Experimental characterization of novel small animal PET detector modules based on scintillation crystal blocks read out by APD arrays

被引:0
|
作者
Maas, MC [1 ]
van der Laan, DJ [1 ]
Schaart, DR [1 ]
Huizenga, J [1 ]
Brouwer, JC [1 ]
Bruyndonckx, P [1 ]
Léonard, S [1 ]
Lemaître, C [1 ]
van Eijk, CWE [1 ]
机构
[1] Delft Univ Technol, IRI Radiat Technol, NL-2629 JB Delft, Netherlands
关键词
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Minimizing dead space is one way to increase the detection efficiency of small-animal PET scanners. By using monolithic blocks of scintillating material (e.g. 20 mm X 10 nun X 10 mm LSO), loss of efficiency due to inter-crystal reflective material is minimized. Readout of such blocks can be performed by means of one or more avalanche photo-diode (APD) arrays optically coupled to the block. The primary event position and depth of interaction (DOI) information are derived from the measured distribution of the scintillation light over the APD array(s). We are studying the feasibility of such detector modules, both by simulation and by experiment. A 64-channel setup for testing the above type of detector modules has been developed. Experiments to verify the effect of crystal surface finish, detector geometry and reconstruction algorithm parameters on the spatial resolution have been carried out. The first results of these experiments are presented in this paper, and compared to simulation results. This research is conducted in collaboration with the Crystal Clear Collaboration (CCC).
引用
收藏
页码:2942 / 2946
页数:5
相关论文
共 20 条
  • [1] Experimental characterization of monolithic-crystal small animal PET detectors read out by APD arrays
    Maas, Marnix C.
    van der Laan, D. J.
    Schaart, Dennis R.
    Huizenga, Jan
    Brouwer, J. C.
    Bruyndonckx, Peter
    Leonard, Sophie
    Lemaitre, Cedric
    van Eijk, Carel W. E.
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (03) : 1071 - 1077
  • [2] A multilayer detector for PET based on APD arrays and continuous crystal elements
    McCallum, S
    Clowes, P
    Welch, A
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 2898 - 2902
  • [3] Wavelength shifter strips and G-APD arrays for the read-out of the z-coordinate in axial PET modules
    Braem, A.
    Chesi, E.
    Joram, C.
    Rudge, A.
    Seguinot, J.
    Weilhammer, P.
    De Leo, R.
    Nappi, E.
    Lustermann, W.
    Schinzel, D.
    Johnson, I.
    Renker, D.
    Albrecht, S.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 586 (02): : 300 - 308
  • [4] Performance of APD-based monolithic-crystal detectors for small animal PET
    Maas, Marnix C.
    van der Laan, D. J.
    Schaart, Dennis R.
    Bruyndonckx, Peter
    Lemaitre, Cedric
    van Eijk, Carel W. E.
    [J]. 2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2017 - 2021
  • [5] Wave length shifter strips and G-APD arrays for the read-out of the z-coordinate in axial PET modules
    Braem, A.
    Chesi, E.
    Joram, C.
    Rudge, A.
    Seguinot, J.
    Weilhammer, P.
    De Leo, R.
    Nappi, E.
    Lustermann, W.
    Schinzel, D.
    Johnson, I.
    Renker, D.
    Albrecht, S.
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2601 - +
  • [6] High resolution detector modules based on NaI(Tl) arrays for small animal imaging
    Weisenberger, AG
    Majewski, S
    Popov, V
    Wojcik, R
    [J]. 2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1540 - 1544
  • [7] Use of single photon counting detector arrays in combined PET/MR: Characterization of LYSO-SiPM detector modules and comparison with a LSO-APD detector
    Spanoudaki, V. C.
    Mann, A. B.
    Otte, A. N.
    Konorov, I.
    Torres-Espallardo, I.
    Paul, S.
    Ziegler, S. I.
    [J]. JOURNAL OF INSTRUMENTATION, 2007, 2
  • [8] Small PET System Using Position Sensitive CdTe Detector Blocks for Experimental Animal Study
    Kikuchi, Y.
    Ishii, K.
    Yamazaki, H.
    Matsuyama, S.
    Nakhostin, M.
    Terakawa, A.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 : S415 - S415
  • [9] Small animal PET prototype based on LSO crystals read out by avalanche photodiodes
    Fries, O.
    Bradbury, S.M.
    Gebauer, J.
    Holl, I.
    Lorenz, E.
    Renker, D.
    Ziegler, S.I.
    [J]. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1997, 387 (1-2): : 220 - 224
  • [10] A small animal PET prototype based on LSO crystals read out by avalanche photodiodes
    Fries, O
    Bradbury, SM
    Gebauer, J
    Holl, I
    Lorenz, E
    Renker, D
    Ziegler, SI
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 387 (1-2): : 220 - 224