Farshad
    Amjadi, Mehrak
    Akouchekian, Mohadese
    Narimani, Tahmineh
    INTERNATIONAL JOURNAL OF ADVANCED BIOTECHNOLOGY AND RESEARCH, 2016, 7 (04): : 1284 - 1290
  • [43] Microleakage of the prosthetic abutment/implant interface with internal and external connection: In vitro study
    Larrucea Verdugo, Carlos
    Jaramillo Nunez, Guido
    Acevedo Avila, Ariel
    Larrucea San Martin, Carlo
    CLINICAL ORAL IMPLANTS RESEARCH, 2014, 25 (09) : 1078 - 1083
  • [44] Micro-leakage at the implant-abutment interface with different tightening torques in vitro
    da Silva-Neto, Joao Paulo
    Prudente, Marcel Santana
    Prado Naves Carneiro, Thiago de Almeida
    de Arruda Nobilo, Mauro Antonio
    Amante Penatti, Mario Paulo
    das Neves, Flavio Domingues
    JOURNAL OF APPLIED ORAL SCIENCE, 2012, 20 (05) : 581 - 587
  • [45] Influence of repeated screw tightening on bacterial leakage along the implant-abutment interface
    do Nascimento, Cassio
    Pedrazzi, Vinicius
    Miani, Paola Kirsten
    Moreira, Larissa Daher
    de Albuquerque Junior, Rubens Ferreira
    CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 (12) : 1394 - 1397
  • [46] Very low certainty in evidence suggesting that internal implant-abutment connections result in negligible bone loss and similar survival and complication rates compared with external implant-abutment connections
    Brignardello-Petersen, Romina
    JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2018, 149 (06): : E95 - E95
  • [47] Fatigue Failure of Narrow Implants with Different Implant-Abutment Connection Designs
    Bordin, Dimorvan
    Witek, Lukasz
    Fardin, Vinicius P.
    Bonfante, Estevam A.
    Coelho, Paulo G.
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2018, 27 (07): : 659 - 664
  • [48] Impact of Intentional Overload on Joint Stability of Internal Implant-Abutment Connection System with Different Diameter
    Lee, Ji-Hye
    Lee, Won
    Huh, Yoon-Hyuk
    Park, Chan-Jin
    Cho, Lee-Ra
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2019, 28 (02): : E649 - E656
  • [49] The Evaluation of Unscrewing Torque Values of Implant-Abutment Connections: An In Vitro Study
    Bruna, Ezio
    Fabianelli, Andrea
    Mastriforti, Giacomo
    Papacchini, Federica
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2017, 30 (01) : 30 - 32
  • [50] Effects of Aging Torque Controllers on Screw Tightening Force and Bacterial Micro-Leakage on the Implant-Abutment Complex
    Jiries, Yousef
    Brosh, Tamar
    Matalon, Shlomo
    Perlis, Vladimir
    Ormianer, Zeev
    MATERIALS, 2022, 15 (02)