共 50 条
In-office assessment of blood pressure in chronic kidney disease: usual measurement versus automated BpTRU measurement
被引:12
|作者:
O'Shaughnessy, Michelle M.
[1
]
Newman, Christine A.
[2
]
Kinsella, Sinead M.
[1
]
Reddan, Donal N.
[1
]
Lappin, David W.
[1
]
机构:
[1] Natl Univ Ireland, Dept Nephrol, Hosp Ground, Galway Univ Hosp, Galway, Ireland
[2] Natl Univ Ireland, Dept Med, Galway, Ireland
关键词:
blood pressure measurement;
BpTRU;
chronic kidney disease;
hypertension;
renal transplant;
WHITE-COAT HYPERTENSION;
PROGNOSTIC IMPORTANCE;
DIABETIC-NEPHROPATHY;
PREVALENCE;
RECORDINGS;
OUTCOMES;
HOME;
D O I:
10.1097/MBP.0b013e328346e0db
中图分类号:
R6 [外科学];
学科分类号:
1002 ;
100210 ;
摘要:
Objectives The automated BpTRU device has been shown to improve the accuracy of in-office blood pressure assessment in hypertensive populations. We aimed to determine whether this was also true for patients with chronic kidney disease. Materials and methods We recorded the blood pressure of 80 hypertensive outpatients with chronic kidney disease by usual automated measurement and BpTRU automated measurement. We established whether there were any statistically significant differences in the absolute blood pressure values measured by either method and whether these differences had any impact on the assessment of blood pressure control. Results Systolic and diastolic blood pressures were significantly lower by BpTRU measurement than by usual measurement, by 10.1 +/- 12.2 mmHg (95% confidence interval: 7.4-12.8 mmHg, P<0.001) and 2.8 +/- 10.6 mmHg (95% confidence interval: 0.4-5.1 mmHg, P = 0.02), respectively. Significantly, more patients achieved their blood pressure target of 130/80 mmHg or less by BpTRU measurement than by usual measurement (72.5 vs. 48.8% for systolic blood pressure, P<0.001; 68.8 vs. 61.3% for diastolic blood pressure, P = 0.02). Systolic blood pressures remained significantly lower by BpTRU measurement than by usual measurement in all predefined study subgroups (estimated glomerular filtration rate <30 vs >= 30 ml/min/1.73m(2); transplant vs. nontransplant). We detected more hypotension by BpTRU measurement than by usual measurement. Conclusion Our study suggests that the BpTRU device can negate white coat effect in patients with chronic kidney disease. The use of this device in routine clinical practice could improve the overall accuracy of in-office blood pressure assessment in this high-risk population, minimizing the potential for undertreatment and overtreatment of hypertension. Blood Press Monit 16:124-128 (C) 2011 Wolters Kluwer Health | Lippincott Williams & Wilkins.
引用
收藏
页码:124 / 128
页数:5
相关论文