Serial intravascular ultrasonographic measurements after implantation of biodegradable polymer-coated stents in porcine coronary arteries

被引:2
|
作者
Fröhlich, G
Strehblow, C
Sperker, W
Yahya, N
Shirazi, M
Hevesi, A
Garamvölgyi, R
Hadjiev, J
Scherzer, T
Glogar, D
Gyöngyösi, M
机构
[1] Univ Vienna, Dept Cardiol, A-1090 Vienna, Austria
[2] Univ Kaposvar, Inst Diagnost Imaging & Radiat Oncol, Kaposvar, Hungary
关键词
coronary stent; drug-eluting stent; biodegradable polymer coating; intimal hyperplasia; in-stent restenosis;
D O I
10.1097/00019501-200308000-00009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Biodegradable stent coatings provide a potential for local drug delivery at the time of vascular injury, while possible tissue toxicity is avoided through constant degradation, leaving behind a bare metal stent. Design Serial three-dimensional (3D) intravascular ultrasonographic results on bare Megaflex stents; and biodegradable polymer-coated Megaflex stents (Hyper stents) (Eurocor, Bonn, Germany) were compared 1 and 4 weeks after intracoronary implantation in pigs. Methods Under general anaesthesia, the left anterior descending and circumflex coronary arteries of domestic pigs were stented with Megaflex and Hyper stents, using right femoral artery access. Control coronary angiography and intravascular ultrasonography (IVUS) were performed 1 and 4 weeks after stent implantation using left femoral artery access and right carotid artery access. After recording of angiographic and IVUS data, the pigs were allowed to recover. Results The 1- and 4-week IVUS follow-ups revealed less neointima formation with Hyper stents than with Megaflex stents: 1-week intimal volume, 11.8 +/- 0.93 compared with 15.02 +/- 4.18 mm(3), P=0.065; intimal area, 0.81 +/- 0.06 compared with 1.1 +/- 0.16 mm(2), p = 0.003; maximal intimal thickness, 0.12 +/- 0.01 compared with 0.14 +/- 0.02 mm, P =0.049; 4-week intimal volume, 12.4 +/- 1.77 compared with 27.32 +/- 12.79 mm(3), P =0.016; intimal area, 0.82 +/- 0.12 compared with 1.95 +/- 0.65 mm(2), P=0.003; and maximal intimal thickness, 0.13 +/- 0.04 compared with 0.30 +/- 0.10 m m, P = 0.003. Conclusions Implantation of biodegradable polymer-coated (Hyper) stents results in significantly less neointima formation 1 and 4 weeks after intracoronary implantation than with bare Megaflex stents. Taking advantage of the good collateralization of femoral and carotid arteries of pigs, the use of different arterial accesses allows serial angiographic and 3D IVUS measurements on neointimal development without sacrificing the animals.
引用
收藏
页码:409 / 412
页数:4
相关论文
共 50 条
  • [1] BIOCOMPATIBILITY OF BIODEGRADABLE AND NONBIODEGRADABLE POLYMER-COATED STENTS IMPLANTED IN PORCINE PERIPHERAL ARTERIES
    DESCHEERDER, IK
    WILCZEK, KL
    VERBEKEN, EV
    VANDORPE, J
    LAN, PN
    SCHACHT, E
    PIESSENS, J
    DEGEEST, H
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 1995, 18 (04) : 227 - 232
  • [2] Biodegradable polymer coated, paclitaxel eluting stents in the porcine coronary arteries
    Milewski, K
    Buszman, P
    Zurakowski, A
    Pajak, J
    Stojko, J
    Abu Samra, M
    AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (7A): : 23H - 23H
  • [3] Intense inflammatory reaction to heparin polymer-coated intravascular Palmaz stents in porcine arteries compared to uncoated Palmaz stents
    Goodwin, SC
    Yoon, HC
    Chen, G
    Abdel-Sayed, P
    Costantino, MM
    Bonilla, SM
    Nishimura, E
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2003, 26 (02) : 158 - 167
  • [4] BIOCOMPATIBILITY OF POLYMER-COATED OVERSIZED METALLIC STENTS IMPLANTED IN NORMAL PORCINE CORONARY-ARTERIES
    DESCHEERDER, IK
    WILCZEK, KL
    VERBEKEN, EV
    VANDORPE, J
    LAN, PN
    SCHACHT, E
    DEGEEST, H
    PIESSENS, J
    ATHEROSCLEROSIS, 1995, 114 (01) : 105 - 114
  • [5] ANGIOPEPTIN LOADED STENTS INHIBIT THE NEOINTIMAL REACTION INDUCED BY POLYMER-COATED STENTS IMPLANTED IN PORCINE CORONARY-ARTERIES
    DESCHEERDER, I
    WILCZEK, K
    VANDORPE, J
    VERBEKEN, E
    SCHACHT, E
    DESMET, W
    PIESSENS, JH
    FOEGH, M
    CIRCULATION, 1994, 90 (04) : 597 - 597
  • [6] Final five-year outcomes after implantation of biodegradable polymer-coated biolimus-eluting stents versus durable polymer-coated sirolimus-eluting stents
    Jakobsen, Lars
    Christiansen, Evald H.
    Maeng, Michael
    Hansen, Knud N.
    Kristensen, Steen D.
    Botker, Hans E.
    Terkelsen, Christian J.
    Jensen, Svend E.
    Raungaard, Bent
    Madsen, Morten
    Lassen, Jens F.
    Jensen, Lisette O.
    EUROINTERVENTION, 2017, 13 (11) : 1337 - 1345
  • [7] Novel biodegradable polymer-coated, paclitaxel-eluting stent inhibits neointimal formation in porcine coronary arteries
    Buszman, Pawel
    Milewski, Krzysztof
    Zurakowski, Aleksander
    Pajak, Jacek
    Liszka, Lukasz
    Buszman, Piotr
    Musiol, Ewa
    AbuSamra, Motaz
    Trznadel, Stanislaw
    Kaluza, Grzegorz
    KARDIOLOGIA POLSKA, 2010, 68 (05) : 503 - 509
  • [8] Implantation of metal phosphorylcholine coated stents in rabbit iliac and porcine coronary arteries
    Kuiper, KKJ
    Robinson, KA
    Chronos, NAF
    Nordrehaug, JE
    CIRCULATION, 1997, 96 (08) : 1607 - 1607
  • [9] Procedural Performance of Ultrathin, Biodegradable Polymer-Coated Stents Versus Durable Polymer-Coated Stents Based on Intracoronary Imaging
    des Plantes, Annemieke C. Ziedses
    Neleman, Tara
    Forero, Maria N. Tovar
    Visser, Leon
    Scoccia, Alessandra
    Groenland, Frederik T. W.
    Ligthart, Jurgen M. R.
    Boersma, Eric
    Nuis, Rutger-Jan
    den Dekker, Wijnand K.
    Wilschut, Jeroen
    Diletti, Roberto
    Zijlstra, Felix
    Van Mieghem, Nicolas M.
    Daemen, Joost
    JOURNAL OF INVASIVE CARDIOLOGY, 2022, 34 (11): : E811 - E819
  • [10] Sustained inhibition of neointimal hyperplasia after implanting novel biodegradable polymer-coated emodin-eluting stents in the porcine coronary model.
    Ge, J
    Liu, X
    Wang, KQ
    Qian, J
    Wang, X
    Jia, J
    Zhao, Y
    AMERICAN JOURNAL OF CARDIOLOGY, 2004, 94 (6A): : 151E - 152E